Scanning
5-34
ACUSON Cypress™ Operator’s Manual
Scale
Changes the velocity
limits
Setting the Scale higher increases the velocity limits
of the color flow signal and reduces aliasing.
Setting the Scale lower decreases the velocity limits
of the color flow signal. This may be useful in
detecting low velocity blood flow.
Press the corresponding knob to select
SCALE
.
To increase the velocity setting, turn the knob
clockwise.
To decrease the velocity setting, turn the knob
counterclockwise.
Filter
Applies a velocity filter
to Color Doppler
The settings are 1—6. A setting of 6 applies the
maximum velocity filter.
Press the corresponding knob to select
FILTER
.
To increase the velocity filter, turn the knob clock-
wise.
To decrease the velocity filter, turn the knob
counterclockwise.
TABLE 18.
Color Flow Scanning Soft Windows Options
Window/Function
Description