See also
Cable data (Page 79)
5.2
Remote installation
Note
Remote installation only
The following applies to remote installation of MAG 5000 / 6000 or MAG 6000 I.
1. Unscrew and remove terminal box lid.
2. Mount the two terminal blocks as shown and insert electrode cable plug
①
(terminals 82, 0
and 83) and coil cable plug
②
(terminals 85 and 86).
Terminals electrode cable: 82, 0 and 83
Terminals special electrode cable: 84, 83, 0, 82 and 81
Terminals coil cable: 85; 86
Note
Special electrode cable
Special electrode cable must be connected to terminals 84, 83, 0, 82 and 81.
Figure 5-1
Terminal box without blocks
Connecting
5.2 Remote installation
SITRANS FM MAG 3100
Operating Instructions, 12/2021, A5E03005599-AE
45