Operate
6.2 Parameter description
SITRANS P500 with HART
92
Operating Instructions, 12/2014, A5E02344528-09
Description:
The parameter defines the characteristic of the output flow.
"linear"
The output flow is proportional to the differential pressure.
"srlin"
Up to the root application point, the output flow is proportional to
the differential pressure and thereafter it corresponds to the root
function.
"sroff"
The output flow is 0 until the root application point, and thereafter it
corresponds to the root function.
A low flow cut-off.
"srlin2"
The output flow is proportional to the flow, two-stage linear up to
the root application point. The first linear section ranges from the
zero point to 0.6% of the output value and 0.6% of the pressure
value. The second linear section runs with a higher gradient to the
root application point and then in accordance with the root function.
See also
Flow characteristic (Page 155)
"2.1.5.3 Squareroot Startingpoint"
Factory setting:
5 %
Setting range:
•
Min 1 %
•
Max 15 %
Purpose:
Setting the root application point
Description:
The parameter defines as of when the standardised output value and hence
the output flow is output as the root function of standardised pressure.
Depending on the decision made under "2.1.5.2 Transfer Function"
(Page 91), this parameter is displayed or hidden.
"2.1.5.4 Mass Flow Unit"
Factory setting:
kg/s
Adjustment range:
g/s, g/min, g/h, kg/s, kg/min, kg/h, kg/day, MetTon/min, MetTon/h, Met-
Ton/day, lb/s, lb/min, lb/day, ShTon/min, ShTon/h, ShTon/day, LTon/h,
LTon/day
Purpose:
Selection of the unit for the mass flow measurement type
Description:
This parameter allows you to select the unit for the mass flow measurement
type. The parameters associated with this unit are automatically converted
into the new unit.
"2.1.5.5 Mass Flow LRV"
Factory setting:
0 kg/s
Adjustment range:
Freely selectable numeric value
Содержание SITRANS P500
Страница 1: ......
Страница 2: ......
Страница 10: ...Table of contents SITRANS P500 with HART 10 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 14: ...Introduction 1 6 Notes on warranty SITRANS P500 with HART 14 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 214: ...Technical data 11 9 HART communication SITRANS P500 with HART 214 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 218: ...Dimension drawings 12 2 Level SITRANS P500 with HART 218 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 221: ...Appendix A 3 Overview SITRANS P500 with HART Operating Instructions 12 2014 A5E02344528 09 221 Continued on next page ...
Страница 222: ...Appendix A 3 Overview SITRANS P500 with HART 222 Operating Instructions 12 2014 A5E02344528 09 Continued on next page ...
Страница 223: ...Appendix A 3 Overview SITRANS P500 with HART Operating Instructions 12 2014 A5E02344528 09 223 Continued on next page ...
Страница 224: ...Appendix A 3 Overview SITRANS P500 with HART 224 Operating Instructions 12 2014 A5E02344528 09 Continued on next page ...
Страница 225: ...Appendix A 3 Overview SITRANS P500 with HART Operating Instructions 12 2014 A5E02344528 09 225 Continued on next page ...
Страница 226: ...Appendix A 3 Overview SITRANS P500 with HART 226 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 228: ...Appendix A 4 Technical support SITRANS P500 with HART 228 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 230: ...List of abbreviations SITRANS P500 with HART 230 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 234: ...Glossary SITRANS P500 with HART 234 Operating Instructions 12 2014 A5E02344528 09 TP Total Performance ...
Страница 238: ......