Mounting the measuring device
INSTRUCTION MANUAL AquaScat 2
11778E/1
21
4.6.
Installation of the optional accessory parts
4.6.1.
Information about installing the optional
accessory parts
Note the following when connecting the optional accessory parts:
All hose connections have to be secured with hose clamps. The integrity of the con-
nections should be checked about 2 weeks after installation to ensure that no air can
be pulled in.
The venting hose of the deaeration tube must be neither shortened nor bent.
Make sure that the water can drain from the deaeration tube; the venting hose
should point downward as shown in Figure 1.
A minimum flow of 1 l/min. must be guaranteed on the overflow of the level control.
The dimensions of the drawings below must be adhered to (course and radii of the
hoses, spacing position of the deaeration tube to the photometer etc.).
4.6.2.
Mounting the deaeration tube
Mounting the complete deaeration tube is carried out in accordance with drawings
AQUASCAT/ER/1-MB
and
AQUASCAT/ER-MB
.
4.6.3.
Mounting the level control
Mounting the level control is carried out in accordance with drawings
AQUASCAT/NR-MB
and
NR2_AS-MB
.
4.6.4.
Mounting with deaeration tube and level control
Mounting the deaeration tube and level control is carried out in accordance with drawings
AQUASCAT/ER/2-MB
,
AQUASCAT/ER-MB
and
NR2_AS-MB
.
CAUTION!