4.5
Installation with grounding rings
NOTICE
Only sensors with PTFE or PFA liners without grounding electrodes
Installation with grounding rings is only relevant for sensors with PTFE or PFA liners
• Mount grounding rings where necessary on non-conductive pipes.
Installation with grounding rings on sensors with
PTFE liner
Installation with flat grounding rings on sensors
with PFA liner
Note
Potential equalization
For potential equalization one grounding ring is sufficient.
By unidirectional flow it is recommended to mount the grounding ring on the inlet side.
• Connect the grounding rings to the sensor with the enclosed grounding straps.
Groundingstrap connection on type E grounding
rings
Grounding strap connection on flat grounding rings
Installing/Mounting
4.5 Installation with grounding rings
SITRANS FM MAG 3100
Operating Instructions, 12/2021, A5E03005599-AE
41