001118v4.0 - US EN
Sirius Medical Systems
Sirius Pintuition Detector
Instructions for Use
31/32
8
T
ROUBLESHOOTING
8.1
T
ROUBLESHOOTING
E
RRORS
On start-up and during normal operation, the Pintuition Detector performs various self-checks
on its internal components. Detected problems will be indicated with an error code. In case of
an error, the user should turn off the power, wait for 30 seconds and turn on the power.
If the error persists, write down the error code and software version identifiers indicated on
the main screen (see section 3.6 (Turning the Device On)) and contact Sirius Medical Systems
or your local distributor for assistance.
8.2
T
ROUBLESHOOTING
T
IPS
Problem
Possible cause
Display feedback
Possible solution
Base unit does not
turn on
AC wall outlet has no
power
Nothing.
Check connections
Blown fuse
Request a technician to
check fuse, see Section
7.2.
Power cable not
connected properly
Check power cable
No sound
System is muted
Adjust the volume (see
Section 5.4)
Speaker not
functioning
N.A.
Restart Base Unit, system
should produce audio on
start-up
Base unit not
responding or
locked inoperable
Interference by
electromagnetic
radiation
Non-responsive
Increase distance between
other electrical devices
and Detector and
1. Reconnect Probe, or
2. Restart Base Unit
Base unit does not
show Probe
connected
Probe not or not
connected correctly to
Base Unit
Reconnect Probe (see
Section 5.1)
Restart Base Unit
No response to
Pintuition Seed
Seed is out of range
N.A.
Maneuver the Probe in
closer proximity to the
Seed, see Section 5.6 for
additional information.
Detection range is
reduced due to large
metallic objects near
Probe tip
Keep any metal
instruments, objects, or
other magnetic
disturbances at least 2-5
centimeters away from the
tip of the Probe
Detector needs
calibration
Perform calibration
(Section 5.2).
Detected distance
deviates more than
prescribed during
repeated system
test
System failure
N.A.
Cease to use this system.
Obtain Sirius Medical
Systems technical
assistance or contact your
local distributor. See next
section.