83
PBD-51041388
SITRANS LG250 - Operating Instructions
46201-EN-150424
Materials
Ʋ
Housing
ABS
Ʋ
Inspection window
Polyester foil
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone Ex factory
CET
Electronics temperature measurement
Resolution
1 °C (1.8 °F)
Accuracy
±1 °C (1.8 °F)
Voltage supply
Operating voltage
Ʋ
Non-Ex instrument, Ex-d instrument
9.6 … 35 V DC
Ʋ
Ex ia instrument
9.6 … 30 V DC
Ʋ
Ex-d-ia instrument
15 … 35 V DC
Operating voltage with illuminated display and adjustment module
Ʋ
Non-Ex instrument, Ex-d instrument
16 … 35 V DC
Ʋ
Ex ia instrument
16 … 30 V DC
Ʋ
Ex-d-ia instrument
20 … 35 V DC
Reverse voltage protection
Integrated
Permissible residual ripple - Non-Ex, Ex-ia instrument
Ʋ
Frequency
16 … 400 Hz
Ʋ for 12 V< U
N
< 18 V
≤ 0.7 V
eff
Ʋ for 18 V< U
N
< 35 V
≤ 1.0 V
eff
Permissible residual ripple - Ex-d-ia instrument
Ʋ
Frequency
16 … 400 Hz
Ʋ for 18 V< U
N
< 35 V
≤ 1.0 V
eff
Load resistor
Ʋ
Calculation
(U
B
- U
min
)/0.0215 A
Ʋ
Example - Non-Ex instrument with
U
B
= 24 V DC
(24 V - 9.6 V)/0.0215 A = 670 Ω
Electrical protective measures
Protection rating
Housing material
Version
IP-protection class
NEMA protection
Plastic
Single chamber
IP 66/IP 67
NEMA 4X
Double chamber
IP 66/IP 67
NEMA 4X