
Electrode resistance check
• Measure the electrode resistance between connections 82 and 0 with a moving coil meter.
With a sensor full of fluid the resistance should be between 5 KΩ and 50 KΩ.
If the sensor is empty the resistance will be infinite.
• Repeat the resistance measurements between connections 83 and 0.
The results should be the same.
If the resistance is low there may be a short on the electrodes or wiring (in the case of a remote
mounted transmitter). Alternatively there may be water ingress or moisture in the terminal box.
If the resistance is high and the pipe is completely full of fluid check the following:
1. Fluid is electrically conductive.
2. Electrodes are not coated with grease or any deposit.
3. Electrode circuit is not open.
4. Remote mounted transmitter has a 3 core cable with an overall shield continuously from
sensor to transmitter, including junction boxes and terminal rails inside panels.
5. Shield is connected to 0 or to earth terminal (PE) on sensor.
Note
Sensors removed from line
For sensors removed from line with dry bore, use megger between terminal 82 and
compression plate, and 83 and compression plate to show any water ingress behind
electrodes or within enclosure.
7.2
Fluctuating process values
Question
Why do the displayed process values fluctuate when the electrode cable is moved?
Troubleshooting/FAQs
7.2 Fluctuating process values
SITRANS FM MAG 3100
56
Operating Instructions, 12/2021, A5E03005599-AE