Operating the Calibrator / Micro Calibration Bath
Setting values for „SLor“ and „SLoF“
Calibrator type
Heating gradient
1)
„SLor“
Cooling gradient
2)
„SLoF“
Heating/Cooling:
TP 17 165 S
< 7 °C/min
< 5 °C/min
TP M 165 S / TP 17 166 /TP 17 200
- with silicone oil 20CS
- with distilled water
- as dry block
- as infrared calibrator
- as surface calibrator
< 3 °C/min
< 5 °C/min
< 3 °C/min
< 3 °C/min
< 3 °C/min
< 6 °C/min
< 4 °C/min
< 4 °C/min
< 4 °C/min
< 4 °C/min
Heating:
TP 17450 S, TP 17650 S
< 35 °C/min
max. … 300 °C
300 °C … 100 °C
< 10 °C/min
< 5 °C/min
TP 17450 S
- as infrared calibrator
- as surface calibrator
< 35 °C/min
< 30 °C/min
max. … 300 °C
300 °C … 100 °C
max. … 300 °C
300 °C … 100 °C
< 10 °C/min
< 5 °C/min
< 10 °C/min
< 5 °C/min
TP M 225 S
- with silicone oil 20CS
- with distilled water
- as dry block
- as infrared calibrator
- as surface calibrator
< 22 °C/min
< 12 °C/min
< 12 °C/min
< 12 °C/min
< 10 °C/min
200 °C … 50 °C
50 °C … 30 °C
90 °C … 50 °C
50 °C … 30 °C
200 °C … 50 °C
50 °C … 30 °C
200 °C … 50 °C
50 °C … 30 °C
200 °C … 50 °C
50 °C … 30 °C
< 4 °C/min
< 0,5 °C/min
< 2 °C/min
< 0,5 °C/min
< 2 °C/min
< 0,5 °C/min
< 2 °C/min
< 0,5 °C/min
< 2 °C/min
< 0,5 °C/min
1)
Heating gradient „SLor“:
The heating gradient "SLor" is active if the reference temperature is lower than the set tem-
perature. Each calibrator type has a max. heating capacity, meaning that only settings < than
this heating capacity are reasonable and extend the time until the set temperature is
achieved.
2)
Cooling gradient „SLoF“:
The cooling gradient "SLor" is active if the reference temperature is higher than the set tem-
perature. Only settings below the cooling capacity of the calibrator have an effect on the cool-
ing gradients.
Duration time „dur.t“:
The duration time "dur.t" is active if the set temperature SP1 has been achieved. Subse-
quently, the calibrator / micro calibration bath automatically switches to set temperature SP2.
ACTIVATE TEMPERATURE PROFILE!
If you have carried out settings for these three settings, the calibrator / micro calibration
bath uses the new values only when modifying the set temperature or switching the cali-
brator off and on again.
A further procedure is to switch off the automatic control prior to modifying parameters
and to switch it on again afterwards (
§ 7.3.1).