SITRANS LC500 – OPERATING INSTRUCTIONS
Page 37
Int
e
rc
on
nect
ion
•
The Upper Sensor Limit (USL) and Lower Sensor Limit (LSL) are set to 3300 and
1.666 pF respectively, and the following conditions apply: the Upper Range Value
(URV) and Lower Range Value (LRV) should be within the USL to LSL range, but
can be set anywhere within that range.
•
An interruption of the measuring connection will be detected: a loose or
interrupted connection results in up to 0.5 pF capacitance, which is below the
adjusted LSL and thus signals a FAULT condition.
TV1 Units
UNDEFINED
TV1 LRV
0 [switch position (0)E, TV1]
TV! URV
1.0 [switch position (0)F, TV1]
Setting
Description