14
PBD-51041388
SITRANS LG250 - Operating Instructions
46201-EN-150424
Fig. 4: Vessel with conical bottom
Plastic vessel/Glass vessel
The guided microwave principle requires a metallic surface on the
process fitting. Therefore, in plastic vessels, etc., use an instru
-
ment version with flange (from DN 50) or place a metal sheet
(ø > 200 mm/8 in) beneath the process fitting when screwing it in.
Make sure that the plate has direct contact with the process fitting.
When installing rod or cable probes in vessels without metal walls,
e.g. in plastic vessels, the measured value can be influenced by
strong electromagnetic fields (emitted interference according to
EN 61326: class A). In this case, use a probe with coaxial version.
1
2
Fig. 5: Installation in non-metallic vessel
1 Flange
2 Metal sheet
If possible, avoid sockets. Mount the sensor flush with the vessel top.
If this is not possible, use short sockets with small diameter.
Type of vessel
Socket