Purpose:
Activate/ deactivate the alarm
Description:
This parameter allows you to activate the limit transmitter alarm.
See also
Limit monitor (Page 166)
"3.8.1.2 Warning enable"
Factory setting:
Off
Adjustment range:
•
Off
•
Lower limit
•
Upper limit
•
Upper and lower limit
Purpose:
Activate/ deactivate a warning
Description:
This parameter allows you to activate the limit transmitter alarm.
See also
Limit monitor (Page 166)
"3.8.1.3 Lo Value"
Factory setting:
0
Purpose:
Setting the lower warning value
Description:
This parameter allows you to set the value of the lower warning limit in the
unit of the selected measurement type.
"3.8.1.4 LoLo Value"
Factory setting:
0
Purpose:
Setting the lower alarm value
Description:
This parameter allows you to set the value of the lower alarm limit in the unit
of the selected measurement type.
"3.8.1.5 Hi Value"
Factory setting:
0
Purpose:
Setting the upper warning value
Description:
This parameter allows you to set the value of the upper warning limit in the
unit of the selected measurement type.
Содержание SITRANS P500
Страница 1: ......
Страница 2: ......
Страница 10: ...Table of contents SITRANS P500 with HART 10 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 14: ...Introduction 1 6 Notes on warranty SITRANS P500 with HART 14 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 214: ...Technical data 11 9 HART communication SITRANS P500 with HART 214 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 218: ...Dimension drawings 12 2 Level SITRANS P500 with HART 218 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 221: ...Appendix A 3 Overview SITRANS P500 with HART Operating Instructions 12 2014 A5E02344528 09 221 Continued on next page ...
Страница 222: ...Appendix A 3 Overview SITRANS P500 with HART 222 Operating Instructions 12 2014 A5E02344528 09 Continued on next page ...
Страница 223: ...Appendix A 3 Overview SITRANS P500 with HART Operating Instructions 12 2014 A5E02344528 09 223 Continued on next page ...
Страница 224: ...Appendix A 3 Overview SITRANS P500 with HART 224 Operating Instructions 12 2014 A5E02344528 09 Continued on next page ...
Страница 225: ...Appendix A 3 Overview SITRANS P500 with HART Operating Instructions 12 2014 A5E02344528 09 225 Continued on next page ...
Страница 226: ...Appendix A 3 Overview SITRANS P500 with HART 226 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 228: ...Appendix A 4 Technical support SITRANS P500 with HART 228 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 230: ...List of abbreviations SITRANS P500 with HART 230 Operating Instructions 12 2014 A5E02344528 09 ...
Страница 234: ...Glossary SITRANS P500 with HART 234 Operating Instructions 12 2014 A5E02344528 09 TP Total Performance ...
Страница 238: ......