7ML19985GE04
SITRANS LC500 – OPERATING INSTRUCTIONS
Page 59
Appendix A
The transmitter variable on which this selection is based is chosen in Menu 01. The
increment/decrement step size is set at Menu 09.
Lower Range Value
Lower Range Value (LRV) is the setting for 0% of the operating range, in most cases an
empty vessel/tank. If the display mode is in % this selection is disabled and the LCD
displays - - - -.
The transmitter variable on which this menu selection operates is chosen in Menu 01. For
TV0 the LRV is in most cases Factory Set to the probe capacitance in free air.
The increment/decrement step size is subject to the setting from Menu 09.
1. Set the rotary switch to E.
2. Press and hold both buttons simultaneously to take the current PV reading as
the new setting.
or: Press the RED (+) or BLUE (–) button to step the value up or down.
or: Press and hold a button for a prolonged time to start a repeat function.
When the new setting exceeds that of the Limit Settings (Menu 0C and 0B), the new value
is rejected and the previous value remains unchanged.
Menu
Item
Affected
by:
Rotary
Switch
Position
Left
Arrow
Mode
Description
Values
0D
01, 07, 09
D
Off
2-state
Delta Range Setting
(enabled)
Span
setting
Analog
*
Display
shows
- - - -
Menu
Item
Affected
by:
Rotary
Switch
Position
Left
Arrow
Mode Description Values
0E
01, 09, 0B, 0C E
Off
Analog
Lower Range
Value
Range: 0.00 to 3300
Factory setting
Probe capacitance in
air
Display percent Display shows
- - - -