25
PBD-51041388
SITRANS LG250 - Operating Instructions
46201-EN-150424
Contact pin
Colour connection ca-
ble in the sensor
Terminal, electronics
module
Pin 1
Brown
5
Pin 2
White
6
Pin 3
Blue
7
Pin 4
Black
8
5.6 Wiring plan, Ex-d-ia double chamber housing
3
1
2
5 6 7 8
1
2
( )
(-)
+
4...20mA
Fig. 17: Electronics compartment, double chamber housing
1 Internal connection to the terminal compartment
2 For display and adjustment module or interface adapter
3 Internal connection to the plug connector for external display and adjust-
ment unit (optional)
4...20mA
1
2
+
( )
(-)
2
1
Fig. 18: Connection compartment with double chamber housing Ex d
1 Voltage supply/Signal output
2 Ground terminal for connection of the cable screen
Electronics compartment
Terminal compartment