Recommended Separation Distances between portable and mobile RG Communications
equipment and bed.
Equipment and Systems that are NOT Life-Supporting
Recommended Separation Distances for the SW Advantage™: The SW Advantage™ is
intended for use in the electromagnetic environment in which radiated disturbances are
controlled. The customer, or user, of the SW Advantage™ can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
Communications Equipment and the bed, as recommended below, according to the maximum
output power of the communications equipment.
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
Compliance Level Cond RF 3 Rad RF-800 MHz 3 Rad RF – 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
General Safety
WARNING:
The bed is to be used in accordance with each facility’s policies
and procedures.
CAUTION:
Possible fire hazard when used with oxygen administering
equipment other than the nasal, mask or 1/2 bed length type tent. Oxygen
tent should not extend below the mattress support level. Lock hand control at
foot of bed when using oxygen administering equipment.
WARNING:
Electric shock may occur when plugging the bed into the
wall outlet. Use ONLY grounded or hospital grade outlets.
DO NOT use the bed if the power cord is cut, frayed or loosely connected to
the bed.