Operating the Calibrator / Micro Calibration Bath
© SIKA • Ba_TP17-TPM-Multi_en.doc • 01/2014
Retrieving the saved set temperatures (S version):
The saved set temperatures can be retrieved in calibration mode.
Press
the
U
key for approx 2 sec., the current set
point memory opens.
The current reference temperature appears on the
upper display.
The set point memory SP… appears on the lower
display for 2 sec. followed by the current set tem-
perature.
To receive another saved set point SP1, SP2, SP3
or SP4, press the
U
key again.
The selected temperature value is immediately
adopted and targeted.
7.3.4 Setting a gradient control and a temperature profile (S version)
It is possible to carry out a gradient control yourself and thus determine the time in which the
set temperature is reached. The time can be shorter or longer than the time usually required
by the calibrator / micro calibration bath.
When modifying the set temperature or switching on the calibrator / micro calibration bath it is
automatically determined which of the gradients (heating gradient “SLor” or cooling gradient
“SLoF”) is to be used.
Additionally, you can ensure that the calibrator / micro calibration bath switches to the set
temperature in set point memory SP2 as soon as the set temperature in set point memory
SP1 has been achieved and after a programmed duration time “dur.t”; this creates a simple
temperature profile.
After switching on the calibrator / micro calibration bath the temperature profile is automati-
cally carried out.
Gradient control and temperature profile
Retrieving the set temperatures display