Operation
4.3 Analyzer Functions
LDS 6
Operating Instructions, 01/2009, A5E00295894-05
53
51 Limits
Upper Limit[30.00] mg/Nm
51 Limits Ch1 NH
3
Lower Limit 0.00 mg/Nm
3
3
Limit Alarm On/Off
You can use this functions to set concentration limits. A limit alarm will be triggered when the
value of a component exceeds the permissible range set in this screen. The alarm can be turned
off here.
The limit alarm is displayed on the status line if this has been set up by function 53. The alarm is
also signalled by a relay if one has been assigned using function 71. The triggering of a limit
alarm is registered in the logbook (function 3).
52 Transmission
Set Nominal Value
52 Transmission Ch1 NH
3
Absolute 285.0 Units
Nominal 301.2 Units
Alarm Level[5 ] %
Transm. Alarm On/Off
Relative 84.93 %
You can use this function to set the transmission parameters. A transmission alarm is triggered
when the transmission is outside a permissible range. The lower alarm level of the transmission
for a channel is set as a percentage of the nominal transmission value, where the nominal value
is registered when the fourth soft key is pressed. The upper alarm level is set to a fixed value
and cannot be changed by the user.
The transmission alarm is displayed on the status line if this has been set up by function 53. The
alarm is also signalled by a relay if this has been set up by function 71. The triggering of a
transmission alarm is registered in the logbook
(function 3).
53 Status Messages
53 Status Messages Ch1 NH
3
[CTRL]
Display Func. Control
[LIM]
Display Limits
[STO]
Display Stored Value
Display Transm. Limits
[TR]
You can use this function to display - within the status line - up to four different statuses which
can be assumed by the analyzer.
Table 4- 7
Op. 3
Function
Status
STO: Stored Value
Analog output connected to memory (see also function 77).
LIM: Limit
Upward or downward violation of limit (see also function 51).
CTRL: Function Control
Start-up mode - Service mode.
TR: Transmission
Upward or downward violation of transmission limit (see also
function 52).