INSTRUCTION MANUAL AquaScat 2
Operation
42
11778E/1
7.8.
Setting the current outputs
MANIPULATION
SUPPLEMENTARY INFORMATION
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Curr. outputs
button.
4.
Select current output 1 or 2.
5.
Select
C1 Turb
source.
For
Inactive
the current output is
switched off.
6.
Select range.
MR1 .. MR8 (see table below for
range definition)
In 1, In 2, Auto 1, Auto 2
Reference Manual
7.
Press the
Meas
button.
Instrument again in normal opera-
tion.
MEASURING
RANGE NO.
MEASURING RANGE
STANDARD
WTM/WTM A
MEASURING
RANGE
STANDARD HT
MEASURING RANGE
(CUSTOMER SPECIFIC)
1
0 .. 500
0 .. 4000
2
0 .. 100
0 .. 2000
3
0 .. 30.0
0 .. 1000
4
0 .. 10.0
0 .. 500
5
0 .. 3.00
0 .. 200
6
0 .. 1.00
0 .. 100
7
0 .. 0.30
0 .. 30
8
0 .. 0.10
0 .. 10
If other measuring ranges are needed, you can re-program the table above as required.