FEC920: USER GUIDE
Page 59
September 18
4.
The display changes to the high value adjust page.
5.
Apply the known high value and wait for the value to stabilize. Enter the High Target Value (the value that
the instrument is to read for the applied input). When all is steady, set ‘Confirm High’ to ‘Yes’.
Figure 52 Channel adjustment procedure (3)
REMOVAL PROCEDURE
1.
Set ‘Remove Adjust’ to ‘Yes’ and operate the scroll button.
2.
Use the scroll and up/down arrow buttons to change the required channel icons from crosses to ticks.
3.
Select Remove IP Adjust to ‘Yes’ and operate the scroll key. The adjustment is removed from all selected
channels without further confirmation.
Figure 53 Channel adjustment removal
DUAL INPUT CHANNELS
For the dual input channel option, input adjust is carried out as described above, except that for any channel
where dual inputs are configured, the user must initiate adjustment to primary and secondary inputs separately.
As shown in Figure 54, a new field ‘Input on Channel’ is introduced for this purpose.
Figure 54 Input adjust top level display (dual input channels)
High Target Value
Confirm High
Channel 1 Value
Abort
Instrument.Input Adjust
10.00
Yes
9.79
No
Channel 1
Channel 2
Channel 3
Channel 4
Apply Adjust
Remove Adjust
Instrument.Input Adjust
Adjusted
Unadjusted
Adjusted
Unadjusted
Yes
No
Channel 1
Channel 2
Channel 3
Channel 4
Apply Adjust
Remove Adjust
Instrument.Input Adjust
Adjusted
Unadjusted
Unadjusted
Unadjusted
Yes
No
Channel 1
Channel 2
Channel 3
Channel 4
Remove IP Adjust
Abort
Instrument.Input Adjust
Yes
No
Input on Channel
Channel 1
Channel 2
Channel 3
Channel 4
Apply Adjust
Remove Adjust
Instrument.Input Adjust
Primary
Unadjusted
Unadjusted
Unadjusted
Unadjusted
No
No
Input on Channel
Channel 1
Channel 3
Apply Adjust
Remove Adjust
Instrument.Input Adjust
Secondary
Unadjusted
Unadjusted
No
No
Only those channels with ‘Type’ set to ‘Dual mA’,
‘Dual mV’ or ‘Dual T/C’ appear in the list of second-
ary channels. In this example, only channels 1 and
3 are configured as dual input. (See Section 6.4.1
for channel Type configuration.)