Revision Date: 3/13/2019
22
Prisma VE11C Software Bugs
This is a list of known software bugs/glitches. If you encounter any other suspected issues,
please report them to [email protected] and/or during study reporting.
AutoAlign Re-Run Failure
This glitch ONLY happens if you run a second autoalign or localizer scout. For example, if the
subject had to come out to use the restroom or needed to be re-positioned. If you require an
additional localizer for any reason, you MUST check all subsequent scans to make sure the
autoalign angle is properly applied. If you do not use autoalign then this does not affect you.
Workaround:
1.
Open each subsequent scan after the new autoalign localizer has been run and then click
the green check mark to apply autoalign
2.
When you open the sequence you should see the yellow FOV adjust
3.
Some sequences do not have a “working man” symbol
on the left side, so they will
not automatically open to be applied
4.
Please be vigilant in opening all subsequent sequences and/or add the working man after
rerunning the new localizer so they will automatically open and require application
5.
For those of you who have scheduled breaks built into to your study we highly
recommend that you turn on or make sure the working man is turned on for all post break
sequences
6.
The below linked SOP was made specifically for the HCP study, but it will give you an
idea of what to look for
http://www.bmap.ucla.edu/docs/HCP_AutoAlign_QA.pdf
***This glitch has been reported to Siemens, but there is no way to know when and if
there will be a fix.
AP/PA Copy Reference Issue
If you are using Siemens product sequences with different phase encoding directions (e.g. AP
then PA), you MUST NOT use the copy reference option. This glitch will change your second
PA sequence to RL. This glitch does not affect CMRR, HCP or ABCD sequences as the phase
encoding direction is controlled on the special card instead of the routine parameter card.
Workaround:
1.
You must manually input the angle for each sequence if you are manually prescribing
your FOV
2.
If you use autoalign and DO NOT manually adjust your FOV in any way then your scans
will not be affected and there would be no need to use copy reference
***This glitch has been reported to Siemens, but there is no way to know when and if
there will be a fix.