40
Programming the Pump
SIGMA Manual Number 41018
Revision AR
Software version 6.01.00 and later:
To titrate flow rates and VTBI without stopping the pump:
1.
Press the
TITRATE
softkey.
2.
To titrate dose, press
DOSE
softkey.
3.
To titrate rate mL/hr, press
mL/hr
softkey.
4.
Observing the displayed hard and soft dose or mL/hr flow rate limits, enter a new dose or
mL/hr flow rate.
5.
Press
OK
.
If soft limits are exceeded, press
YES
to accept new value or
NO
to return to prior value.
6.
To titrate VTBI, press
VTBI
softkey, enter new value, and press
OK
.
ICU
10
options
review
DOPamine*
mcg/kg/min
titrate
bolus