001118v4.0 - US EN
Sirius Medical Systems
Sirius Pintuition Detector
Instructions for Use
27/32
7
T
ECHNICAL
S
PECIFICATIONS
Pintuition Detector
(Probe and Base Unit)
Detection range
Up to 50 mm
Distance accuracy
0-20mm +/-1mm
20-30mm +/-2mm
30-40mm +/-4 mm
40-50mm +/-5 mm
Mode of operation
Continuous operation
Operating temperature range
15
–
25ºC (59
–
77 ºF)
Operating relative humidity range
15% - 90%, non-condensing
Operating altitude
<3000 meters (< 9842 feet)
Operating, Storage and Transportation
atmospheric pressure range
70 - 106 kPa (10.2
–
15.4 psi)
Storage and transport temperature range
-25
–
70
º
C (-13
–
158 ºF)
Storage and Transportation relative humidity
range
<90%, non-condensing
Protection against electrical shock
Class I Protectively Earthed
Service life
5 years
Pintuition Base Unit
Dimensions (W x H x D)
26 x 16 x 32 cm (10.2 x 6.3 x 12.6 in)
Weight
6 kg (13.2 lbs)
Ingress protection
IP3X
Power Lead Connector
IEC 60320 C13
Rated input
100-240 V AC ~ 50/60 Hz 1.2A
Fuse type and Dimensions (Ø x L)
T3.15AH / 250 V
5 mm x 20 mm (0.2 x 0.8 in)
Pintuition Probe (Applied Part; type BF)
Dimensions
Ø 18mm (tip); (0.71 in)
180 mm (length); (7.1 in)
Weight (including cable)
165 g (5.82 oz)
Cable Length
3m (9.8 ft)
Ingress protection
IP44
7.1
C
OMPLIANCE WITH
S
TANDARDS
This product has been designed in accordance with the following standards on electrical
safety:
•
IEC 60601-1 (Edition 3.1).
•
IEC 60601-1-2 (Edition 4).
The Pintuition Detector has no essential performance according to the IEC 60601-1 (Edition
3.1) definition.