3.2.2.6.2 Upper warning limit 

Description 
If mass flow value exceeds this limit a warning is triggered. 

 

Setting  -3682800.00 kg/h to +3682800.00 kg/h 
Default  3682800.00 kg/h 

Related 

 

Path 

Setup → Process values → Mass flow → Alarm and warning limits → Upper warning limit

 

3.2.2.6.3 Lower warning limit 

Description 
If the mass flow value falls below this limit a warning is triggered. 

 

Setting  -3682800.00 kg/h to +3682800.00 kg/h 
Default  -3682800.00 kg/h 

Related 

 

Path 

Setup → Process values → Mass flow → Alarm and warning limits → Lower warning limit

 

3.2.2.6.4 Lower alarm limit 

Description 
If mass flow value falls below this limit an alarm is triggered. 

 

Setting  -3682800.00 kg/h to +3682800.00 kg/h 
Default  -3682800.00 kg/h 

Related 

 

Path 

Setup → Process values → Mass flow → Alarm and warning limits →

 Lower alarm limit 

3.2.2.6.5 Hysteresis 

Description 
Sets the hysteresis for alarm and warning limits. The hysteresis is the distance between the 

limits for activation and deactivation of an alarm/warning. 

 

Setting  0.0 kg/h to 3682800.00 kg/h 
Default  0.0 kg/h 

Содержание SITRANS F

Страница 1: ...Function Manual Answers for industry 04 2017 Edition Ultrasonic Flowmeters FST030 SITRANS F ...

Страница 2: ......

Страница 3: ...on Manual This Function Manual applies to Siemens products SITRANS FST030 with order codes commencing 7ME372 04 2017 A5E35693328 AA Introduction 1 Quick Start 2 Setup 3 Maintenance and diagnostics 4 Communication 5 Security 6 Language 7 Support A ...

Страница 4: ...ay be operated only by personnel qualified for the specific task in accordance with the relevant documentation in particular its warning notices and safety instructions Qualified personnel are those who based on their training and experience are capable of identifying risks and avoiding potential hazards when working with these products systems Proper use of Siemens products Note the following WAR...

Страница 5: ...1 Pipe class 23 3 1 2 2 Pipe size 23 3 1 2 3 Outer pipe diameter 23 3 1 2 4 Wall thickness 24 3 1 2 5 Pipe material 24 3 1 2 6 Wall sound velocity 24 3 1 2 7 Liner settings 25 3 1 2 8 Inner pipe roughness 25 3 1 2 9 Inner pipe diameter 26 3 1 2 10 Active compensations 26 3 1 2 11 Active compensations settings 26 3 1 3 Upstream and downstream conditions 26 3 1 3 1 Disturbed flow profile compensatio...

Страница 6: ...hs 33 3 1 6 2 Path 1 34 3 1 6 3 Path 2 36 3 1 6 4 Necessary number of measuring paths 37 3 1 6 5 Necessary number of measuring paths settings 37 3 1 7 User calibration 38 3 1 7 1 Slope 38 3 1 7 2 Path 1 offset 38 3 1 7 3 Path 2 offset 38 3 1 8 Multipoint calibration 39 3 1 8 1 Enable multipoint calibration 39 3 1 8 2 Asymmetric calibration table 39 3 1 8 3 Calibration point 39 3 1 8 4 Multipoint c...

Страница 7: ...aces 55 3 2 4 5 Low flow cut off 55 3 2 4 6 Alarm and warning limits 55 3 2 5 Sound velocity 57 3 2 5 1 Units 57 3 2 5 2 Custom units 57 3 2 5 3 Custom conversion factor 57 3 2 5 4 Decimal places 58 3 2 5 5 Alarm and warning limits 58 3 2 6 LiquIdent 60 3 2 6 1 Alarm and warning limits 60 3 2 6 2 Temperature slope factor 61 3 2 6 3 Pressure slope factor 61 3 2 7 Density 62 3 2 7 1 Units 62 3 2 7 2...

Страница 8: ...3 3 3 1 3 Custom units 83 3 3 1 4 Custom conversion factor 83 3 3 1 5 Decimal places 84 3 3 1 6 Direction 84 3 3 1 7 Fail safe behaviour 84 3 3 1 8 Reset 84 3 3 1 9 Preset 85 3 3 1 10 Alarm and warning limits 85 3 3 2 Totalizer 2 87 3 3 2 1 Process value 87 3 3 2 2 Units 87 3 3 2 3 Custom units 88 3 3 2 4 Custom conversion factor 88 3 3 2 5 Decimal places 88 3 3 2 6 Direction 89 3 3 2 7 Fail safe ...

Страница 9: ...ulse output 120 3 4 4 7 Status output 124 3 4 5 Channel 3 input output 131 3 4 5 1 Operation mode 131 3 4 5 2 Change operation mode 131 3 4 5 3 Active operation read only 132 3 4 5 4 Fail safe activation condition 132 3 4 5 5 Current output 132 3 4 5 6 Frequency output 136 3 4 5 7 Pulse output 139 3 4 5 8 Status output 143 3 4 5 9 Digital input 151 3 4 5 10 Current input 153 3 4 6 Channel 3 relay ...

Страница 10: ...ms 1 192 3 4 8 4 Sensor alarms 2 193 3 4 8 5 Process alarms 1 193 3 4 8 6 Process alarms 2 193 3 4 8 7 Process alarms 3 194 3 4 8 8 Process alarms 4 194 3 4 8 9 Totalizer alarms 195 3 4 8 10 Device alarms 195 3 4 8 11 Channel 1 alarms 195 3 4 8 12 Input output alarms 196 3 4 8 13 Simulation alarms 196 3 4 8 14 Alarm class 196 3 4 8 15 NAMUR status signal 197 3 4 8 16 Polarity 197 3 4 8 17 On delay...

Страница 11: ...ss value 211 3 6 6 6 4th process value 211 3 6 6 7 5th process value 211 3 6 6 8 6th process value 212 3 6 6 9 Trend scale mode 212 3 6 6 10 Trend log time window 212 3 6 6 11 Trend scale lower limit 213 3 6 6 12 Trend scale upper limit 213 3 6 7 View 3 214 3 6 7 1 Enable or disable 214 3 6 7 2 View 214 3 6 7 3 1st value 214 3 6 7 4 Graphic scale mode 215 3 6 7 5 2nd process value 215 3 6 7 6 3rd ...

Страница 12: ... 3 1st process value 228 3 6 10 4 2nd process value 228 3 6 10 5 3rd process value 229 3 6 10 6 4th process value 229 3 6 10 7 5th process value 229 3 6 10 8 6th process value 230 3 6 10 9 Trend scale mode 230 3 6 10 10 Trend log time window 231 3 6 10 11 Trend scale lower limit 231 3 6 10 12 Trend scale upper limit 231 3 6 11 Status icons 232 4 Maintenance and diagnostics 233 4 1 Identification 2...

Страница 13: ... diagnostic events 243 4 2 2 Diagnostic log 243 4 2 3 Clear diagnostic log 244 4 2 4 Acknowledge mode 244 4 2 5 Transmitter detailed events read only 244 4 2 6 Suppression time 245 4 2 7 Enable alarms 245 4 2 7 1 Sensor events 1 245 4 2 7 2 Sensor events 2 246 4 2 7 3 Process events 1 246 4 2 7 4 Process events 2 247 4 2 7 5 Process events 3 247 4 2 7 6 Process events 4 247 4 2 7 7 Totalizer event...

Страница 14: ...nnel 4 relay 273 4 4 3 7 Calibration picture 273 4 5 Peak values 274 4 5 1 Process value 1 274 4 5 1 1 Process value 274 4 5 1 2 Maximum 274 4 5 1 3 Timestamp at maximum value 274 4 5 1 4 Minimum 275 4 5 1 5 Timestamp at minimum value 275 4 5 1 6 Reset peak values 275 4 5 2 Process value 2 276 4 5 2 1 Process value 276 4 5 2 2 Maximum 276 4 5 2 3 Timestamp at maximum value 276 4 5 2 4 Minimum 277 ...

Страница 15: ...286 4 8 1 3 Channel 3 input output 287 4 8 1 4 Relay 3 289 4 8 1 5 Channel 4 Input output 290 4 8 1 6 Channel 4 relay 291 4 8 2 Simulation process values 292 4 8 2 1 Enable simulation 292 4 8 2 2 Volume flow 293 4 8 2 3 Mass flow 293 4 8 2 4 Standard volume flow 294 4 8 2 5 Flow velocity 294 4 8 2 6 Sound velocity 294 4 8 2 7 LiquIdent 295 4 8 2 8 Density 295 4 8 2 9 Kinematic viscosity 296 4 8 2 ...

Страница 16: ...10 5 2 4 HART device type read only 311 5 2 5 HART device revision 311 5 2 6 Response preambles 311 5 2 7 Dynamic variable mapping 311 5 2 7 1 SV process value 312 5 2 7 2 TV process value 313 5 2 7 3 QV process value 314 5 2 8 Hart units 315 5 2 8 1 Volume flow units 315 5 2 8 2 Mass flow units 315 5 2 8 3 Standard volume flow unit 315 5 2 8 4 Flow velocity units 316 5 2 8 5 Sound velocity units ...

Страница 17: ...y unit 325 5 3 8 5 Sound velocity unit 325 5 3 8 6 Density unit 326 5 3 8 7 Kinematic viscosity unit 326 5 3 8 8 Pressure unit 326 5 3 8 9 Temperature units 327 5 3 8 10 Totalizer 1 unit 327 5 3 8 11 Totalizer 2 unit 328 5 3 8 12 Totalizer 3 unit 328 5 3 8 13 Service Channel USB 329 6 Security 331 6 1 Change user PIN 331 6 2 Change expert PIN code expert 331 6 3 Recovery ID 331 6 4 PIN recovery 33...

Страница 18: ...Table of contents FST030 16 Function Manual 04 2017 A5E35693328 AA ...

Страница 19: ...chapter titles The default settings and the setting ranges are listed in tables for each parameter Furthermore the relation if any between parameters is described As of chapter 3 Setup Page 23 the chapter and subchapter numbers equal the parameter numbers in the HMI structure For parameters with access level other than User for example Read only or Expert the access level is stated in parenthesis ...

Страница 20: ... 00 or later EDD 1 00 00 or later EDD 1 00 00 or later EDD 1 00 00 or later Edition Remarks Product compatibil ity Compatibility of device integration package 04 2017 First edition Modbus FW 1 00 00 HW 02 SIMATIC PDM V8 2 or later AMS Device Manager 12 0 or later SITRANS DTM V4 1 Field communicator V3 8 EDD 1 00 00 or later EDD 1 00 00 or later EDD 1 00 00 or later EDD 1 00 00 or later 1 6 Documen...

Страница 21: ...strument tech nicians plant operators On documentation disk Available for download from homepage Hardcopy can be purchased via PIA Life Cycle Portal Sensor Installa tion Manual Contains all information needed to install Clamp on sensors Mechanical and electrical install ers instrument technicians Included in the delivery On documentation disk Available for download from homepage Compact Operat ing...

Страница 22: ...Introduction 1 7 Device documentation package FST030 20 Function Manual 04 2017 A5E35693328 AA ...

Страница 23: ...cation 2 2 Sensor settings Description The sensor settings wizard will guide you through configuration of essential parameters 2 3 Process values The Process values wizard will guide you through setup of process values for your application The prioritizing of the process values automatically configures the measurement views on the display The process value configured as 1st process value appears a...

Страница 24: ...Quick Start 2 5 Copy configuration FST030 22 Function Manual 04 2017 A5E35693328 AA ...

Страница 25: ...ription Pipe class Read only Can only be changed via the Sensor Wizard Indicates the pipe class chosen from the pipe data library included in the sensor setup wizard 3 1 2 2 Pipe size Description Pipe size Read only Can only be changed via the Sensor Wizard Indicates the pipe class chosen from the pipe data library included in the sensor setup wizard 3 1 2 3 Outer pipe diameter Description Externa...

Страница 26: ... Stainless steel ABS plastic Aluminum Black iron Brass Cast iron Copper nickel 70 30 Copper nickel 90 10 Copper Ductile iron FRP plastic Glass Hastelloy Inconel Kynar plastic Monel Nickel Polyethylene PVC plastic Teflon Titani um Default Can only be changed via the Sensor wizard Related Path Setup Sensor settings Pipe settings Pipe material 3 1 2 6 Wall sound velocity Description Pipe wall materia...

Страница 27: ...ly Can only be changed via the Sensor Wizard Setting Read only Custom Cement Coal tar Enamel Glass Plastic Hi_dens_Poly PTFE Rub ber Default Custom Related Path Setup Sensor settings Pipe settings Liner settings Liner material Liner sound velocity Description Liner material sound velocity clamp on Setting 200 0 m s to 4000 0 m s Default 2000 0 m s Related Path Setup Sensor settings Pipe settings L...

Страница 28: ...lt 0 0 m Related Path Setup Sensor settings Pipe settings Inner pipe diameter 3 1 2 10 Active compensations Description Active compensations 3 1 2 11 Active compensations settings Setting Pipe expansion due to temperature Pipe expansion due to pressure Default Multiselection Related Path Setup Sensor settings Pipe settings Active compensations 3 1 3 Upstream and downstream conditions 3 1 3 1 Distu...

Страница 29: ...ow Double elbow in plane Double elbow out of plane Reducer Expansion Default Straight run Related Path Setup Sensor settings Upstream and downstream conditions Type of upstream condition 3 1 3 4 Upstream distance Description The distance to the upstream flow disturbance The distance is given in pipe diameters Used for clamp on flow profile compensation 3 1 3 5 Upstream distance settings Setting 0 ...

Страница 30: ...ipe diameters Used for clamp on flow profile compensation 3 1 3 8 Downstream distance settings Setting 0 to 40 Default 0 Related Path Setup Sensor settings Upstream and downstream conditions Downstream distance 3 1 4 Medium characteristics 3 1 4 1 Process medium type Description Process medium types Read only Can only be changed via the Sensor Wizard Indicates the fluid medium type chosen from the...

Страница 31: ...p Sensor settings Medium characteristics Process temperature 3 1 4 4 Process pressure Description Pressure value used for compensation Input during setup Setting 0 0 Pa to 100000000 0 Pa Default 700000 0 Pa Related Path Setup Sensor settings Medium characteristics Process pressure 3 1 4 5 Kinematic viscosity Description Viscosity of media used for compensation Input during setup Setting 0 00 m s t...

Страница 32: ... be the same Read only Can only be changed via the Sensor Wizard Indicates the Sensor model chosen from the sensor setup wizard Setting Read only FSS200 high precision FSS200 1011 Universal FSS200 991 high tempera ture 1011 high precision liquid Default FSS200 1011 Universal Related Path Setup Sensor settings Sensor selection Sensor model 3 1 5 2 Clamp on sensor size Description Installed sensor s...

Страница 33: ... m Related Path Setup Sensor settings Sensor settings Sensor crystal projection 3 1 5 5 Sensor phase velocity Description The velocity at which the wave fronts traverse the interface between the wedge and pipe Access level Expert user only Setting 1000 0 m s to 20000 0 m s Default 3695 5 m s Related Path Setup Sensor settings Sensor settings Sensor phase velocity 3 1 5 6 Sensor inactive wedge Desc...

Страница 34: ...g hole offset Description Distance between spacer hole and sensor front face Setting 0 0 m to 0 05 m Default 0 00953 m Related Path Setup Sensor settings Sensor settings Mounting hole offset 3 1 5 9 Spacing offset Description Sensor spacing offset Setting MIN NOM MAX and A B C D for High Temperature sensors Default MIN Related Path Setup Sensor settings Sensor selection Spacing offset 3 1 5 10 Tem...

Страница 35: ...sor selection Temperature compensation factor 3 1 5 12 Length of sensor cables Description Length of sensor cable between transmitter DSL and any sensors All cables must have the same length Setting 0 0 m to 20 0 m Default 0 0 m Related Path Setup Sensor settings Sensor selection Length of sensor cables 3 1 6 Path settings 3 1 6 1 Installed paths Description Bit encoded value containing installed ...

Страница 36: ...r path 1 Read only Setting 80000 0 Hz to 4000000 0 Hz Default Related Path Setup Sensor settings Path settings Path 1 Sensor frequency 3 1 6 2 2 Path geometry Description The number of crossings the signal makes through the pipe Reflect or Direct geometry for Path 1 For example Direct 1 traverse Reflect 2 traverse Direct 3 traverse Reflect 4 traverse Setting 1 to 10 Default 1 Related Path Setup Se...

Страница 37: ...oint adjustment Description Zero point adjustment Can only be changed via the Sensor Wizard Note Important Please stop all flow in the pipe and perform this function The flow velocity is measured prior and after the function has been applied Setting 5 to 300 Default 10 Related Path Setup Sensor settings Path settings Path 1 Zero point adjustment Time duration 3 1 6 2 5 Serial number Description Se...

Страница 38: ...uency 3 1 6 3 2 Path geometry Description The number of crossings the signal makes through the pipe Reflect or direct geometry for path 2 E g Z path 1 V path 2 W path 4 Setting 1 to 10 Default 1 Related Path Setup Sensor settings Path settings Path 2 Path geometry 3 1 6 3 3 Sensor spacing Description Sensor spacing Distance between the front faces of the sensors for path 2 Setting 100 0 m to 100 0...

Страница 39: ...Setup Sensor settings Path settings Path 2 Zero point adjustment Time duration 2 1 6 3 5 Serial number Description Serial number of sensor clamp on Setting Default Related Path Setup Sensor settings Path settings Path 2 Serial number 3 1 6 4 Necessary number of measuring paths Description List of measuring paths If less than the specified number of paths are in a measuring state the meter will rep...

Страница 40: ...r calibration Slope 3 1 7 2 Path 1 offset Description User calibrated zero offset for path 1 Delta time offset correction value which can be determined at zero flow Setting 1250ns to 1250ns Default 0 0 Related Path Setup Sensor settings User calibration Path 1 offset 3 1 7 3 Path 2 offset Description User calibrated zero offset for path 2 Delta time offset correction value which can be determined ...

Страница 41: ...alibrationTable 0 AND PID_EnableUserCalibrationTable 1 Usage of the calibration table Unidirectional All 20 calibration points are used for positive flow calibration and are then mirrored for negative flow Bidirectional The first 10 calibration points are used for positive flow calibration and the 10 last calibration points are used for negative flow calibration Setting Unidirectional Bidirectiona...

Страница 42: ...ess level Expert user only Note Only available for PID_ProtectCalibrationTable 1 OR PID_EnableUserCalibrationTable 1 Setting 0 0 m3 s to 100000 0 m3 s Default 0 0 m3 s Related Path Setup Sensor settings Multipoint calibration Multipoint calibration table Calibration points 1 20 3 1 8 4 2 Calibration value 1 20 Description Calibration correction factor that is associated to Calibration points 1 thr...

Страница 43: ...direction 3 1 10 Miscellaneous 3 1 10 1 Preamplifier Setting Description Preamplifier settings Access level Expert user only Auto default Will automatically turn on the pre amplifier if the sensor type is clamp on Off Turns off the pre amplifier regardless of the sensor type inline or clamp on On Turns on the pre amplifier regardless of the sensor type inline or clamp on Half transmit amplitude Re...

Страница 44: ... ringdown delay setting for optimized signal processing is 2000 000122 μs 3 1 10 4 Electronic time delay Description The time delay induced by the DSL electronics Access level Expert user only Read only The time delay induced by the DSL electronincs is 0 519 μs 3 1 11 Configuration of test blocks 3 1 11 1 About Description NOTICE Configuration data overwrite error Running the test block configurat...

Страница 45: ...ize Sensor path A1 Path 1 Path 2 A2 Path 1 Path 2 B1 Path 1 Path 2 B2 Path 1 Path 2 B3 Path 1 Path 2 C1 Path 1 Path 2 C2 Path 1 Path 2 C3 Path 1 Path 2 D1 Path 1 Path 2 D2 Path 1 Path 2 D3 Path 1 Path 2 Save settings After settings are complete press the right arrow to save sensor test block settings 3 2 Process values 3 2 1 Volume flow 3 2 1 1 Units Description Select units for displayed volume f...

Страница 46: ...L42 d million barrels per day BBL31 5 s 1 barrel 31 5 US gallons BBL31 5 min 1 barrel 31 5 US gallons BBL31 5 h 1 barrel 31 5 US gallons BBL31 5 d 1 barrel 31 5 US gallons ft s cubic feet per second ft min cubic feet per minute ft h cubic feet per hour ft d cubic feet per day Mft3 d million cubic feet per day af min acre feet per minute af h acre feet per hour af d acre feet per day in3 sec cubic ...

Страница 47: ...p Process values Standard volume flow Decimal places 3 2 1 5 Low flow cut off Description Volume flow limit for low flow cut off Below limit volume flow output is forced to zero Setting 0 00 m3 h to 896 40 m3 h depends on sensor Default 0 1 of maximum flow depends on sensor Related Path Setup Process values Volume flow Low flow cut off 3 2 1 6 Alarm and warning limits 3 2 1 6 1 Upper alarm limit D...

Страница 48: ...40 m3 h Default 896 40 m3 h Related Path Setup Process values Volume flow Alarm and warning limits Lower warning limit 3 2 1 6 4 Lower alarm limit Description Set Lower alarm limit Alarm is triggered if measured value is below the limit Setting 896 40 m3 h to 896 40 m3 h Default 896 40m3 h Related Path Setup Process values Volume flow Alarm and warning limits Lower alarm limit 3 2 1 6 5 Hysteresis...

Страница 49: ...ay t min 1 t 1000 kg t h 1 t 1000 kg t d 1 t 1000 kg STon min 1 STon 2000 lb STon h 1 STon 2000 lb STon d 1 STon 2000 lb T h 1 T 2240 lb T d 1 T 2240 lb lb s pounds per second lb min pounds per minute lb h pounds per hour lb d pounds per day custom units see Custom units Custom conversion factor Default kg h kilograms per hour Related Path Setup Process values Mass flow Units 3 2 2 2 Custom units ...

Страница 50: ...Path Setup Process values Mass flow Decimal places 3 2 2 5 Low flow cut off Description Mass flow limit for low flow cut off Below limit mass flow output is forced to zero Setting 0 00 to 3682800 00 kg h depends on sensor Default 0 1 of maximum mass flow depends on sensor Related Path Setup Process values Mass flow Low flow cut off 3 2 2 6 Alarm and warning limits 3 2 2 6 1 Upper alarm limit Descr...

Страница 51: ...tting 3682800 00 kg h to 3682800 00 kg h Default 3682800 00 kg h Related Path Setup Process values Mass flow Alarm and warning limits Lower warning limit 3 2 2 6 4 Lower alarm limit Description If mass flow value falls below this limit an alarm is triggered Setting 3682800 00 kg h to 3682800 00 kg h Default 3682800 00 kg h Related Path Setup Process values Mass flow Alarm and warning limits Lower ...

Страница 52: ...ters per day Sft s standard cubic feet per sec Sft min standard cubic feet per min Sft h standard cubic feet per hour Sft d standard cubic feet per day Sin3 sec cubic inches per second Sin3 min cubic inches per min Sin3 h cubic inches per hour Sin3 d cubic inches per day Custom units see Custom units Custom conversion factor Default Sm s Related Path Setup Process values Standard volume flow Units...

Страница 53: ...mal places 5 decimal places 6 decimal places Default 2 decimal places Related Path Setup Process values Standard volume flow Decimal places 3 2 3 5 Low flow cut off Description Standard volume flow limit for low flow cut off Below limit standard volume flow output is forced to zero Setting Min 0 0 Sm3 s Default 0 0 Sm3 s Related Path Setup Process values Standard volume flow Low flow cut off 3 2 3...

Страница 54: ...ard volume flow Alarm and warning limits Upper warning limit 3 2 3 6 3 Lower warning limit Description Falling below this limit causes a warning Setting 10800000 00 Sm3 h to 10800000 00 Sm3 h Default 3600 00 Sm3 h Related Path Setup Process values Standard volume flow Alarm and warning limits Lower warning limit 3 2 3 6 4 Lower alarm limit Description Falling below this limit causes an alarm Setti...

Страница 55: ... the standard process values Setting 0 0 C 32 F to 100 0 C 212 0 F Default 15 0 C 59 0 F Related Path Setup Process values Standard volume flow Standard temperature Units 3 2 3 8 Standard pressure Description Standard pressure This value is used to calculate standard process values Setting 100000 0 Pa to 200000 0 Pa Default 101325 0 Pa pascal Related Path Setup Process values Standard volume flow ...

Страница 56: ...r second in min inches per minute ft s feet per second ft min feet per minute custom units see Custom units Custom conversion factor Default m s meters per second Related Path Setup Process values Flow velocity Units 3 2 4 2 Custom units Description User specific text string to define custom unit values The custom unit can then be selected in the flow velocity units Setting Default Related Path Se...

Страница 57: ...ult 2 decimal places Related Path Setup Process values Flow velocity Decimal places 3 2 4 5 Low flow cut off Description Flow velocity limit for low flow cut off Below limit flow velocity output is forced to zero Setting Min 0 00 Default 0 0 m s Related Path Setup Process values Flow velocity Low flow cut off 3 2 4 6 Alarm and warning limits 3 2 4 6 1 Upper alarm limit Description If process value...

Страница 58: ...g 9999997952 m s to 9999997952 m s Default 9999997952 m s Related Path Setup Process values Flow velocity Alarm and warning limits Lower warning limit 3 2 4 6 4 Lower alarm limit Description If the process value falls below this limit an alarm is triggered Setting 9999997952 m s to 9999997952 m s Default 9999997952 m s Related Path Setup Process values Flow velocity Alarm and warning limits Lower ...

Страница 59: ...r minute ft s feet per second ft min feet per minute Custom units see Custom units Custom conversion factor Default m s meters per second Related Path Setup Process values Sound velocity Units 3 2 5 2 Custom units Description User specific text string to define a name for custom unit values The custom unit can then be selected in the sound velocity units Setting Default Related Path Setup Process ...

Страница 60: ...ces Related Path Setup Process values Sound velocity Decimal places 3 2 5 5 Alarm and warning limits 3 2 5 5 1 Upper alarm limit Description If process value exceeds this limit an alarm is triggered Setting 9999997952 m s to 9999997952 m s Default 9999997952 m s Related Path Setup Process values Sound velocity Alarm and warning limits Upper alarm limit 3 2 5 5 2 Upper warning limit Description If ...

Страница 61: ...arm limit Description If the process value falls below this limit an alarm is triggered Setting 9999997952 m s to 9999997952 m s Default 9999997952 m s Related Path Setup Process values Sound velocity Alarm and warning limits Lower alarm limit 3 2 5 5 5 Hysteresis Description Sets the hysteresis for alarm and warning limits The hysteresis is the distance between the limits for activation and deact...

Страница 62: ...larm and warning limits Upper alarm limit 3 2 6 1 2 Upper warning limit Description Exceeding this limit causes a warning Setting 9999997952 m s to 9999997952 m s Default 9999997952 m s Related Path Setup Process values LiquIdent Alarm and warning limits Upper warning limit 3 2 6 1 3 Lower warning limit Description Falling below this limit causes a warning Setting 9999997952 m s to 9999997952 m s ...

Страница 63: ...9997952 m s Default 0 0 m s Related Path Setup Process values LiquIdent Alarm and warning limits Hysteresis 3 2 6 2 Temperature slope factor Description Temperature slope factor This factor is used in connection with the standard condition temperature for calculating LiquIdent Setting 0 0 m s K to 8 0 m s K Default 4 1 m s K Related Path Setup Process values LiquIdent Temperature slope factor 3 2 ...

Страница 64: ...lograms per liter kg m kilograms per cubic meter lb in pounds per cubic inch lb gal pounds per US gallon lb ft pounds per cubic foot STon yd 1 STon 2000 lb custom units see Custom units Custom conversion factor Default kg m kilograms per cubic meter Related Path Setup Process values Density Units 3 2 7 2 Custom units Description User specific text string to define custom unit values The custom uni...

Страница 65: ...lt 2 decimal places Related Path Setup Process values Density Decimal places 3 2 7 5 Alarm and warning limits 3 2 7 5 1 Upper alarm limit Description If density value exceeds this limit an alarm is triggered Setting 0 00 kg m to 10000 00 kg m Default 900 00 kg m Related Path Setup Process values Density Alarm and warning limits Upper alarm limit 3 2 7 5 2 Upper warning limit Description If density...

Страница 66: ...alls below this limit an alarm is triggered Setting 0 00 kg m to 10000 00 kg m Default 1 00 kg m Related Path Setup Process values Density Alarm and warning limits Lower alarm limit 3 2 7 5 5 Hysteresis Description Sets the hysteresis for alarm and warning limits The hysteresis is the distance between the limits for activation and deactivation of an alarm warning Setting 0 00 kg m to 10000 00 kg m...

Страница 67: ...places 3 decimal places 4 decimal places 5 decimal places 6 decimal places Default 2 decimal places Related Path Setup Process values Kinematic viscosity Decimal places 3 2 8 3 Alarm and warning limits 3 2 8 3 1 Upper alarm limit Description If process value exceeds this limit an alarm is triggered Setting 9999997952 m s to 9999997952 m s Default 9999997952 m s Related Path Setup Process values St...

Страница 68: ...limits Lower warning limit 3 2 8 3 4 Lower alarm limit Description If process value falls below this limit an alarm is triggered Setting 9999997952 m s to 9999997952 m s Default 9999997952 m s Related Path Setup Process values Kinematic viscosity Alarm and warning limits Lower alarm limit 3 2 8 3 5 Hysteresis Description Sets the hysteresis for alarm and warning limits The hysteresis is the distan...

Страница 69: ... C mH2O meters of water at 4 C cmHg centimeters of mercury at 0 C mHg meters of mercury at 0 C gf cm grams force per square cm kg m kil ograms per square meter lb ft pounds per square foot ftH2O feet water at 4 C ftH2O feet water at 60 F Default Pa pascals Related Path Setup Process values Pressure Units 3 2 9 2 Decimal places Description Sets the decimal places for the process value displayed in ...

Страница 70: ...ressure Alarm and warning limits Upper warning limit 3 2 9 3 3 Lower warning limit Description If the process value falls below this limit a warning is triggered Setting 9999997952 Pa to 9999997952 Pa Default 9999997952 Pa Related Path Setup Process values Pressure Alarm and warning limits Lower warning limit 3 2 9 3 4 Lower alarm limit Description if process value falls below this limit an alarm ...

Страница 71: ...3 2 10 1 Units Description Select units for displayed temperature Units are not applied at communication interface Setting C degrees Celsius F degrees Fahrenheit R degrees Rankine K kelvins Default C degrees Celsius Related Path Setup Process values Medium temperature Units 3 2 10 2 Decimal places Description Sets the decimal places for the process value displayed in the local display Setting Auto...

Страница 72: ...warning limits Upper alarm limit 3 2 10 3 2 Upper warning limit Description Exceeding this limit causes a warning Related Setting 272 77 to 400 00 C Default 270 00 C Related Path Setup Process values Medium temperature Alarm and warning limits Upper warn ing limit 3 2 10 3 3 Lower warning limit Description Falling below this limit causes a warning Setting 272 77 00 to 400 00 C Default 270 00 C Rel...

Страница 73: ... 0 00 C to 200 00 C Default 0 00 C Related Path Setup Process values Medium temperature Alarm and warning limits Hysteresis 3 2 11 Standard density 3 2 11 1 Units Description Select units for displayed standard density Units are not applied at communication interface Related Setting μg l micrograms per liter μg m micrograms per cubic meter mg l milligrams per liter g ml grams per milliliter g cm g...

Страница 74: ...ath Setup Process values Standard density Decimal places 3 2 11 3 Alarm and warning limits 3 2 11 3 1 Upper alarm limit Description Exceeding this limit causes an alarm Setting 0 00 kg m to 10000 00 kg m Default 900 00 kg m Related Path Setup Process values Standard density Alarm and warning limits Upper alarm limit 3 2 11 3 2 Upper warning limit Description Exceeding this limit causes a warning S...

Страница 75: ...limits Lower warning limit 3 2 11 3 4 Lower alarm limit Description Falling below this limit causes an alarm Setting 0 00 kg m to 10000 00 kg m Default 1 00 kg m Related Path Setup Process values Standard density Alarm and warning limits Lower alarm limit 3 2 11 3 5 Hysteresis Description Hysteresis that affects the alarm and warning limits Setting 0 00 kg m to 10000 00 kg m Default 0 00 kg m Rela...

Страница 76: ... Units 3 2 12 2 Decimal places Description Sets the decimal places for the process value displayed in the local display Setting Auto No decimal place 1 decimal place 2 decimal places 3 decimal places 4 decimal places 5 decimal places 6 decimal places Default 2 decimal places Related Path Setup Process values Standard kinematic viscosity Decimal places 3 2 12 3 Alarm and Warning limits 3 2 12 3 1 U...

Страница 77: ...ng Setting 9999997952 m s to 9999997952 m s Default 9999997952 m s Related Path Setup Process values Standard kinematic viscosity Alarm and warning limits Lower warning limit 3 2 12 3 4 Lower alarm limit Description Falling below this limit causes an alarm Setting 9999997952 m s to 9999997952 m s Default 9999997952 m s Related Path Setup Process values Standard kinematic viscosity Alarm and warnin...

Страница 78: ...ption Exceeding this limit causes an alarm Setting 9999997952 API to 9999997952 API Default 9999997952 API Related Path Setup Process values API gravity Alarm and warning limits Upper alarm limit 3 2 13 1 2 Upper warning limit Description Exceeding this limit causes a warning Setting 9999997952 API to 9999997952 API Default 9999997952 API Related Path Setup Process values API gravity Alarm and war...

Страница 79: ...Setting 9999997952 API to 9999997952 API Default 9999997952 API Related Path Setup Process values API gravity Alarm and warning limits Lower alarm limit 3 2 13 1 5 Hysteresis Description Hysteresis that affects the alarm and warning limits Setting 0 00 API to 9999997952 API Default 0 00 API Related Path Setup Process values API gravity Alarm and warning limits Hysteresis 3 2 14 Standard API gravit...

Страница 80: ...I to 9999997952 API Default 9999997952 API Related Path Setup Process values Standard API gravity Alarm and warning limits Upper warn ing limit 3 2 14 1 3 Lower warning limit Description Falling below this limit causes a warning Setting 9999997952 API to 9999997952 API Default 9999997952 API Related Path Setup Process values Standard API gravity Alarm and warning limits Lower warn ing limit 3 2 14...

Страница 81: ...Path Setup Process values Standard API gravity Alarm and warning limits Hysteresis 3 2 15 Specific gravity 3 2 15 1 Alarm and warning limits 3 2 15 1 1 Upper alarm limit Description Exceeding this limit causes an alarm Setting 9999997952 SGU to 9999997952 SGU Default 9999997952 SGU Related Path Setup Process values Specific gravity Alarm and warning limits Upper alarm limit 3 2 15 1 2 Upper warnin...

Страница 82: ...rocess values Specific gravity Alarm and warning limits Lower warning limit 3 2 15 1 4 Lower alarm limit Description Falling below this limit causes an alarm Setting 9999997952 SGU to 9999997952 SGU Default 9999997952 SGU Related Path Setup Process values Specific gravity Alarm and warning limits Lower alarm limit 3 2 15 1 5 Hysteresis Description Hysteresis that affects the alarm and warning limi...

Страница 83: ...larm and warning limits Upper alarm limit 3 2 16 1 2 Upper warning limit Description Exceeding this limit causes a warning Setting 9999997952 SGU to 9999997952 SGU Default 9999997952 SGU Related Path Setup Process values Standard specific gravity Alarm and warning limits Upper warning limit 3 2 16 1 3 Lower warning limit Description Falling below this limit causes a warning Setting 9999997952 SGU ...

Страница 84: ... limit 3 2 16 1 5 Hysteresis Description Hysteresis that affects the alarm and warning limits Setting 0 00 SGU to 9999997952 SGU Default 0 00 SGU Related Path Setup Process values Standard specific gravity Alarm and warning limits Hystere sis 3 3 Totalizers 3 3 1 Totalizer 1 3 3 1 1 Process value Description Select the process value to be totalized Setting Mass flow Volume flow Standard volume flo...

Страница 85: ...BL42 1 barrel 42 US gallons in cubic inches ft cubic feet yd cubic yards bush bushels Process value Standard volume flow Nl normal liters Nm normal cubic meters Sl standard liters Sft standard cubic feet Sm standard cubic meters custom units see Custom units Custom conversion factor Default kg kilograms Related Path Setup Totalizers Totalizer 1 Units 3 3 1 3 Custom units Description User specific ...

Страница 86: ...tion Totalizer 1 direction Setting Forward and backward Forward Backward Hold Default Forward Related Path Setup Totalizers Totalizer 1 Direction 3 3 1 7 Fail safe behaviour Description Behaviour of the totalizer during the occurrence of bad input value Setting RUN totalization is continued using the bad input value HOLD totalization is stopped MEMORY totalization is continued based on the last in...

Страница 87: ...low 9999997952 Sm to 9999997952 Sm Setting Cancel OK Default Cancel Related Path Setup Totalizers Totalizer 1 Preset 3 3 1 10 Alarm and warning limits 3 3 1 10 1 Upper alarm limit Description Alarm is triggered if Totalizer 1 value is above the limit Setting Process value Mass flow 9999997952 kg to 9999997952 kg Process value Volume flow 9999997952 m to 9999997952 m Process value Standard volume f...

Страница 88: ...cess value Mass flow 9999997952 kg to 9999997952 kg Process value Volume flow 9999997952 m to 9999997952 m Process value Standard volume flow 9999997952 Sm to 9999997952 Sm Default Process value Mass flow 9999997952 kg Process value Volume flow 9999997952 m Process value Standard volume flow 9999997952 Sm Related Path Setup Totalizers Totalizer 1 Alarm and warning limits Lower warning limit 3 3 1 ...

Страница 89: ...Default Process value Mass flow 0 00 kg Process value Volume flow 0 00 m Process value Standard volume flow 0 00 Sm Related Path Setup Totalizers Totalizer 1 Alarm and warning limits Hysteresis 3 3 2 Totalizer 2 3 3 2 1 Process value Description Select the process value to be totalized Setting Mass flow Volume flow Standard volume flow only Hydrocarbon and gas applications Default Mass flow Relate...

Страница 90: ...l cubic meters Sl standard liters Sft standard cubic feet Sm standard cubic meters custom units see Custom units Custom conversion factor Default kg kilograms Related Path Setup Totalizers Totalizer 2 Units 3 3 2 3 Custom units Description User specific string to define custom unit Totalizer 2 values Setting Default Related Path Setup Totalizers Totalizer 2 Custom units 3 3 2 4 Custom conversion f...

Страница 91: ...ult Forward Related Path Setup Totalizers Totalizer 2 Direction 3 3 2 7 Fail safe behaviour Description Behavior of the totalizer during the occurrence of bad input value Setting RUN totalization is continued using the bad input value HOLD totalization is stopped MEMORY totalization is continued based on the last incoming good value Default RUN totalization is continued using the bad input value R...

Страница 92: ...Path Setup Totalizers Totalizer 2 Preset 3 3 2 10 Alarm and warning limits 3 3 2 10 1 Upper alarm limit Description Alarm is triggered if Totalizer 2 value is above the limit Setting Process value Mass flow 9999997952 kg to 9999997952 kg Process value Volume flow 9999997952 m to 9999997952 m Process value Standard volume flow 9999997952 Sm to 9999997952 Sm Default Process value Mass flow 999999795...

Страница 93: ...Process value Volume flow 9999997952 m to 9999997952 m Process value Standard volume flow 9999997952 Sm to 9999997952 Sm Default Process value Mass flow 9999997952 kg Process value Volume flow 9999997952 m Process value Standard volume flow 9999997952 Sm Related Path Setup Totalizers Totalizer 2 Alarm and warning limits Lower warning limit 3 3 2 10 4 Lower alarm limit Description Alarm is triggere...

Страница 94: ...Process value Mass flow 0 00 kg Process value Volume flow 0 00 m Process value Standard volume flow 0 00 Sm Related Path Setup Totalizers Totalizer 2 Alarm and warning limits Hysteresis 3 3 3 Totalizer 3 3 3 3 1 Process value Description Select the cut off filtered process value to be totalized Setting Mass flow Volume flow Standard volume flow only Hydrocarbon and gas applications Default Mass fl...

Страница 95: ...l cubic meters Sl standard liters Sft standard cubic feet Sm standard cubic meters custom units see Custom units Custom conversion factor Default kg kilograms Related Path Setup Totalizers Totalizer 3 Units 3 3 3 3 Custom units Description User specific string to define custom unit Totalizer 3 values Setting Default Related Path Setup Totalizers Totalizer 3 Custom units 3 3 3 4 Custom conversion f...

Страница 96: ...efault Forward Related Path Setup Totalizers Totalizer 3 Direction 3 3 3 7 Fail safe behaviour Description Behavior of the totalizer during the occurrence of bad input value Setting RUN totalization is continued using the bad input value HOLD totalization is stopped MEMORY totalization is continued based on the last incoming good value Default RUN totalization is continued using the bad input valu...

Страница 97: ...Path Setup Totalizers Totalizer 3 Preset 3 3 3 10 Alarm and warning limits 3 3 3 10 1 Upper alarm limit Description Alarm is triggered if Totalizer 3 value is above the limit Setting Process value Mass flow 9999997952 kg to 9999997952 kg Process value Volume flow 9999997952 m to 9999997952 m Process value Standard volume flow 9999997952 Sm to 9999997952 Sm Default Process value Mass flow 999999795...

Страница 98: ...Process value Volume flow 9999997952 m to 9999997952 m Process value Standard volume flow 9999997952 Sm to 9999997952 Sm Default Process value Mass flow 9999997952 kg Process value Volume flow 9999997952 m Process value Standard volume flow 9999997952 Sm Related Path Setup Totalizers Totalizer 3 Alarm and warning limits Lower warning limit 3 3 3 10 4 Lower alarm limit Description Alarm is triggere...

Страница 99: ...ess value Volume flow 0 00 m Process value Standard volume flow 0 00 Sm Related Path Setup Totalizers Totalizer 3 Alarm and warning limits Hysteresis 3 3 4 Reset all totalizers Description Reset command to reset totalizers 1 to 3 to zero Setting Cancel OK Default Cancel Related Path Setup Totalizers Reset all totalizers 3 4 Inputs and outputs The available configuration of the hardware functionali...

Страница 100: ...requency ouput Pulse output Status output Alarm class Alarm item NAMUR status signals 3 Signal input output Inputs Current input 0 4 20 mA Digital input Totalizer reset Remote zero adjust Force outputs Freeze process values Outputs Current output 0 4 20 mA Frequency ouput Pulse output Status output Alarm class Alarm item NAMUR status signals Relay output Status Output Alarm class Alarm item NAMUR ...

Страница 101: ...put Status output Alarm class Alarm item NAMUR status signals Relay output Status Output Alarm class Alarm item NAMUR status signals Current output All four channels can be configured as current output Current output configuration The following process values can be assigned to the current output Volume flow Mass flow Sound velocity Flow velocity Density Medium temperature Pressure Kinematic visco...

Страница 102: ...y specific flow values Linear control range Lower saturation limit Upper saturation limit Lower fault current value Recommended setting range for lower fault current Recommended setting range for upper fault current Measuring range Figure 3 1 Current limits for NAMUR configuration The fail safe current output signal can be selected to Lower fault current defined in the Loop current scale selection...

Страница 103: ...ities for a current output Positive flow with negative scaling Low flow cut off Upper range value Maximum output current Upper fault current Lower range value Lower saturation limit Span Current output setting Process value Mass flow Direction Positive Loop current scale 4 20 mA maximum 25 mA Upper range value 100 kg h Lower range value 400 kg h Fail safe behaviour Upper fault current Low flow cut...

Страница 104: ...ge value Low flow cut off Upper range value Upper saturation limit Lower saturation limit Lower fault current Span Current output setting Process value Mass flow Direction Bidirectional Loop current scale 4 20 mA NAMUR Upper range value 400 kg h Lower range value 100 kg h Fail safe behaviour Upper fault current Low flow cut off 25 kg h ...

Страница 105: ...er range value Low flow cut off Upper range value Upper saturation limit Lower saturation limit Lower fault current Span Current output setting Process value Mass flow Direction Bidirectional Loop current scale 4 20 mA US Upper range value 400 kg h Lower range value 100 kg h Fail safe behaviour Minimum current Low flow cut off 25 kg h ...

Страница 106: ...s value Mass flow Direction Bidirectional Symmetric Loop current scale 4 20 mA NAMUR Upper range value 400 kg h Lower range value 100 kg h Fail safe behaviour Upper fault current Low flow cut off 25 kg h Pulse output The pulse output function supplies pulses equivalent to a configured amount of accumulated volume or mass The pulse width is configured and the pulse repetition is proportional to the...

Страница 107: ... Pulse width must be selected with the view that remaining time is always greater than pulse width at the highest measured flow Frequency output The frequency output function supplies a frequency 50 duty cycle proportional to the selected process value Frequency is calculated as follows Example This example shows how to calculate the output frequency for any measured flowrate Frequency output conf...

Страница 108: ...arms can be signaled on the output and selected from the alarm class or the alarm item lists Alarm class Alarm will be signaled if alarm within the selected alarm class occurs Alarm item Alarm will be signaled if selected alarm item occurs It is possible to select multiple alarms to be signaled Note Alarm class NAMUR status signals The options depend on the setting selected in Staus icons either N...

Страница 109: ...nnel 2 shifted by 90 or 180 of the functional width of the pulse If set to redundancy mode channel 3 inherits all channel 2 settings The functional width of the pulse is two times the pulse On duration The flow direction will determine whether channel 3 is shifted before or after channel 2 The following examples describe the pulse functionalities for channel 2 and 3 in redundancy mode Channel 2 co...

Страница 110: ...Off Disables the diagnosis of the current input channel Pressure Allocates the input current to the Process value Pressure Medium temperature Allocates the input current to the Process value Medium temperature Density Allocates the input current to the Process value Density Kinematic viscosity Allocates the input current to the Process value Kinematic viscosity Sensor temperature Allocates the inp...

Страница 111: ...configuration Reset all totalizers simultaneously Freeze process values Freezes all currently measured values in the display and outputs Force process values Forces all outputs to adopt the value selected in the menu If the value 100 is selected the current output will show 20 mA and the frequency output will show 10 000 kHz when the external output is activated Note Changing polarity Changing the...

Страница 112: ...cosity Pressure Medium temperature Standard density only hydrocarbon and gas applications Standard kinematic viscosity only hydrocarbon applications API gravity only hydrocarbon applications Standard API gravity only hydrocarbon applications Specific gravity only hydrocarbon applications Standard specific gravity only hydrocar bon and gas applications Auxiliary temperature Sensor temperature Stand...

Страница 113: ...iption Sets value at 100 of measuring span and corresponds to a loop current of 20 mA Setting and Default are sensor specific depending on the setting in PV selection and depending on the Units setting Setting 10800000000 00 kg h to 10800000000 00 kg h Default 21600000000 00 kg h Related Path Setup Inputs and outputs Channel 1 output with HART Upper range value 3 4 3 7 Lower range value Descriptio...

Страница 114: ...ponse to sudden changes in measurement An increase to Damping value increases response time of device and affects digital value and loop current If output values are noisy increase Damping value For faster response time decrease Damping value Find a value that meets requirements of signal stability and response time Setting 0 0 s to 100 0 s Default 0 0 s Related Path Setup Inputs and outputs Chann...

Страница 115: ...hannel 1 output with HART Fail safe value 3 4 3 12 Fail safe minimum time Description Minimum time the output stays in fail safe mode Setting 0 s to 100 s Default 0 s Related Path Setup Inputs and outputs Channel 1 output with HART Fail safe minimum time 3 4 4 Channel 2 output 3 4 4 1 Operation mode Description Operation mode Sets the output functionality for channel operation Setting Off Current ...

Страница 116: ...nel 2 output Fail safe activation condition 3 4 4 4 Current output 3 4 4 4 1 Process value Description Select process value for the current output Setting Volume flow Mass flow Standard volume flow only hydrocarbon and gas applications Flow velocity Sound velocity LiquIdent only hydrocarbon applications Density Kinemat ic viscosity Pressure Medium temperature Standard density only hydrocarbon and ...

Страница 117: ...lower fault current lower saturation limit to upper saturation limit upper fault current Description Setting 4 to 20 mA 3 50 3 8 to 20 5 22 6 NAMUR 4 to 20 mA 3 75 4 0 to 20 8 22 6 US 4 to 20 mA 0 00 4 0 to 20 5 22 0 4 to 20 mA 2 00 4 0 to 24 0 25 0 0 to 20 mA 0 00 0 0 to 20 5 22 0 0 to 20 mA 0 00 0 0 to 24 0 25 0 Default 4 to 20 mA 3 50 3 8 to 20 5 22 6 NAMUR Related Path Setup Inputs and outputs...

Страница 118: ...ange value 2 4 4 4 6 Damping value Description Sets the Damping filtering to smooth out response to sudden changes in measurement An increase to Damping value increases response time of device and affects digital value and loop current If output values are noisy increase Damping value For faster response time decrease Damping value Find a value that meets requirements of signal stability and respo...

Страница 119: ...tput Current output Fail safe minimum time 3 4 4 5 Frequency output 3 4 4 5 1 Process value Description Select process value for the frequency output Setting Volume flow Mass flow Standard volume flow only hydrocarbon and gas applications Flow velocity Sound velocity LiquIdent only hydrocarbon applications Density Kinemat ic viscosity Pressure Medium temperature Standard density only hydrocarbon a...

Страница 120: ...ctional symmetric Default Positive Related Path Setup Inputs and outputs Channel 2 output Frequency output Direction 3 4 4 5 3 Frequency value high Description Upper frequency value Upper nominal output range boundary Setting 0 0 Hz to 12500 0 Hz Default 10000 0 Hz Related Path Setup Inputs and outputs Channel 2 output Frequency output Frequency value high 3 4 4 5 4 Frequency value low Description...

Страница 121: ...5 6 Lower range value Description Sets value at 0 of measuring span and corresponds to Frequency value low Setting and Default are sensor specific depending on the setting in Process value and depending on the Units setting Setting 0 00 kg h to 3682800 00 kg h Default 0 00 kg h Related Path Setup Inputs and outputs Channel 2 output Frequency output Lower range value 3 4 4 5 7 Damping value Descrip...

Страница 122: ...red to Fail safe value Setting 0 0 Hz to 12500 0 Hz Default 0 0 Hz Related Path Setup Inputs and outputs Channel 2 output Frequency output Fail safe value 3 4 4 5 10 Fail safe minimum time Description Minimum time the output stays in fail safe mode Setting 0 s to 100 s Default 0 s Related Path Setup Inputs and outputs Channel 2 output Frequency output Fail safe minimum time 3 4 4 6 Pulse output 3 ...

Страница 123: ...rial gal lons BBL31 5 1 barrel 31 5 US gallons BBL31 1 barrel 31 US gallons BBL42 1 barrel 42 US gallons in cubic inches ft cubic feet yd cubic yards bush bush els Mass g grams kg kilograms t 1 t 1000 kg STon 1 STon 2000 lb T 1 T 2240 lb oz ounces lb pounds Standard volume Nl normal liters Nm normal cubic meters Sl standard liters Sft standard cubic feet Sm standard cubic meters cus tom units see ...

Страница 124: ...ate an output pulse Setting 0 0 l to 99999997952 l Default 1 00 I Related Path Setup Inputs and outputs Channel 2 output Pulse output Amount per pulse 3 4 4 6 7 Pulses per amount Description Number of pulses which will be output when configured amount is reached Setting 1 to 65535 Default 1 Related Path Setup Inputs and outputs Channel 2 output Pulse output Pulses per amount 3 4 4 6 8 Pulse width ...

Страница 125: ...of pulse output Setting Active high level Active low level Default Active high level Related Path Setup Inputs and outputs Channel 2 output Pulse output Polarity 3 4 4 6 11 Fail safe behaviour Description Current output reaction in case of a fault Setting Last valid value Hold Current value Fail safe value Maximum pulses s only if CH3 oper ation mode is configured to Pulse output and CH3 redundanc...

Страница 126: ...escription Minimum time the output stays in fail safe mode Short events will be prolonged Setting 0 s to 100 s Default 0 s Related Path Setup Inputs and outputs Channel 2 output Pulse output Fail safe minimum time 3 4 4 7 Status output 3 4 4 7 1 Status mode Description Select the functionality for the status output Setting Alarm class Alarm item Default Alarm class Related Path Setup Inputs and ou...

Страница 127: ...6 Configuration 2 37 DSL system monitor Default None Related Path Setup Inputs and outputs Channel 2 output Status output Sensor alarms 2 3 4 4 7 4 Process alarms 1 Description Selection of alarm items that impact the output Applicable if Status mode Alarm item Setting 96 Mass flow above alarm limit 97 Mass flow above warning limit 98 Mass flow below warning limit 99 Mass flow below alarm limit 10...

Страница 128: ...ound velocity above alarm limit 229 Sound velocity above warning limit 230 Sound velocity below warning limit 231 Sound velocity below alarm limit 232 Flow velocity above alarm limit 233 Flow velocity above warning limit 234 Flow velocity below warning limit 235 Flow velocity below alarm limit 236 Pressure above alarm limit 237 Pressure above warning limit 238 Pressure below warning limit 239 Pres...

Страница 129: ...below alarm limit Default None Related Path Setup Inputs and outputs Channel 2 output Status output Process alarms 4 3 4 4 7 8 Totalizer alarms Description Selection of alarm items that impact the output Applicable if Status mode Alarm item Setting 136 Totalizer 1 above alarm limit 137 Totalizer 1 above warning limit 138 Totalizer 1 below warning limit 139 Totalizer 1 below alarm limit 140 Totaliz...

Страница 130: ... output frequency too high 200 CH2 pulse overflow 201 CH3 loop cur in lower saturation 202 CH3 loop cur in upper saturation 203 CH3 cable break 204 CH3 output frequency too low 205 CH3 output fre quency too high 206 CH3 pulse overflow 207 CH4 loop cur in lower saturation 208 CH4 loop cur in upper saturation 209 CH4 cable break 210 CH4 output frequency too low 211 CH4 output frequency too high 212 ...

Страница 131: ...alarms 2 Setting 289 Sound velocity simulated 290 Flow velocity simulated 291 Pressure simulated 292 Kin viscosity simulated 293 Aux temperature simulated 294 Sensor temperature simu lated 296 CH5 simulated 297 CH6 simulated 298 Std volume flow simulated 299 Standardizing factor simulated 300 Std kin viscosity simulated 301 Std density simulat ed 303 API gravity simulated 304 Std API gravity simul...

Страница 132: ...utput Each diagnostic event is allocated to a NAMUR status signal Application if Status mode Alarm class and Status icons NAMUR Setting Failure Function check Out of specification Maintenance required Default None Related Path Setup Inputs and outputs Channel 2 output Status output NAMUR status signal 3 4 4 7 17 Polarity Description Logical polarity of status output Setting Active high level Activ...

Страница 133: ...hannel 2 output Status output Off delay 3 4 5 Channel 3 input output 3 4 5 1 Operation mode Description Shows the configured Operation mode Use Change operation mode to set the mode Read only Setting Off Current output Frequency output Pulse output Status output Digital input Current input Default Off Related Path Setup Inputs and outputs Channel 3 input output Operation mode 3 4 5 2 Change operat...

Страница 134: ...nt loop Default Related Path Setup Inputs and outputs Channel 3 input output Active operation 3 4 5 4 Fail safe activation condition Description Defines the condition the fail safe behaviour of the channel becomes active and determines the channel availability Setting Invalid process value Maintenance alarm or Failure NAMUR Default Maintenance alarm or Failure NAMUR Related Path Setup Inputs and o...

Страница 135: ...3 input output Current output Process value 3 4 5 5 2 Direction Description Flow direction filter applicable for process values mass flow volume flow flow velocity standard volume flow rate only hydrocarbon and gas applications Setting Positive Negative Bidirectional Bidirectional symmetric Default Positive Related Path Setup Inputs and outputs Channel 3 input output Current output Direction 3 4 5...

Страница 136: ...ing and Default are sensor specific depending on the setting in Process value and depending on the Units setting Setting 10800000000 0 kg h to 10800000000 0 kg h Default 0 00 kg h Related Path Setup Inputs and outputs Channel 3 input output Current output Lower range value 3 4 5 5 6 Damping value Description Sets the Damping filtering to smooth out response to sudden changes in measurement An incr...

Страница 137: ...input output Current output Fail safe behav iour 3 4 5 5 8 Fail safe value Description Output value in case of a fault and when fail safe behaviour configured to Fail safe value Setting 0 0 mA to 25 0 mA Default 0 0 mA Related Path Setup Inputs and outputs Channel 3 input output Current output Fail safe value 3 4 5 5 9 Fail safe minimum time Description Minimum time the output stays in fail safe m...

Страница 138: ...ns Flow velocity Sound velocity LiquIdent only hydrocarbon applications Density Kinemat ic viscosity Pressure Medium temperature Standard density only hydrocarbon and gas applications Standard kinematic viscosity only hydrocarbon applications API gravity only hydrocarbon applications Standard API gravity only hydrocarbon applications Specific gravity only hydrocarbon applications Standard specific...

Страница 139: ...ncy value low Description Lower frequency value Lower nominal output range boundary Setting 0 0 Hz to 12500 0 Hz Default 0 0 Hz Related Path Setup Inputs and outputs Channel 3 input output Frequency output Frequency value low 3 4 5 6 6 Upper range value Description Sets value at 100 of measuring span and corresponds to Frequency value high Setting and Default are sensor specific depending on the s...

Страница 140: ... output Lower range value 3 4 5 6 8 Damping value Description Time constant of damping filter for frequency output signal Setting 0 0 s to 100 0 s Default 0 0 s Related Path Setup Inputs and outputs Channel 3 input output Frequency output Damping value 3 4 5 6 9 Fail safe behaviour Description Frequency output reaction in case of a fault Setting Minimum frequency Maximum frequency Last valid value...

Страница 141: ...puts and outputs Channel 3 input output Frequency output Fail safe value 3 4 5 7 Pulse output 3 4 5 7 1 Redundancy mode Description Sets the redundancy mode The pulses on channel 3 will follow pulses on channel 2 with 90 or 180 shift Setting Off Channel 3 runs in independently 90 Channel 3 follows pulses on channel 2 with a 90 shift 180 Channel 3 follows pulses on channel 2 with a 180 shift Defaul...

Страница 142: ...duration values of Ch3 in pulse mode Setting Volume l liters hl hectoliters m cubic meters gal US gallons i gal Imperial gal lons BBL31 5 1 barrel 31 5 US gallons BBL31 1 barrel 31 US gallons BBL42 1 barrel 42 US gallons in cubic inches ft cubic feet yd cubic yards bush bush els Mass g grams kg kilograms t 1 t 1000 kg STon 1 STon 2000 lb T 1 T 2240 lb oz ounces lb pounds Standard volume Nl normal ...

Страница 143: ...s Channel 3 output Pulse outputs Custom conversion factor 3 4 5 7 7 Amount per pulse Description Amount of value to generate an output pulse Setting 0 0 l to 99999997952 l Default 1 00 I Related Path Setup Inputs and outputs Channel 3 input output Pulse output Amount per pulse 3 4 5 7 8 Pulses per amount Description Number of pulses which will be output when configured amount is reached Setting 1 ...

Страница 144: ...10 Pulse width Description Pulse duration Setting 0 00 s to 4 00 s Default 0 10 s Related Path Setup Inputs and outputs Channel 3 input output Pulse output Pulse width 3 4 5 7 11 Polarity Description Logical polarity of pulse output Setting Active high level Active low level Default Active high level Related Path Setup Inputs and outputs Channel 3 input output Pulse output Polarity 3 4 5 7 12 Fail...

Страница 145: ...lses s Related Path Setup Inputs and outputs Channel 3 input output Pulse output Fail safe value 3 4 5 7 14 Fail safe minimum time Description Minimum time the output stays in fail safe mode Setting 0 s to 100 s Default 0 s Related Path Setup Inputs and outputs Channel 3 input output Pulse output Fail safe minimum time 3 4 5 8 Status output 3 4 5 9 1 Status mode Description Select the functionalit...

Страница 146: ...he output Mainly sensor alarms Applicable if Status mode Alarm item Setting 32 Auxiliary input 34 Invalid flow meas 35 Failure on CH5 or CH6 36 Configuration 2 37 DSL system monitor Default None Related Path Setup Inputs and outputs Channel 3 input output Status output Sensor alarms 2 3 4 5 8 4 Process alarms 1 Description Selection of alarm items that impact the output Applicale if Stats mode Ala...

Страница 147: ...larm item For hydrocarbon and gas applications only Setting 228 Sound velocity above alarm limit 229 Sound velocity above warning limit 230 Sound velocity below warning limit 231 Sound velocity below alarm limit 232 Flow velocity above alarm limit 233 Flow velocity above warning limit 234 Flow velocity below warning limit 235 Flow velocity below alarm limit 236 Pressure above alarm limit 237 Press...

Страница 148: ...ath Setup Inputs and outputs Channel 3 input output Status output Process alarms 4 3 4 5 8 8 Totalizer alarms Description Selection of alarm items that impact the output Applicable if Status mode Alarm item For hydrocarbon and gas applications only Setting 136 Totalizer 1 above alarm limit 137 Totalizer 1 above warning limit 138 Totalizer 1 below warning limit 139 Totalizer 1 below alarm limit 140...

Страница 149: ...tting 195 CH2 loop cur in lower saturation 196 CH2 loop cur in upper saturation 197 CH2 cable break 198 CH2 output frequency too low 199 CH2 output frequency too high 200 CH2 pulse overflow 201 CH3 loop cur in lower saturation 202 CH3 loop cur in upper saturation 203 CH3 cable break 204 CH3 output frequency too low 205 CH3 output frequency too high 206 CH3 pulse overflow 207 CH4 loop cur in lower ...

Страница 150: ...h 191 CH4 external failure Default None Related Path Setup Inputs and outputs Channel 3 input output Status output Input output alarms 2 3 4 5 8 13 Simulation alarms 1 Description Selection of alarm items that impact the output Applicable if Status mode Alarm item Setting 160 Mass flow simulated 161 Volume flow simulated 162 Density simulated 163 Medium temp simulated 166 Std volume flow simulated...

Страница 151: ...larm class Description Selection of alarm classes which impact the status output Each diagnostic event is allocated to an alarm class Applicable if Status mode Alarm class and Status icons Standard Bit 0 Process value alarm PA Bit 1 Process value warning PW Bit 2 Maintenance alarm MA Bit 3 Maintenance warning MW Bit 4 Maintenance required MR Bit 5 Function check FC Setting Maintenance alarm Functi...

Страница 152: ... Inputs and outputs Channel 3 input output Status output Polarity 3 4 5 8 18 On delay Description Sets the time delay between alarm condition and activating the output Setting 0 0 s to 100 0 s Default 0 0 s Related Path Setup Inputs and outputs Channel 3 input output Status output On delay 3 4 5 8 19 Off delay Description Set the time delay between resetting alarm condition and de activating the o...

Страница 153: ...lated Path Setup Inputs and outputs Channel 3 input output Digital input Input function 3 4 5 9 2 Forced value Ch1 Description Forced value when forcing is active Only applicable for HART option Setting 3 50 mA to 25 00 mA Default 4 00 mA Related Path Setup Inputs and outputs Channel 3 input output Digital input Forced value Ch1 3 4 5 9 3 Forced value Ch2 Description Forced value when forcing is a...

Страница 154: ...put output Digital input Forced value Ch2 3 4 5 9 4 Forced value Ch4 Description Forced value when forcing is active Setting 3 50 mA to 25 00 mA Default 4 00 mA Current output Setting 0 0 Hz to 12500 00 Hz Default 1 00 Hz Frequency output Setting 0 00 pulses s to 12500 00 pulses s Default 1 00 pulses s Pulse output Setting 0 1 Default 0 Status output Related Path Setup Inputs and outputs Channel 3...

Страница 155: ...Description Logical polarity of discrete input signal Setting Active high level Active low level Default Active high level Related Path Setup Inputs and outputs Channel 3 input output Digital input Polarity 3 4 5 10 Current input 3 4 5 10 1 Process value Description Assignment of the current input value to a process value Read only Setting Off Pressure Density Kinematic viscosity Medium temperatur...

Страница 156: ...surement range and fail safe mode Setting 4 to 20 mA 3 50 3 8 to 20 5 22 6 NAMUR 4 to 20 mA 3 75 4 0 to 20 8 22 6 US 4 to 20 mA 0 00 4 0 to 20 5 22 0 4 to 20 mA 2 00 4 0 to 24 0 25 0 0 to 20 mA 0 00 0 0 to 20 5 22 0 0 to 20 mA 0 00 0 0 to 24 0 25 0 Default 4 to 20 mA 3 50 3 8 to 20 5 22 6 NAMUR Related Path Setup Inputs and outputs Channel 3 input output Current input Loop current scale 3 4 5 10 4...

Страница 157: ... outputs Channel 3 input output Current input Lower range value 3 4 5 10 6 Medium temperature Description Medium temperature Setting 50 0 C to 200 0 C Default 200 0 C Related Path Setup Inputs and outputs Channel 3 input output Current input Medium tempera ture 3 4 5 10 7 Density Description Density Setting 14000 00 kg m to 14000 00 kg m Default 14000 00 kg m Related Path Setup Inputs and outputs ...

Страница 158: ...sor temperature Setting 50 0 C to 200 0 C Default 200 0 C Related Path Setup Inputs and outputs Channel 3 input output Current input Sensor tempera ture 3 4 5 10 10 Medium temperature Description Medium temperature Setting 50 0 C to 200 0 C Default 50 0 C Related Path Setup Inputs and outputs Channel 3 input output Current input Medium tempera ture 3 4 5 10 11 Density Description Density Setting 1...

Страница 159: ...input output Current input Kinematic vis cosity 3 4 5 10 13 Sensor temperature Description Sensor temperature Setting 50 0 C to 200 0 C Default 200 0 C Related Path Setup Inputs and outputs Channel 3 input output Current input Sensor tempera ture 3 4 5 10 14 Fail safe behaviour Description Reaction in case of a fault at the current input Setting Last valid value Fail safe value Default Last valid ...

Страница 160: ...outputs Channel 3 input output Current input Fail safe value 4 5 3 10 16 Fail safe minimum time Description Minimum time the input stays in fail safe mode Setting 0 s to 100 s Default 0 s Related Path Setup Inputs and outputs Channel 3 input output Current input Fail safe mini mum time 3 4 6 Channel 3 relay 3 4 6 1 Operation mode Description Select the operation mode for the relay output Setting A...

Страница 161: ...al 09 Oil Density 10 Oil Viscosity 11 Oil spec gravity 14 Path 1 No signal 15 Path 2 No signal 16 Path 3 No signal 17 Path 4 No signal 19 Electronics temp 22 Density calculation 23 Medium temp calc 24 Pressure calculation 25 Viscosity calculation 26 Sensor temp comp 28 Default None Related Path Setup Inputs and outputs Channel 3 relay Sensor alarms 1 3 4 6 4 Sensor alarms 2 Description Selection o...

Страница 162: ... impact the output Applicable if Status mode Alarm item For hydrocarbon and gas applications only Setting 132 Std vol flow above alarm limit 133 Std vol flow above warning limit 134 Std vol flow below warning limit 135 Std vol flow below alarm limit Default None Related Path Setup Inputs and outputs Channel 3 relay Process alarms 2 3 4 6 7 Process alarms 3 Description Selection of alarm items that...

Страница 163: ...ow warning limit 271 Specific gravity below alarm limit 272 Std spec grav above alarm limit 273 Std spec grav above warning limit 274 Std spec grav bel warning limit 275 Std spec grav bel alarm limit 276 LiquIdent above alarm limit 277 LiquIdent above warning limit 278 LiquIdent below warning limit 279 LiquIdent below alarm limit Default None Related Path Setup Inputs and outputs Channel 3 relay P...

Страница 164: ...ransmitter Default None Related Path Setup Inputs and outputs Channel 3 relay Device alarms 3 4 6 12 Channel 1 alarms Description Channel 1 alarms The output is active when one of the alarm items occurs Setting 153 CH1 loop cur in lower saturation 154 CH1 loop cur in upper saturation 155 CH1 loop current deviation 158 CH1 cable break Default None Related Path Setup Inputs and outputs Channel 3 rel...

Страница 165: ... CH4 pulse overflow 214 Channel 2 simulated 215 Channel 3 simulated 216 Channel 4 simulated 217 Process values frozen 218 All outputs forced 219 CH2 loop current deviation 220 CH3 loop current deviation 221 CH4 loop current deviation 222 Invalid register mapping 223 Invalid coil configuration Default None Related Path Setup Inputs and outputs Channel 3 relay Input output alarms 3 4 6 14 Input outp...

Страница 166: ...atus mode Alarm item Setting 289 Sound velocity simulated 290 Flow velocity simulated 291 Pressure simulated 292 Kin viscosity simulated 293 Aux temperature simulated 294 Sensor temperature simu lated 296 CH5 simulated 297 CH6 simulated 288 Std volume flow simulated 299 Standardizing factor simulated 300 Std Kin viscosity simulated 301 Std density simulat ed 302 LiquIdent simulated 303 API gravity...

Страница 167: ... status output Each diagnostic event is allocated to a NAMUR status signal Application if Status mode Alarm class and Status icons NAMUR Setting Failure Function Check Out of specification Maintenance Required Default None Related Path Setup Inputs and outputs Channel 3 relay NAMUR status signal 3 4 6 20 Polarity Description Logical polarity of status output Setting Active high level Active low le...

Страница 168: ...annel 3 relay Off delay 3 4 7 Channel 4 input output 3 4 7 1 Operation mode Description Shows the configured Operation mode Use Change operation mode to set the mode Read only Setting Off Current output Frequency output Pulse output Signal output Discrete input Analog input Default Off Related Path Setup Inputs and outputs Channel 4 input output Operation mode 3 4 7 2 Change operation mode Descrip...

Страница 169: ...t loop Default Related Path Setup Inputs and outputs Channel 4 input output Active operation 3 4 7 4 Fail safe activation condition Description Defines the condition the fail safe behaviour of the channel becomes active and determines the channel availability Setting Invalid process value Maintenance alarm or Failure NAMUR Default Maintenance alarm or Failure NAMUR Related Path Setup Inputs and ou...

Страница 170: ...4 input output Current output Process value 3 4 7 5 2 Direction Description Flow direction filter applicable for process values mass flow volume flow flow velocity standard volume flow rate only hydrocarbon and gas applications Setting Positive Negative Bidirectional Bidirectional symmetric Default Positive Related Path Setup Inputs and outputs Channel 4 input output Current output Direction 3 4 7...

Страница 171: ...ing and Default are sensor specific depending on the setting in Process value and depending on the Units setting Setting 10800000000 0 kg h to 10800000000 0 kg h Default 0 00 kg h Related Path Setup Inputs and outputs Channel 4 input output Current output Lower range value 3 4 7 5 6 Damping value Description Sets the Damping filtering to smooth out response to sudden changes in measurement An incr...

Страница 172: ...4 7 5 8 Fail safe value Description Output value in case of a fault and when fail safe mode configured to Fail safe value Setting 0 0 mA to 25 0 mA Default 0 0 mA Related Path Setup Inputs and outputs Channel 4 input output Current output Fail safe value 3 4 7 5 9 Fail safe minimum time Description Minimum time the output stays in fail safe mode Setting 0 s to 100 s Default 0 s Related Path Setup ...

Страница 173: ...zing factor only hydrocarbon and gas applications Default Volume flow Related Path Setup Inputs and outputs Channel 4 input output 4 Frequency output Process value 3 4 7 6 2 Direction Description Flow direction filter applicable for process values mass flow volume flow flow velocity standard volume flow rate only hydrocarbon and gas applications Setting Positive Negative Bidirectional Bidirectiona...

Страница 174: ...ing 10800000000 00 kg h to 10800000000 00 kg h Default 10800000000 00 kg h Related Path Setup Inputs and outputs Channel 4 input output Frequency output Upper range value 3 4 7 6 6 Lower range value Description Sets value at 0 of measuring span and corresponds to Frequency value low Setting and Default are sensor specific depending on the setting in Process value and depending on the Units setting...

Страница 175: ...value Current valid fail safe value Default Last valid value Related Path Setup Inputs and outputs Channel 4 input output Frequency output Fail safe behaviour 3 4 7 7 9 Fail safe minimum time Description Minimum time the output stays in fail safe mode Setting 0 s to 100 s Default 0 s Related Path Setup Inputs and outputs Channel 4 input output Frequency output Fail safe minimum time 3 4 7 6 10 Fai...

Страница 176: ...Bidrectional Bidirectional symmetric Default Positive Related Path Setup Inputs and outputs Channel 4 input output Pulse output Direction 3 4 7 7 3 Pulse units Description Units of pulse duration values of Ch4 in pulse mode Setting Volume l liters hl hectoliters m cubic meters gal US gallons i gal Imperial gal lons BBL31 5 1 barrel 31 5 US gallons BBL31 1 barrel 31 US gallons BBL42 1 barrel 42 US ...

Страница 177: ...and outputs Channel 4 input output Pulse output Custom units 3 4 7 7 5 Custom conversion factor Description Conversion factor for user specific pulse values Setting Default Related Path Setup Inputs and outputs Channel 4 input output Pulse outputs Custom conver sion factor 3 4 7 7 6 Amount per pulse Description Amount of value to generate an output pulse Setting 0 0 l to 99999997952 l Default 1 00...

Страница 178: ...ue used to configure Ch4 pulse width Setting s seconds ms milliseconds μs micro seconds Default s seconds Related Path Setup Inputs outputs Channel 4 input output Pulse output Pulse width units 3 4 7 7 9 Polarity Description Logical polarity of pulse output Setting Active high level Active low level Default Active high level Related Path Setup Inputs and outputs Channel 4 input output Pulse output...

Страница 179: ...lses s Related Path Setup Inputs and outputs Channel 4 input output Pulse output Fail safe value 3 4 7 7 12 Fail safe minimum time Description Minimum time the output stays in fail safe mode Setting 0 s to 100 s Default 0 s Related Path Setup Inputs and outputs Channel 4 input output Pulse output Fail safe minimum time 3 4 7 8 Status output 3 4 7 8 1 Status mode Description Select the functionalit...

Страница 180: ...he output Mainly sensor alarms Applicable if Status mode Alarm item Setting 32 Auxiliary input 34 Invalid flow meas 35 Failure on CH5 or CH6 36 Configuration 2 37 DSL system monitor Default None Related Path Setup Inputs and outputs Channel 4 input output Status output Sensor alarms 2 3 4 7 8 4 Process alarms 1 Description Selection of alarm items that impact the output Applicale if Stats mode Ala...

Страница 181: ... Alarm item For hydrocarbon and gas applications only Setting 228 Sound velocity above alarm limit 229 Sound velocity above warning limit 230 Sound velocity below warning limit 231 Sound velocity below alarm limit 232 Flow velocity above alarm limit 233 Flow velocity above warning limit 234 Flow velocity below warning limit 235 Flow velocity below alarm limit 236 Pressure above alarm limit 237 Pre...

Страница 182: ...ath Setup Inputs and outputs Channel 4 input output Status output Process alarms 4 3 4 7 8 8 Totalizer alarms Description Selection of alarm items that impact the output Applicable if Status mode Alarm item For hydrocarbon and gas applications only Setting 136 Totalizer 1 above alarm limit 137 Totalizer 1 above warning limit 138 Totalizer 1 below warning limit 139 Totalizer 1 below alarm limit 140...

Страница 183: ...cur in lower saturation 196 CH2 loop cur in upper saturation 197 CH2 cable break 198 CH2 output frequency too low 199 CH2 output frequency too high 200 CH2 pulse overflow 201 CH3 loop cur in lower saturation 202 CH3 loop cur in upper saturation 203 CH3 cable break 204 CH3 output frequency too low 205 CH3 output frequency too high 206 CH3 pulse overflow 207 CH4 loop cur in lower saturation 208 CH4 ...

Страница 184: ...h 191 CH4 external failure Default None Related Path Setup Inputs and outputs Channel 4 input output Status output Input output alarms 2 3 4 7 8 13 Simulation alarms 1 Description Selection of alarm items that impact the output Applicable if Status mode Alarm item Setting 160 Mass flow simulated 161 Volume flow simulated 162 Density simulated 163 Medium temp simulated 166 Std volume flow simulated...

Страница 185: ...larm class Description Selection of alarm classes which impact the status output Each diagnostic event is allocated to an alarm class Applicable if Status mode Alarm class and Status icons Standard Bit 0 Process value alarm PA Bit 1 Process value warning PW Bit 2 Maintenance alarm MA Bit 3 Maintenance warning MW Bit 4 Maintenance required MR Bit 5 Function check FC Setting Maintenance alarm Functi...

Страница 186: ...up Inputs and outputs Channel 4 input output Status output Polarity 3 4 7 8 18 On delay Description Sets the time delay between alarm condition and activating the output Setting 0 0 s to 100 0 s Default 0 0 s Related Path Setup Inputs outputs Channel 4 input output Status output On delay 3 4 7 8 19 Off delay Description Set the time delay between resetting alarm condition and de activating the out...

Страница 187: ...d Path Setup Inputs and outputs Channel 4 input output Digital input Input function 3 4 7 9 2 Forced vlaue Ch1 Description Forced value when forcing is active Only applicable for HART option Setting 3 50 mA to 25 00 mA Default 4 00 mA Related Path Setup Inputs and outputs Channel 4 input output Digital input Forced value Ch1 3 4 7 9 3 Forced value Ch2 Description Forced value when forcing is activ...

Страница 188: ...ption The debounce time used to delay the hardware input signal can be configured Setting 0 0 ms 0 5 ms 1 ms 1 5 ms 2 ms 2 5 ms 3 ms 3 5 ms 4 ms 4 5 ms 5 ms 5 5 ms 6 ms 6 5 ms 7 ms 7 5 ms Default 0 0 ms Related Path Setup Inputs and outputs Channel 4 input output Digital input Delay time 3 4 7 9 6 Polarity Description Logical polarity of discrete input signal Setting Active high level Active low l...

Страница 189: ...ge process value selection enable the changing of the process value Setting Off Pressure Medium temperature Density Kinematic viscosity Sensor temperature Auxiliary temperature Default Off Related Path Setup Inputs and outputs Channel 4 input output Current input Change process value 3 4 7 10 3 Loop current scale Description Loop current scale Selects current input scaling according to desired mea...

Страница 190: ...lated Path Setup Inputs and outputs Channel 4 input output Current input Upper range value 3 4 7 10 5 Lower range value Description Lower pressure value that is mapped to the upper nominal range boundary selected by current mode Applicable if process value Pressure Setting Default 0 0 Pa Related Path Setup Inputs and outputs Channel 4 input output Current input Lower range value 3 4 7 10 6 Medium ...

Страница 191: ... Kinematic viscosity Description Kinematic viscosity Setting Default 1 00 m s Related Path Setup Inputs and outputs Channel 4 input output Current input Kinematic vis cosity 3 4 7 10 9 Sensor temperature Description Sensor temperature Setting 50 0 C to 200 0 C Default 200 0 C Related Path Setup Inputs and outputs Channel 3 input output Current input Sensor tempera ture 3 4 7 10 10 Medium temperatu...

Страница 192: ... m Related Path Setup Inputs and outputs Channel 3 input output Current input Density 3 4 7 10 12 Kinematic viscosity Description Kinematic viscosity Setting Default 0 00 m s Related Path Setup Inputs and outputs Channel 4 input output Current input Kinematic vis cosity 3 4 7 10 13 Sensor temperature Description Sensor temperature Setting 50 0 C to 200 0 C Default 200 0 C Related Path Setup Inputs...

Страница 193: ...ehav iour 3 4 7 10 15 Fail safe value Description Input value in case of a fault and when fail safe behaviour is configured to Fail safe value Applicable if process value Fail safe value Setting Default 0 0 Pa Related Path Setup Inputs and outputs Channel 4 input output Current input Fail safe value 3 4 7 10 16 Fail safe minimum time Description Minimum time the input stays in fail safe mode Setti...

Страница 194: ...Path Setup Inputs and outputs Channel 4 relay Status mode 3 4 8 3 Sensor alarms 1 Description Selection of alarm items that impact the output Mainly sensor alarms Applicable if Status mode Alarm item Setting 05 DSL voltages 06 DSL Storage 07 Flow measurement 08 DSL internal 09 Oil Density 10 Oil Viscosity 11 Oil spec gravity 14 Path 1 No signal 15 Path 2 No signal 16 Path 3 No signal 17 Path 4 No ...

Страница 195: ... limit 105 Density above warning limit 106 Density below warning limit 107 Density below alarm limit 108 Medium temp above alarm limit 109 Medium temp above warning limit 110 Medium temp below warning limit 111 Medium temp below alarm limit Default None Related Path Setup Inputs and outputs Channel 4 relay Process alarms 1 3 4 8 6 Process alarms 2 Description Selection of alarm items that impact t...

Страница 196: ...ocess alarms 3 3 4 8 8 Process alarms 4 Description Selection of alarm items that impact the output Applicable if Status mode Alarm item For hydrocarbon and gas applications only Setting 256 Standard density above alarm limit 257 Standard density above warning limit 258 Standard density below warning limit 259 Standard density below alarm limit 260 API gravity above alarm limit 261 API gravity abo...

Страница 197: ...talizer 3 below alarm limit Default None Related Path Setup Inputs and outputs Channel 4 relay Totalizer alarms 3 4 8 10 Device alarms Description Device alarms If one of the selected events appears this affects the status icon the related status output and the alarm on the communication interface Setting 148 Transm temp above alarm limit 149 Transm temp below alarm limit 150 Sensor signal disrupt...

Страница 198: ... impact the output Mainly simulation alarms Applicable if Status mode Alarm item Setting 160 Mass flow simulated 161 Volume flow simulated 162 Density simulated 163 Medium temp simulated 166 Std volume flow simulated 167 Totalizer 1 simulated 168 Totalizer 2 simulated 169 Totalizer 3 simulated 170 Loop current simulated Default None Related Path Setup Inputs and outputs Channel 4 relay Simulation ...

Страница 199: ...c event is allocated to a NAMUR status signal Application if Status mode Alarm class and Status icons NAMUR Setting Failure Function check Out of specification Maintenance required Default None Related Path Setup Inputs and outputs Channel 4 relay NAMUR status signal 3 4 8 16 Polarity Description Logical polarity of status output Setting Active high level Active low level Default Active high level...

Страница 200: ...h Setup Inputs and outputs Channel 4 relay Off delay 3 4 9 Channel 5 input 3 4 9 1 Process value Description Assignment of the current input value to a process value Setting Off Medium temperature RTD Sensor temperature RTD Auxiliary temperature RTD Default Off Related Path Setup Inputs and outputs Channel 5 Input Process value 3 4 9 2 RTD configuration 3 4 9 2 1 RTD type Description Type of conne...

Страница 201: ...Path Setup Inputs and outputs Channel 5 input RTD configuration RTD wire configu ration 3 4 9 2 3 RTD wire resistance Description Resistance of cables in a 2 wire RTD system Setting 0 0 ohms to 50 0 ohms Default 0 0 ohms Related Path Setup Inputs and outputs Channel 5 input RTD configuration RTD wire re sistance 3 4 9 3 Calibration 3 4 9 3 1 Offset Description Calibration offset Setting 100 C 148 ...

Страница 202: ... Description Assignment of the current input value to a process value Setting Off Medium temperature RTD Sensor temperature RTD Auxiliary temperature RTD Default Off Related Path Setup Inputs and outputs Channel 6 input Process value 3 4 10 2 RTD configuration 3 4 10 2 1 RTD type Description Type of connected RTD Setting 100 ohms RTD 500 ohms RTD 1000 ohms RTD Default 100 ohms RTD Related Path Set...

Страница 203: ... RTD wire configu ration 3 4 10 2 3 RTD wire resistance Description Resistance of cables in a 2 wire RTD system Setting 0 0 ohms to 50 0 ohms Default 0 0 ohms Related Path Setup Inputs and outputs Channel 6 input RTD configuration RTD wire re sistance 3 4 10 3 Calibration 3 4 10 3 1 Offset Description Calibration offset Setting 100 C 148 F to 100 C 212 F Default 0 C 32 F Related Path Setup Inputs ...

Страница 204: ...e Description Displays the current date and time Read only Setting Default Related Path Setup Date and time Current date and time 3 5 2 Set date and time Description Sets current date and time for the device This date and time is used for all timestamps of logged information Setting Default Related Path Setup Date and time Set date and time 3 6 Local display The system can be configured to show up...

Страница 205: ...aph 1 value and graph Six values Operation view Totalizer Alarm view Alarm list Diagnostic view Six diagnostic values 3 6 1 Brightness Description Adjust backlight level to improve local display conditions Setting 0 10 20 30 40 50 60 70 80 90 100 Default 50 Related Path Setup Local display Brightness 3 6 2 Backlight Description Select between always on or backlight switch off after 10 minutes Sett...

Страница 206: ...Damping 3 6 4 1 Damping value Description The damping filter time constant Setting 0 0 to 100 0 Default 10 s Related Path Setup Local display Damping Damping value 3 6 4 2 Process values 1 Description Available process values for the meter Setting Volume flow Mass flow Totalizer 1 Totalizer 2 Totalizer 3 Sound velocity Flow velocity Pressure Kinematic viscosity Standard volume flow Standardizing f...

Страница 207: ...hree values 1 value and bargraph 1 value and graph Six values Six diag nostic values Default Three values Related If Single value is selected 1st process value must be configured If Three values is selected 1st process value 2nd process value and 3rd process value must be configured If 1 value and bargraph is selected 1st process value must be configured If 1 value and graph is selected 1st proces...

Страница 208: ...d API gravity Specific gravity Standard specific gravity Default Volume flow Related Path Setup Local display View 1 1st value 3 6 5 3 Graph scale mode Description Graph scale mode Setting Auto Fixed Default Auto Related Path Setup Local display View 1 Graph scale mode 3 6 5 4 2nd value Description The display 2nd process value can be selected Setting Volume flow Mass flow Flow velocity Sound velo...

Страница 209: ...ting Volume flow Mass flow Flow velocity Sound velocity Density Process viscosity Process pressure Fluid temperature Reference density Totalizer 1 Totalizer 2 Totalizer 3 Default Sound velocity Related This parameter is only available if View 1 is set to Six values or Six diagnostic values Path Setup Local display View 1 4th process value 3 6 5 7 5th process value Description The display 5th proce...

Страница 210: ...ue 3 6 5 9 Trend scale mode Description The Trend scale mode can be selected with automatic scaling or fixed scaling Setting Auto Fixed Default Auto Related This parameter is only available if View is set to 1 value and graph If Fixed is selected Trend scale lower limit and Trend scale upper limit must be configured Path Setup Local display View 1 Trend scale mode 3 6 5 10 Trend log time window De...

Страница 211: ... 3 6 5 12 Trend scale upper limit Description The Trend scale upper limit defines the scaling upper limit of the value axis for the fixed mode Setting options Default setting 0 Related This parameter is only available if Trend scale mode is set to Fixed Path Setup Local display View 1 Trend scale upper limit 3 6 6 View 2 3 6 6 1 Enable or disable Description In this parameter View 2 can be enabled...

Страница 212: ...s value must be configured If Six values or Six diagnostic values is selected 1st process value 2nd process value 3rd process value 4th process value 5th process value and 6th process value must be configured Path Setup Local display View 2 View 3 6 6 3 1st value Description The display 1st Process Value can be selected This process value is shown in all view types Setting Volume flow Mass flow St...

Страница 213: ...e Corrected Volume flow Reference Density Totalizer 1 Totalizer 2 Totalizer 3 Default Fluid Temperature Related This parameter is only available if View is set to Three Values Six Values or Six Diagnostic Values Path Setup Display View 2 3rd Process Value 3 6 6 6 4th process value Description The display 4th Process Value can be selected Setting Mass flow Volume flow Density Fluid Temperature Corr...

Страница 214: ...d Volume flow Reference Density Totalizer 1 Totalizer 2 Totalizer 3 Default Totalizer 2 Related This parameter is only available if View is set to Six Values or Six Diagnostic Values Path Setup Display View 2 6th Process Value 3 6 6 9 Trend scale mode Description The Trend Scale Mode can be selected with automatic scaling or fixed scaling Setting Auto Fixed Default Auto Related This parameter is o...

Страница 215: ...n The Trend Scale Lower Limit defines the scaling lower limit of the value axis for the fixed mode Setting Default 0 Related This parameter is only available if Trend Scale Mode is set to Fixed Path Setup Display View 2 Trend Scale Lower Limit 3 6 6 12 Trend scale upper limit Description The Trend Scale Upper Limit defines the scaling upper limit of the value axis for the fixed mode Setting option...

Страница 216: ...ue and bargraph Related If Single value or Dosing is selected 1st process value must be configured If Three values is selected 1st process value 2nd process value and 3rd process value must be configured If Totalizer or 1 value and graph is selected 1st process value Trend scale mode and Trend log time window must be configured If 1 value and bargraph is selected 1st process value must be configur...

Страница 217: ...ixed Default Auto Related Path Setup Display View 3 Graphic scale mode 3 6 7 5 2nd process value Description The display 2nd Process Value can be selected Setting Volume flow Mass flow Standard volume flow Flow velocity Sound velocity Liquid identi fier LiquIdent Density Kinematic viscosity Totalizer 1 Totalizer 2 Totalizer 3 Pressure Medium Temperature Standard density Standard kinematic viscosit...

Страница 218: ...rocess Value 3 6 7 7 4th process value Description The display 4th Process Value can be selected Setting Mass flow Volume flow Density Fluid Temperature Corrected Volume flow Reference Density Totalizer 1 Totalizer 2 Totalizer 3 Default Volumeflow Related This parameter is only available if View is set to Six Values or Six Diagnostic Values Path Setup Display View 3 4th Process Value 3 6 7 8 5th p...

Страница 219: ...tion The Trend Scale Mode can be selected with automatic scaling or fixed scaling Setting Auto Fixed Default Auto Related This parameter is only available if View is set to Totalizer or 1 Value and Graph If Fixed is selected Trend Scale Lower Limit and Trend Scale Upper Limit must be configured Path Setup Display View 3 Trend Scale Mode 3 6 7 11 Trend log time window Description The Trend log time...

Страница 220: ...mit 3 6 7 13 Trend scale upper limit Description The Trend Scale Upper Limit defines the scaling upper limit of the value axis for the fixed mode Setting options Default setting 0 Related This parameter is only available if Trend Scale Mode is set to Fixed Path Setup Display View 3 Trend Scale Upper Limit 3 6 8 View 4 3 6 8 1 Enable or disable Description In this parameter View 4 can be enabled Se...

Страница 221: ...ss Value 3rd Process Value 4th Process Value 5th Process Value and 6th Process Value must be configured Path Setup Display View 4 View 3 6 8 3 1st value Description The display 1st Process Value can be selected This process value is shown in all view types Setting Transmitter internal temperature DSL temperature Measured current CH5 Measured current CH6 Ch1 value Ch3 value Ch4 value Auxiliary temp...

Страница 222: ... or Six Diagnostic Values Path Setup Display View 4 2nd Value 3 6 8 5 3rd value Description The display 3rd Process Value can be selected Setting Transmitter internal temperature DSL temperature Measured current CH5 Measured current CH6 Ch1 value Ch3 value Ch4 value Auxiliary temperature Sensor temperature Reynolds number RxGain up path 1 RxGain down path 1 SNR up path 1 SNR down path 1 Sound velo...

Страница 223: ...or temperature Reynolds number RxGain up path 1 RxGain down path 1 SNR up path 1 SNR down path 1 Sound velocity path 1 Flow velocity path 1 Delta time path 1 Travel time up path 1 Travel time down path 1 Correlation factor path 1 Percentage of bursts accepted path1 Peak amplitude up path 1 Peak amplitude down path 1 Min acc travel time path 1 Min acc travel time path 1 Max acc travel time path 1 D...

Страница 224: ...h If Fixed is selected Trend Scale Lower Limit and Trend Scale Upper Limit must be configured Path Setup Display View 4 Trend Scale Mode 3 6 8 10 Trend log time window Description The Trend log time window logging period time axis length can be selected Setting 1 Minute 5 Minutes 15 Minutes 30 Minutes 1 Hour 2 Hours 3 Hours Default 5 Minutes Related This parameter is only available if View is set ...

Страница 225: ...This parameter is only available if Trend Scale Mode is set to Fixed Path Setup Display View 4 Trend Scale Upper Limit 3 6 9 View 5 3 6 9 1 Enable or disable Description In this parameter View 5 can be enabled Setting Disabled Enabled Default Enabled Related If Enabled is selected all parameters for View 5 must be configured Path Setup Display View 5 Enable or disable 3 6 9 2 View Description The ...

Страница 226: ...and 6th Process Value must be configured Path Setup Local display View 5 View 3 6 9 3 1st value Description The display 1st Process Value can be selected This process value is shown in all view types Setting Volume flow Mass flow Flow velocity Sound velocity Density Kinematic viscosity Total izer 1 Totalizer 2 Totalizer 3 Pressure Medium temperature Default Totalizer 1 Related Path Setup Display V...

Страница 227: ...isplay 4th Process Value can be selected Setting Mass flow Volume flow Density Fluid Temperature Corrected Volume flow Reference Density Totalizer 1 Totalizer 2 Totalizer 3 Default Volumeflow Related This parameter is only available if View is set to Six Values or Six Diagnostic Values Path Setup Display View 5 4th Process Value 3 6 9 7 5th process value Description The display 5th Process Value c...

Страница 228: ...tion The Trend Scale Mode can be selected with automatic scaling or fixed scaling Setting Auto Fixed Default Auto Related This parameter is only available if View is set to Totalizer or 1 Value and Graph If Fixed is selected Trend Scale Lower Limit and Trend Scale Upper Limit must be configured Path Setup Display View 5 Trend Scale Mode 3 6 9 10 Trend log time window Description The Trend log time...

Страница 229: ...it 3 6 9 12 Trend scale upper limit Description The Trend Scale Upper Limit defines the scaling upper limit of the value axis for the fixed mode Setting options Default setting 0 Related This parameter is only available if Trend Scale Mode is set to Fixed Path Setup Display View 5 Trend Scale Upper Limit 3 6 10 View 6 3 6 10 1 Enable or disable Description In this parameter View 6 can be enabled S...

Страница 230: ...s selected 1st Process Value must be configured If Six Values or Six Diagnostic Values is selected 1st Process Value 2nd Process Value 3rd Process Value 4th Process Value 5th Process Value and 6th Process Value must be configured Path Setup Local display View 6 View 3 6 10 3 1st process value Description The display 1st Process Value can be selected This process value is shown in all view types Se...

Страница 231: ...ew is set to Three Values Six Values or Six Diagnostic Values Path Setup Display View 6 3rd Process Value 3 6 10 6 4th process value Description The display 4th Process Value can be selected Setting Mass flow Volume flow Density Fluid Temperature Corrected Volume flow Reference Density Totalizer 1 Totalizer 2 Totalizer 3 Default Volumeflow Related This parameter is only available if View is set to...

Страница 232: ...erence Density Totalizer 1 Totalizer 2 Totalizer 3 Default Totalizer 2 Related This parameter is only available if View is set to Six Values or Six Diagnostic Values Path Setup Display View 6 6th Process Value 3 6 10 9 Trend scale mode Description The Trend Scale Mode can be selected with automatic scaling or fixed scaling Setting Auto Fixed Default Auto Related This parameter is only available if...

Страница 233: ...og time window 3 6 10 11 Trend scale lower limit Description The Trend Scale Lower Limit defines the scaling lower limit of the value axis for the fixed mode Setting Default 0 Related This parameter is only available if Trend Scale Mode is set to Fixed Path Setup Display View 6 Trend Scale Lower Limit 3 6 10 12 Trend scale upper limit Description The Trend Scale Upper Limit defines the scaling upp...

Страница 234: ... FST030 232 Function Manual 04 2017 A5E35693328 AA 3 6 11 Status icons Description Sets the type of status icons used for diagnostic events Setting Standard NAMUR Default Standard Related Path Setup Local display Status icons ...

Страница 235: ...Path Maintenance and diagnostics Identification Long tag 4 1 2 Tag Description Sets a unique short tag name for device or measurement point Limited to 8 characters Setting User configurable Default characters A B C X Y Z 1 2 3 7 8 9 Related Path Maintenance diagnostics Identification Tag 4 1 3 Descriptor Description Sets a unique description for the measurement point Limited to 16 characters Setti...

Страница 236: ...2 characters A B C X Y Z 1 2 3 7 8 9 Related Path Maintenance and diagnostics Identification Message 4 1 5 Location Description Sets device location Limited to 32 characters Setting User configurable Default 32 characters A B C X Y Z 1 2 3 7 8 9 Related Path Maintenance and diagnostics Identification Location 4 1 6 Installation date Description Sets the installation date of the device Setting User...

Страница 237: ...ys name of product Also shown on device nameplate Setting Default SITRANS FS230 Related Path Maintenance and diagnostics Identification Product name 4 1 9 Version read only Description The product Version according to order can be viewed Setting Default WM int DSL Related Path Maintenance and diagnostics Identification Version 4 1 10 Order number read only Description System order number MLFB can ...

Страница 238: ...tion Displays transmitter serial number Also shown on the device nameplate Setting Default Related Path Maintenance and diagnostics Identification Serial number 4 1 12 FW version Description Displays the firmware version of the device Setting Default Related Path Maintenance and diagnostics Identification FW version 4 1 13 HW version Description Displays the electronics hardware version of the dev...

Страница 239: ...components used in final device assembly Setting 0 to 4294967295 Default 0 Related Path Maintenance and diagnostics Identification Final assembly number 4 1 15 System type Description System type Setting SITRANS FUS SITRANS FUH Default SITRANS FUS Related Path Maintenance and diagnostics Identification System type 4 1 16 Transmitter electronics 4 1 16 1 HW version Description Displays the electron...

Страница 240: ...ification FW version 4 1 16 3 Serial number Description Displays transmitter serial number Also shown on the device nameplate Setting Default Related Path Maintenance and diagnostics Identification Transmitter Serial number 4 1 16 4 Order number Description Displays the order number of the current device configuration Setting Default Related Path Maintenance and diagnostics Identification Transmit...

Страница 241: ...display Description Local display Setting HW version 1 01 00 Default Setting FW version 1 04 13 09 Default Setting HMI cfg version 1 00 01 39 Default Related Path Maintenance and diagnostics Identification Local display 4 1 18 I O electronics Description I O electronics Setting HW version Default Setting FW version Default Setting Serial number Default Related Path Maintenance and diagnostics Iden...

Страница 242: ...tion The name of the digital sensor Path Maintenance and diagnostics Identification Sensor Type 4 1 20 2 Sensor type read only Description Identification of the sensor type Used by the transmitter to enable technology specific functionalities Path Maintenance and diagnostics Identification Sensor Sensor type 4 1 20 3 Serial number read only Description Unique frontend serial number Setting Default...

Страница 243: ...Serial number read only Description Unique serial number of sensor set path 2 clamp on Setting Default Related Path Maintenance and diagnostics Identification Sensor Sensors serial number path 2 4 1 20 6 Serial number read only Description Unique serial number of sensors set path 3 clamp on Setting Default Related Path Maintenance and diagnostics Identification Sensor Sensors serial number path 3 ...

Страница 244: ...p on SITRANS FSS600 Clamp on Default Related Path Maintenance and diagnostics Identification Sensor Sensor type 4 1 20 9 Sensor firmware version read only Description Sensor firmware version Setting Default Related Path Maintenance and diagnostics Identification Sensor Sensor firmware version 4 1 20 10 Sensor hardware version read only Description Sensor hardware version Setting Default Related Pa...

Страница 245: ...riant 4 1 20 12 Serial number read only Description Unique frontend PCBA serial number Setting Default Related Path Maintenance and diagnostics Identification Sensor Frontend PCBA serial number 4 2 Diagnostic events 4 2 1 Active diagnostic events Description Displays list of active events Path Maintenance and diagnostics Diagnostic events Active events 4 2 2 Diagnostic log Description Displays lis...

Страница 246: ...e and diagnostics Diagnostic events Clear diagnostic log 4 2 4 Acknowledge mode Description The system can be configured to acknowledge the alarms automatically when they are gone Auto or it can be configured to require an acknowledgement from the operator Manual Setting Auto Manual Default Auto Path Maintenance and diagnostics Diagnostic events Acknowledge mode 4 2 5 Transmitter detailed events r...

Страница 247: ...ule not found Path Maintenance and diagnostics Diagnostic events Transmitter detailed events 4 2 6 Suppression time Description A coming event can be suppressed in order to prevent alarm flooding Setting 0 to 300 s Default 0 s Path Maintenance and diagnostics Diagnostic events Suppression time 4 2 7 Enable alarms 4 2 7 1 Sensor events 1 Description Various sensor alarms can be enabled to show in t...

Страница 248: ...h Maintenance and diagnostics Diagnostic events Enable alarms Sensor events 2 4 2 7 3 Process events 1 Description Various sensor alarms can be enabled to show in the display and on the output if they occur Setting 228 Sound velocity above alarm limit 229 Sound velocity above warning limit 230 Sound velocity below warning limit 231 Sound velocity below alarm limit 232 Flow velocity above alarm lim...

Страница 249: ...arning limit 230 Sound velocity below warning limit 231 Sound velocity below alarm limit 232 Flow velocity above alarm limit 233 Flow velocity above warning limit 234 Flow velocity below warning limit 235 Flow velocity below alarm limit 236 Pressure above alarm limit 237 Pressure above warning limit 238 Pressure below warning limit 239 Pressure below alarm limit 244 Kin viscosity above alarm limit...

Страница 250: ...elow alarm limit 144 Totalizer 3 above alarm limit 145 Totalizer 3 above warning limit 146 Totalizer 3 below warning limit 147 Totalizer 3 below alarm limit Default Related The selected events will be shown in the display when they occur Furthermore they will be indicated on the output if Output 2 Input output 3 and or Input output 4 is set to Status output and Status mode is set to Alarm item Pat...

Страница 251: ...simulated 292 Kin viscosity simulated 293 Aux temperature simulated 294 Sensor temperature simu lated 296 CH5 simulated 297 CH6 simulated 298 Std volume flow simulated 299 Standardizing factor simulated 300 Std Kin viscosity simulated 301 Std density simulat ed 302 LiquIdent simulated 303 API gravity simulated 304 Std API gravity simulated 305 Specific gravity simulated 306 Std spec grav simulated...

Страница 252: ...cation interface Setting 186 CH3 input current too low 187 CH3 input current too high 188 CH3 external failure Default Related The selected events will be shown in the display when they occur Furthermore they will be indicated on the output if Output 2 Input output 3 and or Input output 4 is set to Status output and Status mode is set to Alarm item Path Maintenance and diagnostics Diagnostic event...

Страница 253: ...ilure Default Process value alarm PA resp Out of specification Path Maintenance and diagnostics Diagnostic events Assign alarm class 148 Transm temp above alarm limit 4 2 8 4 149 Transm temp below alarm limit Description This event can be mapped to an alarm class respective NAMUR status signal Setting Process value alarm PA resp Out of specification Maintenance alarm resp Failure Default Process a...

Страница 254: ...ce transmitter 4 3 2 1 2 Replace transmitter cassette Description Replace transmitter cassette Only Siemens authorized personnel may replace spare parts This action will restore the data from the SensorFlash on the transmitter cassette module Access level Expert user only Setting Cancel OK Default Cancel Path Maintenance and diagnostics Maintenance Spare part replacement Transmitter Replace transm...

Страница 255: ...ly Setting Cancel OK Default Cancel Path Maintenance and diagnostics Maintenance Spare part replacement Sensor Re place sensor 4 3 2 3 Operating time 4 3 2 3 1 Operating time Description Operating time Indicates the operating time of the transmitter since the last power up 4 3 2 3 2 Operating time total Description Operating time total Indicates the total operating time of the transmitter since th...

Страница 256: ...thorized personnel may replace spare parts Setting Cancel OK Default Cancel Path Maintenance and diagnostics Maintenance Electronic part identification Transmitter cassette HW version 4 3 3 1 2 FW version expert Description Only Siemens authorized personnel may replace spare parts Setting Cancel OK Default Cancel Path Maintenance and diagnostics Maintenance Electronic part identification Transmitt...

Страница 257: ...ance Electronic part identification Transmitter cassette Comm HW version 4 3 3 1 5 Comm serial number expert Description Only Siemens authorized personnel may replace spare parts Setting Cancel OK Default Cancel Path Maintenance and diagnostics Maintenance Electronic part identification Transmitter cassette Comm serial number 4 3 3 2 Local display 4 3 3 2 1 HW version expert Description Only Sieme...

Страница 258: ... Description Only Siemens authorized personnel may replace spare parts Setting Cancel OK Default Cancel Path Maintenance and diagnostics Maintenance Electronic part identification Local dis play HMI config version 4 3 3 3 I O cassette 4 3 3 3 1 HW version expert Description Only Siemens authorized personnel may replace spare parts Setting Cancel OK Default Cancel Path Maintenance and diagnostics M...

Страница 259: ...ts Setting Cancel OK Default Cancel Path Maintenance and diagnostics Maintenance Electronic part identification I O cassette Serial number 4 3 3 4 Sensor cassette 4 3 3 4 1 HW version expert Description Only Siemens authorized personnel may replace spare parts Setting Cancel OK Default Cancel Path Maintenance and diagnostics Maintenance Electronic part identification Sensor cassette HW version 4 3...

Страница 260: ...d personnel may replace spare parts Setting Cancel OK Default Cancel Path Maintenance and diagnostics Maintenance Electronic part identification Sensor cassette Serial number 4 4 Diagnostics 4 4 1 Sensor 4 4 1 1 Receiver signal Description Receiver signal Path Maintenance and diagnostics Diagnostics Sensor Receiver signal 4 4 1 2 Reynolds number read only Description Calculated Reynolds number Pat...

Страница 261: ...iagnostics Diagnostics Sensor Path 1 RxGain down 4 4 1 3 3 SNR up read only Description Signal to noise ration of upstream signal Path Maintenance and diagnostics Diagnostics Sensor Path 1 SNR up 4 4 1 3 4 SNR down read only Description Signal to noise ration of downstream signal Path Maintenance and diagnostics Diagnostics Sensor Path 1 SNR down 4 4 1 3 5 Sound velocity read only Description Soun...

Страница 262: ... Diagnostics Sensor Path 1 Travel time up 4 4 1 3 9 Travel time down read only Description Travel time of downstream signal Path Maintenance and diagnostics Diagnostics Sensor Path 1 Travel time down 4 4 1 3 10 Correlation factor read only Description Correlation factor Path Maintenance and diagnostics Diagnostics Sensor Path 1 Correlation factor 4 4 1 3 11 Actual sensor frequency read only Descri...

Страница 263: ... 4 1 4 1 RxGain up read only Description Rx gain of upstream signal Path Maintenance and diagnostics Diagnostics Sensor Path 2 RxGain up 4 4 1 4 2 RxGain down read only Description Rx gain of downstream signal Path Maintenance and diagnostics Diagnostics Sensor Path 2 RxGain down 4 4 1 4 3 SNR up read only Description Signal to noise ration of upstream signal Path Maintenance and diagnostics Diagn...

Страница 264: ...time path 2 read only Description Delta time Path Maintenance and diagnostics Diagnostics Sensor Path 2 Delta time 4 4 1 4 8 Travel time up read only Description Travel time of upstream signal Path Maintenance and diagnostics Diagnostics Sensor Path 2 Travel time up 4 4 1 4 9 Travel time down read only Description Travel time of downstream signal Path Maintenance and diagnostics Diagnostics Sensor...

Страница 265: ...time of frontend since last power up Path Maintenance and diagnostics Diagnostics Sensor Path 2 Peak amplitude down 4 4 1 4 13 Peak amplitude down read only Description Operating time of frontend since last power up Path Maintenance and diagnostics Diagnostics Sensor Path 2 Peak amplitude down 4 4 2 Temperature monitoring 4 4 2 1 Medium temperature Description Medium temperature Setting 4 00 C Def...

Страница 266: ...ure Minimum 4 4 2 2 3 Timestamp at minimum value read only Description Shows timestamp of recorded minimum transmitter internal temperature Path Maintenance and diagnostics Diagnostics Temperature monitoring Transmitter elec tronics temperature Timestamp at minimum value 4 4 2 2 4 Maximum read only Description Drag indicator of maximum transmitter internal temperature Path Maintenance and diagnost...

Страница 267: ...g 0 0 C Default Related Path Maintenance and diagnostics Diagnostic DSL temperature Current value 4 4 2 3 2 Minimum Description Minimum DSL temperature Setting 200 0 C Default Related Path Maintenance and diagnostics Diagnostics DSL temperature Minimum 4 4 2 3 3 Maximum Description Maximum DSL temperature Setting 200 0 C Default Related Path Maintenance and diagnostics Diagnostics DSL temperature ...

Страница 268: ...utput with HART Loop current 4 4 3 1 2 Error status read only Description Fault status if configured to current output Options Upper saturation Lower saturation Fail safe behaviour active Cable break Loop current deviation Related Path Maintenance and diagnostics Diagnostics Inputs and outputs Channel 1 output with HART Error status 4 4 3 2 Channel 2 output 4 4 3 2 1 Operation mode read only Descr...

Страница 269: ...Related if Operation mode is set to Current output this parameter will show the current output value if Operation mode is set to Frequency output this parameter will show the frequency output value if Operation mode is set to Pulse output this parameter will show the pulse output value if Operation mode is set to Status output this parameter will show the status output value Path Maintenance diagn...

Страница 270: ...e and diagnostics Diagnostics Inputs and outputs Channel 2 output Error status 4 4 3 2 5 Pulse counter read only Description This parameter shows the number of pulses since startup last reset Related This parameter is only available if Operation mode is set to Pulse output Path Maintenance and diagnostics Diagnostics Inputs and outputs Channel 2 output Pulse counter 4 4 3 2 6 Reset pulse counter D...

Страница 271: ...ut Path Maintenance and diagnostics Diagnostics Inputs and outputs Channel 3 input output Operation mode 4 4 3 3 2 Current input value read only Description Current input value Read only Options Read only Related Path Maintenance and diagnostics Diagnostics Inputs and outputs Channel 3 input output Current input value 4 4 3 3 3 Error status read only Description The error status of the channel can...

Страница 272: ...e pulse output value if Operation mode is set to Status output this parameter will show the status output value Path Maintenance and diagnostics Diagnostics Inputs and outputs Channel 3 in put output Channel value 4 4 3 3 5 Pulse counter read only Description This parameter shows the number of pulses since startup last reset Related Path Maintenance and diagnostics Diagnostics Inputs and outputs C...

Страница 273: ...viewed in this parameter Setting Off Current output Frequency output Pulse output Status output Digital input Current input Default Off Related In this parameter the setting of Operation mode is read only Path Maintenance and diagnostics Diagnostics Inputs and outputs Channel 4 in put output Operation mode 4 4 3 5 2 Channel value read only Description The channel value can be viewed in this parame...

Страница 274: ...e behaviour is activated Operation mode is set to Frequency output Options The pulse output has reached maximum pulse output frequency Fail safe behaviour is activated Operation mode is set to Pulse output Options The pulse output has reached maximum pulse output frequency Fail safe behaviour is activated Operation mode is set to Pulse output Options Loop current too high Loop current too low Fail...

Страница 275: ...t Cancel Related This parameter is only available if Operation mode is set to Pulse output Path Maintenance and diagnostics Diagnostics Inputs and outputs Channel 4 in put output Reset pulse counter 4 4 3 6 Channel 4 relay 4 4 3 6 1 Status output signal read only Description The channel value can be viewed in this parameter Related This parameter is only available if Operation mode is set to On Pa...

Страница 276: ...e RTD Auxiliary temperature RTD Totalizer 1 Totalizer 2 Totalizer 3 Default Disabled Related If this parameter is set to any other setting than Disabled parameters Maximum value Timestamp at maximum value Minimum value Timestamp at Minimum value and Reset logging will be available Path Maintenance and diagnostics Peak values Process value 1 Process value 4 5 1 2 Maximum Description Displays maximu...

Страница 277: ...ue 1 Minimum value 4 5 1 5 Timestamp at minimum value Description Displays the timestamp when minimum value was measured Related This parameter is only available if Process value is not set to Disabled Path Maintenance and diagnostics Peak values Process value 1 Minimum value timestamp 4 5 1 6 Reset peak values Description This command is used to reset the logging of the recorded maximum and minim...

Страница 278: ...ting than Disabled parameters Maximum value Maximum value timestamp Minimum value Minimum value timestamp and Reset logging will be available Path Maintenance and diagnostics Peak values Process value 2 Process value 4 5 2 2 Maximum Description Displays maximum measured value Related This parameter is only available if Process value is not set to Disabled Path Maintenance and diagnostics Peak valu...

Страница 279: ...s parameter is only available if Process value is not set to Disabled Path Maintenance and diagnostics Peak values Process value 2 Minimum value timestamp 4 5 2 6 Reset peak values Description This command is used to reset the logging of the recorded maximum and minimum values monitored The drag pointers will be set to the current value Setting Cancel Ok Default Cancel Related This parameter is on...

Страница 280: ...e and diagnostics Peak values Process value 3 Process value 4 5 3 2 Maximum Description Displays maximum measured value Related This parameter is only available if Process value is not set to Disabled Path Maintenance and diagnostics Peak values Process value 3 Maximum value 4 5 3 3 Timestamp at maximum value Description Displays the timestamp when maximum value was measured Related This parameter...

Страница 281: ...elated This parameter is only available if Process value is not set to Disabled Path Maintenance and diagnostics Peak values Process value 3 Reset logging 4 5 4 Process value 4 4 5 4 1 Process value Description Assign the process value to be monitored Setting Disabled Volume flow Mass flow Standard volume flow Sound velocity Flow velocity LiquIdent Density Medium temperature Pressure Kinematic vis...

Страница 282: ...eter is only available if Process value is not set to Disabled Path Maintenance and diagnostics Peak values Process value 4 Maximum value timestamp 4 5 4 4 Minimum Description Displays minimum measured value Related This parameter is only available if Process value is not set to Disabled Path Maintenance and diagnostics Peak values Process value 4 Minimum value 4 5 4 5 Timestamp at minimum value D...

Страница 283: ...isabled Path Maintenance and diagnostics Peak values Process value 4 Reset logging 4 6 Characteristics 4 6 1 Transmitter 4 6 1 1 Design read only Description Information about the main design features of the transmitter Path Maintenance and diagnostics Characteristics Transmitter Design 4 6 1 2 Hazardous area approval read only Description Text describing the hazardous area approval of the transmi...

Страница 284: ...n Eject Sensor Flash card Setting Cancel Ok Default Cancel Related Path Maintenance and diagnostics SensorFlash Eject 4 7 2 Installed read only Description Shows the status of the SD card Setting SensorFlash not installed SensorFlash installed SensorFlash used as mass storage de vice Default SensorFlash installed Related Path Maintenance and diagnostics SensorFlash Installed 4 7 3 Capacity total r...

Страница 285: ...acity of installed SensorFlash Shows INSERT CARD if card is not installed Setting Default GB Related Path Maintenance and diagnostics SensorFlash Capacity available 4 7 5 Data logging 4 7 5 1 Activation Description Activation turns on and off data logging activation Setting Off On Default Off Related If this parameter is set to Enabled the parameters Logging value 1 through Logging value 20 can be...

Страница 286: ... 4 h 6 h 12 h 24 h Default 5 s Related Path Maintenance and diagnostics SensorFlash Data logging Logging interval 4 7 5 4 Process values Logging value 1 20 Description The process value to be logged can be selected in this parameter Logged process values are an indicator that you have to stop logging before you can change the logged process values Setting Disabled Volume flow Mass flow Standard vo...

Страница 287: ...ensorFlash Data logging Advance logging expert Register 1 90 4 8 Simulation 4 8 1 Simulation inputs and outputs 4 8 1 1 Channel 1 output with HART 4 8 1 1 1 Simulation Description Enable or disable simulation of the current output Setting Disabled Enabled Default Disabled Related If this parameter is set to Enabled the parameter Simulation value can be configured Path Maintenance and diagnostics S...

Страница 288: ...ate Setting Off Current output Frequency output Pulse output Status output Default Off Related The setting of this parameter affects the setting options in parameter Simulation value Path Maintenance and diagnostics Simulation Simulation inputs and outputs Channel 2 output Operation mode 4 8 1 2 2 Simulation Description Enables or disables simulation of channel 2 current output Setting Disabled En...

Страница 289: ...tput simulation is enabled Setting 0 mA to 25 mA Default 0 mA Operation mode is set to Current output Setting 0 0 Hz to 12500 0 Hz Default 1 0 Hz Operation mode is set to Frequency output Setting 0 0 pulses s to 12500 0 pulses s Default Operation mode is set to Pulse output Setting 0 1 Default Operation mode is set to Status output Related Only if parameter Simulation is set to Enabled the setting...

Страница 290: ...elated Only if this parameter is set to Enabled then the setting of parameter Simulation value will be relevant Path Maintenance and diagnostics Simulate Simulate inputs and outputs Channel 3 input output Simulation 4 8 1 3 3 Simulation value Description Set the current value to be output on the channel when current output or current input simulation is enabled Set the frequency value to be output...

Страница 291: ...ting of this parameter will be relevant Path Maintenance and diagnostics Simulation Simulation inputs and outputs Channel 3 input output Simulation value 4 8 1 4 Relay 3 4 8 1 4 1 Simulation Description Enables or disables simulation of channel 3 relay output Setting Disabled Enabled Default Disabled Related Only if this parameter is set to Enabled then the setting of parameter Simulation value wi...

Страница 292: ...g Off Current output Frequency output Pulse output Status output Digital input Current input Default Off Related The setting of this parameter affects the setting options in parameter Simulation value Path Maintenance and diagnostics Simulation Simulation inputs and outputs Channel 4 input output Operation mode 4 8 1 5 2 Simulation Description Enables or disables simulation of Channel 4 input outp...

Страница 293: ...put simulation is enabled Setting 0 mA to 25 mA Default 0 mA Operation mode is set to Current output or Current input Setting 0 0 Hz to 12500 0 Hz Default 1 0 Hz Operation mode is set to Frequency output Setting 0 0 pulses s to 12500 0 pulses s Default 0 0 Operation mode is set to Pulse output Setting 0 1 Default 0 Operation mode is set to Status output or Digital input Related Only if parameter S...

Страница 294: ...ated Only if parameter Simulation is set to Enabled then the setting of this parameter will be relevant Path Maintenance and diagnostics Simulation Simulation inputs and outputs Channel 4 relay Simulated value 4 8 2 Simulation process values 4 8 2 1 Enable simulation Description Select the process value s to be simulated Setting Volume flow Mass flow Standard volume flow Flow velocity Sound veloci...

Страница 295: ...if Volume flow is marked in Enable simulation then the setting of this parameter will be relevant The range and default setting are dependent on the selected Unit Path Maintenance and diagnostics Simulation Simulation process values Volume flow Simulation value 4 8 2 3 Mass flow 4 8 2 3 1 Simulation value Description Set the Mass flow value to be simulated Setting 10800000000 000001 kg h to 108000...

Страница 296: ...ted Unit Path Maintenance and diagnostics Simulation Simulation process values Standard vol ume flow Simulation value 4 8 2 5 Flow velocity 4 8 2 5 1 Simulation value Description Set the Flow velocity value to be simulated Setting 400 m s to 400 m s Default 0 0 m s Related Only if Flow velocity is marked in Enable simulation then the setting of this parameter will be relevant The range and default...

Страница 297: ... values Sound velocity Simulation value 4 8 2 7 LiquIdent 4 8 2 7 1 Simulation value Description Set the Liquident value to be simulated Note Available with Oil Software option only Setting 100 0 m s to 2500 0 m s Default 1500 0 m s Related Only if Liquident is marked in Enable simulation then the setting of this parameter will be relevant Path Maintenance and diagnostics Simulation Simulation pro...

Страница 298: ...the HMI Default 0 000001 m2 s Related Only if Kinematic viscosity is marked in Enable simulation then the setting of this parameter will be relevant The range and default setting are dependent on the selected Unit Path Maintenance and diagnostics Simulation Simulation process values Kinematic vis cosity Simulation value 4 8 2 10 Pressure 4 8 2 10 1 Simulation value Description Set the Pressure val...

Страница 299: ...nd diagnostics Simulation Simulation process values Medium tem perature Simulation value 4 8 2 12 Standard density 4 2 8 12 1 Simulation value Description Set the Standard density value to be simulated Setting 0 0 kg m3 to 100000 0 kg m3 Default 1000 0 kg m3 Related Only if Standard density is marked in Enable simulation then the setting of this parameter will be relevant The range and default set...

Страница 300: ...tion process values Standard kine matic viscosity Simulation value 4 8 2 14 API gravity 4 8 2 14 1 Simulation value Description Set the API gravity value to be simulated Setting 150 0 API to 200 API Default 0 0 API Related Only if API gravity is marked in Enable simulation then the setting of this parameter will be relevant Path Maintenance and diagnostics Simulation Simulation process values API ...

Страница 301: ... SGU Default 0 0 SGU Related Only if Specific gravity is marked in Enable simulation then the setting of this parameter will be relevant Path Maintenance and diagnostics Simulation Simulation process values Specific gravity Simulation value 4 8 2 17 Standard specific gravity 4 8 2 17 1 Simulation value Description Set the Standard specific gravity value to be simulated Setting 0 0 SGU to 10 0 SGU ...

Страница 302: ...ing are dependent on the selected Unit Path Maintenance and diagnostics Simulation Simulation process values Auxiliary tem perature Simulation value 4 8 2 19 Standardizing factor 4 8 2 19 1 Simulation value Description Standardizing factor Setting 0 0 to 1000000 0 Default 0 0 Related Only if Standardizing factor is marked in Enable simulation then the setting of this parameter will be relevant The...

Страница 303: ...ms 4 8 3 1 Simulation mode Description The alarm simulation mode can be configured in this parameter Setting Off Simulate alarm ID number Simulate alarm class Default Off Related If Simulation mode is set to Simulate alarm ID number Alarms and its subcategories will be available If Simulation mode is set to Simulate alarm class Alarm class will be available Path Maintenance and diagnostics Simulat...

Страница 304: ...lable if Simulation mode is set to Simulate alarm ID number Path Maintenance and diagnostics Simulation Simulation alarms Alarms Sensor alarms 2 4 8 3 2 3 Process alarms 1 Description The simulation of single and multiple Process alarms 1 can be configured Setting Various process alarms can be selected Default None selected Related This parameter is only available if Simulation mode is set to Simu...

Страница 305: ...larms 4 Description The simulation of single and multiple Process alarms 4 can be configured Setting Various process alarms can be selected Default None selected Related This parameter is only available if Simulation mode is set to Simulate alarm ID number Path Maintenance and diagnostics Simulation Simulation alarms Alarms Process alarms 4 4 8 3 2 7 Totalizer alarms Description The simulation of ...

Страница 306: ... mode is set to Simulate alarm ID number Path Maintenance and diagnostics Simulation Simulation alarms Alarms Channel 1 alarms 4 8 3 2 9 Input output alarms Description The simulation of single and multiple Input output alarms can be configured Setting Various input output alarms can be selected Default None selected Related This parameter is only available if Simulation mode is set to Simulate al...

Страница 307: ...ance required Default None selected Related This parameter can only be accessed if Status icons is set to NAMUR Path Maintenance and diagnostics Simulation Simulation alarms Alarm class 4 9 Audit trail 4 9 1 Parameter change log Description Displays list of logged parameter changes Log shows parameter value before the change not the current value Path Maintenance and diagnostics Audit trail Parame...

Страница 308: ...irmware update change log 4 9 4 Clear FW update change log expert Description The firmware update log can be reset using this parameter Setting Cancel Ok Default Cancel Related Path Maintenance and diagnostics Audit trail Clear FW update change log 4 10 Self test 4 10 1 Display test Description The display can be tested using a wizard Path Maintenance and diagnostics Self test Display test 4 11 Re...

Страница 309: ...tenance and diagnostics Resets Set to default 4 11 2 Device restart write only Description Command to restart device Setting Cancel Ok Default Cancel Related Path Maintenance and diagnostics Resets Device restart 4 12 Firmware update Description The firmware of the device can be updated using a wizard Path Maintenance and diagnostics Firmware update ...

Страница 310: ...Maintenance and diagnostics 4 12 Firmware update FST030 308 Function Manual 04 2017 A5E35693328 AA ...

Страница 311: ...ice channel USB USB mode 5 1 2 Auto mode Description The Auto mode can be enabled disabled When auto mode is enabled a device connected via a USB cable is always considered an MSD Setting Disabled Enabled Default Related Path Communications Service channel USB Auto mode 5 1 3 MSD connect write only Description This command is used to connect disconnect to a mass storage device for example a PC Set...

Страница 312: ...ss Setting 0 to 63 Default 0 Path Communication HART Channel 1 Polling address SW 5 2 2 Polling address HW read only Description Hardware device address The HW device address is configured by DIP switch in the device Path Communication HART Channel 1 Polling address HW 5 2 3 TAG Description A unique short TAG name for the device or measurement point can be entered Setting User configurable Default...

Страница 313: ...nel 1 HART revision 5 2 6 Response preambles Description The number of response preambles can be defined Setting 5 to 20 Default 5 Path Communication HART Channel 1 Response preambles 5 2 7 Dynamic variable mapping Mapping of measured process variables The assignment of the measured process values to HART device variables PV primary variable SV secondary variable TV tertiary variable and QV quater...

Страница 314: ...PI gravity Base API gravity Specific gravity Base specific gravity Current output channel 1 Current output channel 2 Pulse output channel 2 Frequency output channel 2 Status output channel 2 Current input chan nel 3 Digital input channel 3 Current output channel 3 Pulse output channel 3 Fre quency output channel 3 Status output channel 3 Current input channel 4 Digital input channel 4 Current outp...

Страница 315: ...ic gravity Base specific gravity Current output channel 1 Current output channel 2 Pulse output channel 2 Frequency output channel 2 Status output chan nel 2 Current input channel 3 Digital input channel 3 Current output channel 3 Pulse output channel 3 Frequency output channel 3 Status output channel 3 Current input channel 4 Digital input channel 4 Current output channel 4 Pulse output chan nel ...

Страница 316: ...fic gravity Base specific gravity Current output channel 1 Current output channel 2 Pulse output channel 2 Frequency output channel 2 Status output chan nel 2 Current input channel 3 Digital input channel 3 Current output channel 3 Pulse output channel 3 Frequency output channel 3 Status output channel 3 Current input channel 4 Digital input channel 4 Current output channel 4 Pulse output chan nel...

Страница 317: ... ft3 h ft3 d Imperial gallons i gal s i gal min i gal h i gal d Default m3 h Related Path Communications HART Channel 1 HART units Volume flow units 5 2 8 2 Mass flow units Description The mass flow unit for the communication interface mass time can be configured Setting Metric Gram g s g min g h Kilogram kg s kg min kg h kg d Tonne t min t h t d US Short ton STon min STon h STon d Ton long ton T ...

Страница 318: ...The flow velocity units for the communication interface length time can be configured Setting Metric m m s m h US Inches in s in min Feet ft s ft min Default m s Related Path Communication HART Channel 1 HART units Flow velocity units 5 2 8 5 Sound velocity units Description The sound velocity units for the communication interface length time can be configured Setting Metric m m s m h US Inches in...

Страница 319: ... 8 Pressure units Description The pressure units for the communication interface can be configured Setting Metric mbar bar Pa kPa hPa MPa kgf cm2 mmH2O millimeters of water at 68 F mmH2O millimeters of water at 4 C mmHg millimeters of mercury at 0 C hPa hectopascal atm torr cmH2O centimeters of water at 4 C mH2O meters of water at 4 C cmHg centimeters of mercury at 0 C mHg meters of mercury at 0 C...

Страница 320: ...ed at the local display Setting Metric g grams kg kilograms t 1t 1000 kg US STon 1 STon 2000 lb T 1T 2240 lb oz ounces lb pounds Default kg Totalizer 1 Process value is set to Mass flow Setting Metric l hl m3 US gal i gal BBL31 5 BBL31 BBL42 in3 ft3 yd3 bush Default l Totalizer 1 Process value is set to Volume flow Related Path Communication HART Channel 1 HART units Totalizer 1 units 5 2 8 11 Tot...

Страница 321: ...mmunication interface can be configured The unit is not applied at the local display Setting Metric g grams kg kilograms t 1 t 1000 kg US STon 1STon 2000 lb T 1 T 2240 lb oz ounces lb pounds Default kg Totalizer 3 Process value is set to Mass flow Setting Metric l hl m3 US gal i gal BBL31 5 BBL31 BBL42 in3 ft3 yd3 bush Default l Totalizer 3 Process value is set to Volume flow Related Path Communic...

Страница 322: ...dizing factor Standard kine matic viscosity Standard density LiquIdent Density Medium temperature Default Volume flow Related Path Communication Modbus Channel 1 Damping Process values 1 5 2 9 3 Process values 2 Description Select the process values Setting API gravity Standard API gravity Specific gravity Standard specific gravity Rate of change Default API gravity Related Path Communication Modb...

Страница 323: ...rameter This address is used if the DIP switch is set to a value 0 Options Read only Related Path Communication Modbus Channel 1 Slave address HW 5 3 3 Baud rate Description Baud rate of Modbus interface Setting 9600 Bit s 19200 Bit s 115200 Bit s 38400 Bit s 57600 Bit s 76800 Bit s 1200 Bit s 2400 Bit s 4800 Bit s Default Related Path Communication Modbus Channel 1 Baudrate 5 3 4 Parity framing D...

Страница 324: ...o the right most byte Setting 1 0 3 2 0 1 2 3 2 3 0 1 3 2 1 0 Default Related Path Communication Modbus Channel 1 Floating point byte order 5 3 6 Integer byte order Description The integer byte order used in Modbus messages MSB most significant byte high byte LSB least significant byte low byte Setting MSB to LSB big endian LSB to MSB little endian Default Related Path Communication Modbus Channel...

Страница 325: ...ing 5 3 7 2 Source register 1 20 Description Modbus register that appears within Modbus request is redirected to the parameter specified by Target register 1 20 Setting Default Related Path Communication Modbus Channel 1 Register mapping Source register 1 20 5 3 7 3 Target register 1 20 Description Register of an existing product parameter to which a Modbus request is redirected Setting Default Re...

Страница 326: ... Communications Modbus Channel 1 Modbus units Mass flow unit 5 3 8 2 Volume flow unit Description Set the unit for volume flow values Setting Metric Litre l s l min l h l d Ml d hl s hl min hl h hl d m3 m3 s m3 min m3 h m3 d US Gallons Ugal s Ugal min Ugal h Ugal d Mgal d gal s gal min gal h gal d Barrels BBPS BBPM BBPH BBPD BOPS BOPM BOPH BOPD BLPS BLPM BLPH BLPD ft3 ft3 s ft3 min ft3 h ft3 d Cus...

Страница 327: ...bus Channel 1 Modbus units Standard volume flow 5 3 8 4 Flow velocity unit Description Set the unit for flow velocity values Setting Metric Meter m s m h US Foot ft s ft min Inch in s in min Custom unit Default m s Related Path Communications Modbus Channel 1 Modbus units Flow velocity unit 5 3 8 5 Sound velocity unit Description Set the unit for sound velocity values Setting Metric Meter m s m h ...

Страница 328: ...s Modbus Channel 1 Modbus units Density unit 5 3 8 7 Kinematic viscosity unit Description Set the unit for kinematic viscosity values Setting cSt Default cSt Related Path Communications Modbus Channel 1 Modbus units Kinematic viscosity unit 5 3 8 8 Pressure unit Description Set the unit for pressure values Setting Metric 0 C mmHg 0 C cmHg 0 C inHg 0 C mHg 0 C 4 C mmH2O 4 C cmH2O 4 C mH2O 4 C inH2O...

Страница 329: ...ons Modbus Channel 1 Modbus units Temperature units 5 3 8 10 Totalizer 1 unit Description Set the unit for totalizer 1 quantities Setting Metric l hl m3 US Ugal gal BBL42 BBL31 BBL31 5 bush yd3 ft3 in3 Custom unit Default m3 is set to Volume flow Setting Metric g kg t US lb STon T oz Custom unit Default kg is set to Mass flow Setting Metric Nl Nm3 Sl Sm3 US Sft3 Default Nl is set to Corrected volu...

Страница 330: ... Setting Metric g kg t US lb STon T oz Custom unit Default kg is set to Mass flow Setting Metric Nl Nm3 Sl Sm3 US Sft3 Default Nl is set to Corrected volume flow Related Path Communications Modbus Channel 1 Modbus units Totalizer 2 unit 5 3 8 12 Totalizer 3 unit Description Set the unit for totalizer 3 quantities Setting Metric l hl m3 US Ugal gal BBL42 BBL31 BBL31 5 bush yd3 ft3 in3 Custom unit D...

Страница 331: ... Nl Nm3 Sl Sm3 US Sft3 Default Nl is set to Corrected volume flow Related Path Communications Modbus Channel 1 Modbus units Totalizer 3 unit 5 3 8 13 Service Channel USB Description The USB mode can be read in this parameter Options USB communication Mass storage device MSD Related Path Communications Service channel USB USB mode ...

Страница 332: ...Communication 5 3 Modbus Channel 1 FST030 330 Function Manual 04 2017 A5E35693328 AA ...

Страница 333: ...user PIN code 6 2 Change expert PIN code expert Description The Change expert PIN code function can be accessed Setting Default Related Path Security Access management Change expert PIN code 6 3 Recovery ID Description Displays an ID number required to reset PIN if lost Record this ID number for use when contacting Siemens customer support Setting Default Related Path Security Access management Un...

Страница 334: ...ed Path Security Access management PUK 6 5 Activate user PIN Description The Enable access level control function can be configured Setting Disable Enable Default Enable Related Path Security Access management Enable access level control 6 6 Deactivate user PIN Description The Disable access level control function can be configured Setting Disable Enable Default Enable Related Path Security Access...

Страница 335: ...ault you will not be prompted for password for 10 minutes after the last key press When the parameter is set to On you will be prompted to enter the password before each HMI operation Setting Off On Default Off Related Path Security Access management Auto logout 6 8 Logout Description The Logout function can be configured Setting Default Related Path Security Access management Logout ...

Страница 336: ...Security 6 8 Logout FST030 334 Function Manual 04 2017 A5E35693328 AA ...

Страница 337: ...FST030 Function Manual 04 2017 A5E35693328 AA 335 Language 7 7 1 Language Description Select the local display language Setting English Deutsch Default English Related Language Language ...

Страница 338: ...Language 7 1 Language FST030 336 Function Manual 04 2017 A5E35693328 AA ...

Страница 339: ... on the Internet In addition to our documentation we offer a comprehensive knowledge base online on the Internet at Service and support http www siemens com automation service support There you will find The latest product information FAQs downloads tips and tricks Our newsletter providing you with the latest information about your products Our bulletin board where users and specialists share thei...

Страница 340: ...Support A 1 Technical support FST030 338 Function Manual 04 2017 A5E35693328 AA ...

Страница 341: ... current option list selection When you access an option list you can move the asterisk with the Up or Down Arrows to a new selection then press ENTER to select the item CLR Clear Key Use the CLR key to erase a numeric value or clear a selection from a multiple select option list Cursor This refers to the highlighted text and the arrow cursor that you move via the arrow direction when navigating t...

Страница 342: ... meter recovers its Initial Makeup parameters after a fault condition interrupts operation Installation Menu The flow meter s overall menu structure It allows you to define all aspects of operation for the flow meter Interface m s Refers to an alarm function that declares the passage of a liquid or gas interface by a comparison analysis of the relative sonic velocities of the two liquids or two ga...

Страница 343: ...he estimated sonic velocity measurement This Index can not be overridden by installer Numeric Data Refers to a value entered into a menu cell An example would be the pipe outer diameter Numeric Entry Refers to a number you type into menu cell that stores numeric data Numeric Keys Use the Numeric keys to type a numeric value where appropriate Op Sys ROM The Read Only Memory that stores the basic op...

Страница 344: ...arameters used by the flow meter to service a specific site or location The flow meter allows you to store several independent Site Setups Site Storage Memory Section of RAM allocated for permanent data storage This memory segment stores inactive site setups including a backup of active site The flow meter s Site Setup storage capacity depends on the dynamic memory allocation as dictated by each a...

Страница 345: ...signed by the flow meter This can be overridden by the installer TOTCNT A Totalizer pulse count function used for Batching or Sampling Transducer Also known as sensor Vaer The flow meter s aeration percent output Vps The sonic propagation velocity of a pipe Vs The sonic velocity of a liquid or gas ...

Страница 346: ...Glossary FST030 344 Function Manual 04 2017 A5E35693328 AA ...

Страница 347: ... 337 I Inputs and outputs 97 Internet Contact person 337 Support 337 L Languages 335 P Paths 33 Pipe settings 23 Process values 43 Q Quick commissioning 21 Quick start wizards 17 S Sensor model 30 Sensor sizes 43 Sensor test blocks 43 Sensor diagnostics 258 Service 337 Service channel USB 309 329 Support 337 T Temperature monitoring 263 Totalizers 82 U User calibration 38 ...

Страница 348: ...Index FST030 346 Function Manual 04 2017 A5E35693328 AA ...

Страница 349: ......

Страница 350: ...rives Process Automation 76181 Karlsruhe GERMANY Subject to change without prior notice Order No A5E35693328 Lit No A5E35693328 AA Siemens AG 04 2017 www siemens com processautomation A5E35693328 For more information www siemens com flow ...

Отзывы: