INSTRUCTION MANUAL AquaScat 2
Servicing
66
11778E/1
8.9.
Change battery
Life threatening voltage inside the instrument:
Connecting the electrical lines is extremely dangerous. Parts of the system can also be dam-
aged. Local regulations must be observed at all times for electrical installations.
WORKSTEP
PROCEDURE
1.
Interrupt the service voltage to the pho-
tometer.
Section 5
2.
Loosen the five screws on the front cover
with a 7 mm key and remove the front
cover.
3.
Remove the old battery and replace with a
new one (circle).
The battery is built into the front
cover on the connection print (AQ2Conn).
DANGER!