7ML19985CN01
Milltronics SF 500 – INSTRUCTION MANUAL
Page 37
mm
mmm
Reca
libr
at
ion
Initial Span
Perform a zero calibration prior to performing a span calibration.
Access P388 and enter EDIT mode
Press
If
Press
Press
Remove the test weight and return to RUN.
Direct Span
Direct span entry (P368) is intended for use when replacing software or hardware, and
when it is not convenient to perform an initial span. A record of the last valid span count
is required.
Access P368 and enter EDIT mode
Note:
Perform an initial span when a calibration out of range message is shown.
P388-01 Initial Span
E
Enter 1 to start INitial Span
0
Span Calibration. Current Span
41440
Setup test. Press ENTER to start
Zero should be done prior to Span
Setup test. Press ENTER to start
Initial Span Calibration in progress
Current Reading:
#####
Calibration complete. Deviation
0.00
Press ENTER to accept value
41900
Span Calibration. Current Span
41900
Setup test. Press ENTER to start
P368 Direct Span Entry
E
Enter Span Count
0
the deviation is reset
do a zero calibration or press
the span count being calculated
while calibration is in progress
e.g. the current span count
e.g. the new span value if accepted
e.g. the current span count
if unacceptable, press
to restart