19
External magnetic fields
Failures and data loss occur if strong magnetic fields influence the internal
measuring system.
`
Protect the magnetic sensor from impact by external magnets.
3 Identification
Please check the particular type of unit and type number from the identifi-
cation plate. Type number and the corresponding version are indicated in
the delivery documentation.
e. g.
MSK210
-
0023
variant number
type of unit
Identification by printing on the tape.
e. g.
MBxxxx GEK WT RP NNNNNN
batch number
reference point
material carrier tape
accuracy
MB type
4 Installation
4.1 Mechanical mounting
Magnetic sensor failure
`
When mounting pay attention to the IP type of protection (see chap-
ter
).
`
Do not open the sensor yourself.
`
Avoid blows to the unit.
`
Do not modify the device in any way.
`
The cable must be provided with strain relief. If necessary use drag
chain or protective sleeve.
Loss of measured values
`
The tolerance and gap measures be observed over the whole measur-
ing length.
Exemplary sensor figures are valid for all sensor types unless described
separately.
CAUTION
CAUTION
NOTICE
NOTICE