Servicing
INSTRUCTION MANUAL AquaScat 2
11778E/1
63
WORKSTEP
PROCEDURE
8.
Put the checking unit in the
adjustment position.
8.1: Change over the instrument to service operation
as described in Section 7.5.
8.2: In the
Simulation
menu, select the
Motor posi-
tion
submenu. Now select the
Adjustment
menu
item.
The checking unit now moves to the adjustment posi-
tion.
9.
Remove the optics unit of
the photometer and fasten
onto the docking station.
Section 4.4
10. Clean the head of the check-
ing unit.
10.1: Remove the two screws (arrows) and then re-
move the head of the checking unit.
10.2: Clean the two glass panes (arrows) with a cot-
ton-tipped applicator moistened with ethanol.
Position glass panel 1
Position glass panel 2