Reconstruction Error (aka Blob error)
This intermittent issue causes the data to not reconstruct. The scanner will acquire the data and it
will even show up in the inline display, but it does not create the reconstructed saved images.
Workaround:
1.
The only way to confirm that the data has reconstructed is to review your data in the
browser
2.
If it populates in the browser then it should be ok, however, as good general practice we
also recommend pulling it into the viewer to QA each sequence
3.
This glitch has only happened 3 times so far at BMC, but it has happened at many other
prisma sites as well, so please be sure to check your data as you are acquiring
4.
If this error does occur, you must stop scanning and do a
standby shutdown
on the
scanner to fix the glitch
5.
Note, this will not recover the data so it is important to check each sequence as you go
***This glitch has been reported to Siemens, but there is no way to know when and if
there will be a fix. This issue has not occurred since Dec 2017.