Select Pipe Configuration
1. Scroll down to [Pipe Config] and press the <Right Arrow>.
2. Select a configuration that approximates the conditions upstream of your sensor mounting
location. (Refer to the definitions below.)
3. Press <ENT> to save selection.
①
Use this menu cell to enter the number of pipe diameters between the upstream configuration
and the sensor installation.
②
Use this menu cell to select the pipe configuration that most accurately represents the
upstream pipe condition.
4. Press the <Left Arrow> and return to the main menu.
Table 5-1
Pipe Configuration Option List Definitions
Options
Definitions
Fully Developed
Fully developed flow, as would be expected for very long
straight pipe runs or installation downstream of a flow
condition.
1 Elbow
Single 90 degree Elbow upstream of sensor installation.
Dble Elbow+
Double out-of-plane Elbows upstream of sensor installation.
Dble Elbow-
Double in-plane Elbows upstream of sensor installation.
Valve
Not available at this time.
Expander
Pipe expansion upstream of sensor installation.
Reducer
Pipe reduction upstream of sensor installation.
Norm Entry
Not available at this time.
Header Inlet
Header or pipe manifold upstream of sensor installation.
Intrusions
Not available at this time.
Connecting
5.2 Sensor Wiring
FUP1010 IP67 Portable Flowmeter
36
Operating Instructions, 02/2010, A5E02951522A Revision 01
Содержание SITRANS FUP1010
Страница 2: ......
Страница 66: ......
Страница 136: ......
Страница 138: ......
Страница 150: ......
Страница 153: ...1010WP 7 21614 C ...
Страница 154: ...1010WP 7 21614 C ...
Страница 155: ...1010WDP 7 21614 C ...
Страница 156: ...1010WDP 7 21614 C ...
Страница 157: ...1010WP 8 21614 C ...
Страница 158: ...1010WDP 8 21614 C ...
Страница 159: ......
Страница 161: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 162: ...1011PPS 8 TRANSDUCER PART NUMBER L H W NET WT PAIR X Z 21614 C SIZES A B C D SIZE E ...
Страница 163: ...21614 C ...
Страница 164: ......
Страница 165: ......
Страница 166: ......
Страница 167: ......
Страница 168: ...21614 C ...
Страница 169: ...21614 C ...
Страница 170: ...21614 C ...
Страница 171: ...21614 C ...
Страница 172: ...21614 C ...
Страница 173: ...21614 C ...
Страница 174: ...21614 C ...
Страница 175: ...1015BC 1W 8 21614 C ...
Страница 176: ......
Страница 177: ...21614 C ...
Страница 178: ...21614 C ...
Страница 179: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Страница 180: ...REFLECT MODE DIRECT MODE 21614 C ...
Страница 181: ...21614 C ...
Страница 182: ...1015BC 1 8 21614 C ...
Страница 183: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Страница 184: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Страница 185: ......
Страница 186: ......
Страница 187: ......
Страница 188: ......
Страница 190: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Страница 191: ......
Страница 192: ......
Страница 193: ...L TEMPERATURE BELOW 225 F ONLY SUITABLE FOR AMBIENT TEMPERATURE GLAND SUPPLIED WITH CABLE IS 21614 C PART No L ...
Страница 194: ......
Страница 204: ......
Страница 206: ......
Страница 207: ......