001118v4.0 - US EN
Sirius Medical Systems
Sirius Pintuition Detector
Instructions for Use
15/32
ID
Item
Description
A2
Distance to Seed
A beeping sound is generated, the pause between
beeps is dependent on the detected distance
between the Seed and the Probe
3.3
T
RANSPORT AND
S
TORAGE
Caution.
Care should be taken not to drop the Pintuition Probe, or subject it to
any form of rough physical handling, either during normal use or during storage
and transportation. If the Probe is dropped, confirm correct functioning with the
supplied Test Kit.
Caution.
Care should be taken not to drop the Pintuition Base Unit, or subject it to
any form of rough physical handling, either during normal use or during storage
and transportation. If the Pintuition Base Unit is dropped confirm correct
functioning with a specialist
3.4
I
NSTALLATION
The Pintuition Base Unit (
1
) should be placed on a stable and level working surface, preferably
out of direct sunlight. Refer to Section 7 (Technical Specifications) for environmental
limitations of use.
Warning.
The Pintuition Detector should not be used adjacent to or stacked with
other equipment; if adjacent or stacked use is necessary, the Pintuition Detector
needs to be observed to verify normal operation in the configuration in which it
will be used.
3.5
P
OWERING THE
D
EVICE
Warning.
The Base Unit must be properly grounded to ensure patient safety. To
avoid the risk of electrical shock, this equipment must only be connected to
mains supply with protective earth. To avoid electrical shock, plug the power cord
into a properly wired receptacle, use only the power cord supplied with the Base
Unit and make sure the power cord is in good condition.
1.
Make sure that the Pintuition Base Unit
(1)
is switched off before connecting the
power cord, by switching the Power Switch
(6)
on the back side
to ‘Off’ (or O).
2.
The instrument is powered via a standard mains outlet supply. Make sure that the AC
outlet is properly grounded and supplies the specified voltage and frequency, see
Section 7 (Technical Specifications) for details.
3.
Connect the female connector end of the power cord to the AC power inlet
(8)
on the
back of the Base Unit
(1)
. Plug the male connector end of the power cord into a
properly grounded AC power outlet.
3.6
T
URNING THE
D
EVICE
O
N
To turn the instrument on:
1.
Push the Power Switch
(6)
at the back of the Pintuition Base Unit
(1)
from the ‘0’
position to the ‘1’ position
.
2.
While the Pintuition Base Unit
(1)
is powering up, the display
(2)
will show a start-up
screen containing the Sirius Medical Systems logo and the Base Unit software version
(left bottom corner) and Probe software version (right bottom corner).