Page EN-12
XPS Series Transducers – QUICK START MANUL
A5E32282889
For Canadian Hazardous Installation Only
Product Group
FOR INTERNAL USE ONLY
SIEMENS CANADA LIMITED
SIEMENS MILLTRONICS
PROCESS INSTRUMENTS
PETERBOROUGH, ONTARIO, CANADA
00027003
T. VU
PETERBOROUGH
01 / APRIL /1996
ULTRASONICS
FOR CONSTRUCTION
ADDED CANADIAN INSTALLATION STATEMENT
PER ECN 96-000-0-M028
ADDED CLASS 1, DIV 1, HAZARDOUS LOCATIONS ONLY
TO NOTE 2 PER ECN 98-885-0-0011
REVISED TITLE BLOCK PER ECN P5A00004758
REMOVED MILLTRONICS REFERENCE
0
1
2
3
S. NGUYEN
B. GRAY
1 : 1
1
1
NTS
A
XPS/XCT TRANSDUCER SERIES
INTERCONNECTION DIAGRAM
0-8850004Z-DI-A
3
BG
BG
RPC
PJF
TV
ALC
SN
MC
APR
01/96
AUG
008/96
AUG
12/98
JULY
03/2012
NON - HAZARDOUS LOCATION
(SAFE)
HAZARDOUS LOCATION
CSA CLASS I, GROUP A,B,C,D
CSA CLASS II, GROUP E,F,G
JUNCTION BOX
CLASS I & II
XPS/XCT SERIES
TRANSDUCER
XPS/XCT SERIES
TRANSDUCER
METAL CONDUIT
METAL CONDUIT
CERTIFIED
CONDUIT
SEAL
FITTING
CERTIFIED
CONDUIT
SEAL
FITTING
CERTIFIED
CONDUIT
SEAL
FITTING
CERTIFIED
CONDUIT
SEAL
FITTING
SIEMENS
CONTROLLER
SIEMENS
CONTROLLER
1) THE TRANSDUCER CABLE MUST BE TERMINATED IN AN AREA EXTERNAL
TO THE HAZARDOUS LOCATIONS, CLASS I & II GROUP A,B,C,D,E,F & G
2) A SEAL SHALL BE INSTALLED WITHIN 50mm OF THE TRANSDUCER CONDUIT HUB
FOR CLASS I, DIV. 1 HAZARDOUS LOCATIONS ONLY.
3) INSTALLATION SHALL BE DONE IN ACCORDANCE WITH CEC
PART 1 REQUIREMENTS.
NOTE:
NON COAXIAL
COAXIAL
See Siemens Milltronics drawing number 0-8850004Z-DI-A for current drawing