Electrical installation
INSTRUCTION MANUAL AquaScat 2
11778E/1
23
5.2.
Installation procedure
You can access the terminals by removing the front cover. Proceed as follows:
WORKSTEP
PROCEDURE
1.
Loosen the five screws on the front cover
with a 7 mm key and remove the front cover.
2.
Establish the electrical connections as de-
scribed in Section 5.3.
3.
When the installation is completed, assemble
the photometer in the reverse order.
In doing so, carefully mount the front cover
and fasten with the five screws.
Damage to the threaded inserts in
the housing due to excessive tightening
of the front cover screws:
Use a hex key without a T-handle to tighten
the screws of the front cover finger-tight
(approx. tightening torque 1 Nm).
7 mm hex key
The use of operating signals is described in the Reference Manual.
Section 2