Servicing
INSTRUCTION MANUAL AquaScat 2
11778E/1
59
WORKSTEP
PROCEDURE
7.2: Tighten the inlet tube (B) with the conduit gland (A).
The inlet tube should protrude 5 mm out of the holder.
8.
Replace the optics unit
back on the measuring
cell unit and lock with the
four mounting clips.
Pay attention to the guide
pins (see picture).
8.6.
Check and clean optional accessory parts
Damage to components due to aggressive cleaning agents.
Do not use aggressive cleaning agents when cleaning optional accessory parts. Commercially
available dishwashing liquid can be used if needed.
When checking for cleanliness of the optional accessory parts, the following components
must be checked for particle residues:
Level control tube
Flow meter and regulator valve if fitted
Deaeration tube (pay particular attention to the connections to the hoses)
Hoses
If particle residues are found, rinse or backflush the concerned components with clean wa-
ter. Use commercially available dishwashing liquid if needed.
CAUTION!