Mounting – Liquid Applications (cont’d)
Installation
•
Do not route cable openly. For optimum isolation against electrical noise, run cable
separately in a grounded metal conduit. Seal all thread connections to prevent
ingress of moisture.
•
Do not run cable near high voltage or current runs, contactors, and SCR control
drives. For pressure tight applications, install transducers hand tight plus ½ to 1½
turns.
•
PTFE tape or other appropriate sealant may be used to aid in sealing the threads for
use in pressure applications.
Note:
Installation shall only be performed by qualified personnel and in accordance with
local governing regulations.
WARNING:
Never attempt to loosen, remove, or disassemble process connection
while vessel contents are under pressure.
Flanged
coupling
bolt
gasket
customer flange, flat face only
Customer flanged standpipe. If a metal flange must be welded to pipe, refer
to Liquid Applications - Standpipes in the
Transducer Applications Manual
.
factory flanged transducer
nut
Blind Flange
nipple welded to blind flange
coupling
flange, gasket, and hardware
supplied by customer.
Note:
Tighten the flange bolts
evenly in order to ensure a good
seal between the mating flanges.
Caution: Overtightening can
cause performance
degradation.