Messages
and Diagnosis
7.5 Message List Data and Operating Errors (Synchronous Messages)
Error
No.
Data and operating errors
- Meaning
Description
1
Unknown command code
The SIWAREX does not recognize the command
code or the data record or is unable to process the
command or data record in the current operating
status.
2
Unknown data record
The specified data record does not exist.
4 Reserved
5
Parameter transfer, transfer of
int. process values and
adjustment commands not
permitted with active write
protection
The adjustment parameters (DS3), the int. process
values (DS26) and the adjustment commands may
only be transferred if write protection is inactive
(jumper inserted between connecting terminals 1-5).
First deactivate the write protection.
7
Command can only be
executed at standstill
The command can only be executed when the scale
is at a standstill. Wait until it is at a standstill or
change the standstill parameters in DS3.
8
Minimum time span between
two commands not observed
There must be a waiting time of at least 5 seconds
between two successive commands. This applies to
the command codes 1, 2, 3, 4, 5, 8.
20
Command only permissible with
adjusted scale
Scale commands can only be executed if the
module has been calibrated.
First calibrate the module.
21
Module failure
Command is not permissible due to a module failure.
77
Reserve in DS3 – Content not
permitted.
Enter “----”.
78
Reserved
Reserved
80
Incorrect standstill range
The specified standstill range is not plausible.
81
Incorrect decimal place
The defined decimal point is not plausible for the
remote display. 0.....5 is permitted.
82
Numeral step not plausible
Numeral step is incorrect in DS 3.
The following values are possible as the numeral
step: 20, 10, 5, 2, 1.
83
Filter parameter implausible
The filter parameter definition is not plausible in DS 3.
Check the limit frequency (selection 0...9).
84
Characteristic value not
plausible
The characteristic value in DS3 is incorrect. Set the
parameter to a valid value (1, 2 or 4).
85
Adjustment weight implausible
At least one weight setting for scale adjustment is
not plausible.
86
Adjustment weight error
The defined adjustment weights are incorrect. The
weight values must be in ascending order or 0 if not
used.
87
Defined percentage incorrect
A percentage definition is incorrect. A definition
exceeding 100% is not permitted for defining the
zero set value or the maximum tare value.
88
Negative value not permitted
A parameter contains a negative value that is not
permitted.
89
Selection code not permitted for
the TTY interface
The selection code for the TTY interface is not
permitted. Set a valid value (0 or 1).
SIWAREX MS 7-39