Replacement parts, such as cables and accessories, must be purchased through Sizewise to
ensure proper compliance requirements.
WARNING:
Using other manufacturer’s cables and accessories may affect EMC
performance. Unauthorized use of these items will void warranty and could result
in patient and/or user injury as well as damage to equipment or other property.
The use of cables or accessories other than those for which the bed was designed
or tested can significantly degrade emissions and immunity performance.
Cables and Accessories
Specifications and Requirements
Hand Control
Length:
10
.
49
feet (
3.2
meters)
Usage temperature: 5º C to 40º C
Storage temperature: -10º C to +50º C
Approvals: IEC60601-1, ANSI/AAMI ES60601-1 and
CAN/CSA-22.2 No 60601-1.
HB80 is designed and tested in accordance
with EN60601-2-52 cl. 201.11.6.6.101 (Machine washable medical beds)
Power Cord
Hospital Grade
Length:
10.49 feet (3.2 meters)
SJT 16/3C 60°C (140°F)
End Types: YP-18L-2 and YC-12
UL and CSA certified 3 prong IEC C13 power cord receptacle. Built
to the IEC 60320-C13 standard. Rated 13A 125V/250V. This IEC
C13 power cord receptacle is fully molded with a low profile
ergonomic design and fully RoHS and REACH compliant.
Electromagnetic Interference (EMI)
The SW
Advantage™ has been tested and is intended for safe use with other components
compliant to IEC 601 standards for medical devices. The bed can cause interference with
non-601 regulated equipment and/or other non-601 regulated equipment can cause interference
with the bed. This section is in accordance with IEC 60601-1-2 section 5.2.2.
WARNING:
Radio wave sources such as radio stations, TV stations, amateur
radio (HAM) transmitters, two-way radios and cellular phones can affect this bed.
Following the warnings listed for electromagnetic interference should reduce the
chance of the bed’s unintended movement that could result in serious injury.
DO NOT turn ON hand-held personal communication devices, such as citizens band (CB) radios
and cellular phones while the bed is turned ON.
Be aware of nearby transmitters, such as radio or TV stations, and try to avoid coming close to
them.
If unintended movement occurs, turn the bed OFF as soon as it is safe.