Technical Information
2.4 Configuration Examples
LDS 6
20
Operating Instructions, 01/2009, A5E00295894-05
2.4
Configuration Examples
A feature of the in-situ analytical procedure is that the physical measurement takes place
directly in the stream of process gas, and usually also directly in the actual process gas line.
All process parameters such as gas matrix, pressure, temperature, moisture, dust load, flow
velocity and mounting orientation can influence the measuring properties of the LDS 6 and
must therefore be systematically investigated for each new application.
&HQWUDOXQLW
+\EULGFDEOH
PD[OHQJWKP
6XSSOHPHQWDU\FKDQQHORSWLRQ
6XSSOHPHQWDU\FKDQQHORSWLRQ
7UDQVPLWWHUXQLW
5HFHLYHU
3URFHVVIODQJH
6HQVRUFRQQHFWLQJFDEOH
0HDVXUHPHQW
SDWKOHQJWK
*DVFRQFHQWUDWLRQ
)OXHJDV
FRPSRVLWLRQ
6WHDP
'XVWORDG
*DVYHORFLW\
*DVWHPSHUDWXUH
*DVSUHVVXUH
Figure 2-7
Typical transmitted light setup of LDS 6, in-situ
A feature of the standard applications defined in the ordering data of the LDS 6 is that the
typical process conditions are well-known and documented, and that the guaranteed
measuring properties can be proven by reference installations. If you cannot find your
application among the standard applications, please contact Siemens. We will be pleased to
check your possible individual application of the LDS 6. You can find an application
questionnaire on the LDS product sites on the Internet.
To avoid contamination of sensor openings on the process side, clean gaseous purging
media are used such as instrument air, N
2
or steam. Purging air tubes on the sensor heads,
which slightly penetrate into the process gas stream, define the effective measuring path
length.
7UDQVPLWWHUXQLW
&HQWUDOXQLW
5HFHLYHU
6DPSOHJDV
RXWOHW
6DPSOH
JDV
LQOHW
7HPSHUDWXUH
VHQVRU
+\EULGFDEOH
PD[OHQJWKNP
6XSSOHPHQWDU\
FKDQQHORSWLRQ
6XSSOHPHQWDU\
FKDQQHORSWLRQ
6HQVRUFRQQHFWLQJFDEOH
Figure 2-8
Typical transmitted light setup of LDS 6, in bypass
The LDS 6 can measure in both the transverse and longitudinal directions of the process gas
flow. In certain cases, the process conditions make it necessary to condition the sample gas
stream in a bypass line with respect to process temperature, pressure and/or optical path
length. Further treatment of the process gas, such as drying or dust precipitation, is usually
not necessary.