Operating the Calibrator / Micro Calibration Bath
© SIKA • Ba_TP17-TPM-Multi_en.doc • 01/2014
7.2 Setting a temporary set temperature (set point mode)
In this operating mode it is possible to temporarily modify a saved set temperature.
Press the P key shortly.
The currently active set point memory, e.g. SP 2
(set point 2), is displayed by the upper display.
The respective set temperature is displayed by the
lower display.
Press
the
key to increase the set temperature.
Press the
key to decrease the set temperature.
Press the P key again to confirm the new set point.
NOTES:
B Press the
and
key to raise and lower the value by 0.1 respectively. If the
keys are held pressed for at least one second, the value increases or decreases
quickly and after two seconds even more quickly; this means the desired value can
be reached rapidly.
If no key is pressed in the
set point mode
for approx. 15 seconds, the device auto-
matically returns to the
calibration mode
.
Temporary set temperature setting