MAG 8000 & MAG 8000 CT
27
SFIDK.PS.026.D9.02
4. Commissioning
FT/PDM Meter
Display Parameter/data type
Factory settings
Data range
number
version view
Fix parameter or meter data that
not
are
changeable
717
All
-
Log period 22 totalized (2)
718
All
-
Log period 22 fault status
See 168
719
All
-
Log period 22 status information
See 169
720
All
L23
Date of log period 23
721
All
L23
Log period 23 totalized (1)
722
All
-
Log period 23 totalized (2)
723
All
-
Log period 23 fault status
See 168
724
All
-
Log period 23 status information
See 169
725
All
L24
Date of log period 24
726
All
L24
Log period 24 totalized (1)
727
All
-
Log period 24 totalized (2)
728
All
-
Log period 24 fault status
See 168
729
All
-
Log period 24 status information
See 169
730
All
L25
Date of log period 25
731
All
L25
Log period 25 totalized (1)
732
All
-
Log period 25 totalized (2)
733
All
-
Log period 25 fault status
See 168
734
All
-
Log period 25 status information
See 169
735
All
L26
Date of log period 26
736
All
L26
Log period 26 totalized (1)
737
All
-
Log period 26 totalized (2)
738
All
-
Log period 26 fault status
See 168
739
All
-
Log period 26 status information
See 169
800
Advanced
-
Insulation test enable
No
Yes / No
801
Advanced
-
Insulation test interval
30
0 to 65535
802
Advanced
-
Insulation value
803
Advanced
-
Insulation test date
2000-01-01 T00:00
year-month- day T
hours:minutes:seconds
804
Advanced
-
Insulation tests fulfilled
0
810
Advanced
-
Leakage detection mode
Off
Off/ fix value/ fix + lowest value
811
Advanced
-
Leakage source
Flow rate
Flow rate / volume
812
Advanced
-
Start period for leakage
120min=2:00 [24:00]
0 to 1430
detection minutes (0-23:50)
813
Advanced
-
Duration leakage detection
120min=2hours
1 to 144 (1440 minutes)
814
Advanced
-
Leakage value unit
Flow / volume unit
815
Advanced
-
Leakage limit
1
0 to 1*10^9
816
Advanced
-
Leakage excitation frequency
1.5625 Hz
1/15Hz, 1/5Hz, 1.5625Hz,
3.125Hz, 6.25Hz, 1/30Hz, 1/60Hz
817
Advanced
-
Leakage status
Leakage status; 1: Finished
successfully, 2: Leakage detection
running, 3: Leakage detection failed
(SystemStatus have fatal error), 4:
Leakage detection failed (Empty-pipe
detection disabled), 5: Leakage detection
failed (Coil current off), 6: Leakage
detection failed (Insulation test was
active during detection), 7: Leakage
detection stopped because leakage
parameter was changed.
818
Advanced
-
Periods with possible leakage
819
Advanced
-
Leakage periods before alarm
3
0 to 255
820
Advanced
-
Reset leakage period information No
Yes / No
821
Advanced
St1
Latest leakage period flowrate
0
822
Advanced
St1
Latest leakage period volume
0
823
Advanced
-
Lowest measured leakage value
1000000000
824
Advanced
-
Date of lowest leakage value
2000-01-01 T00:00
year-month-day T
hours:minutes:seconds
825
Advanced
-
Highest measured leakage value
0
826
Advanced
-
Date of highest leakage value
2000-01-01 T00:00
year-month-day T
hours:minutes:seconds
830
Advanced
R8
Next settling date
PS3 production date and time
year-month-day T
hours:minutes:seconds
831
Advanced
R9
Latest settling date
PS3 production date and time
year-month-day T
hours:minutes:seconds
832
Advanced
R9
Latest totalizer 1 value
0
Содержание SITRANS F M MAG 8000
Страница 60: ...MAG 8000 MAG 8000 CT 60 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Страница 61: ...MAG 8000 MAG 8000 CT 61 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Страница 62: ...MAG 8000 MAG 8000 CT 62 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Страница 63: ...MAG 8000 MAG 8000 CT 63 SFIDK PS 026 D9 02 10 Approvals and certificates ...