Transferring Studies
9-12
ACUSON Cypress™ Operator’s Manual
DICOM Setup For Storage Commitment
If the optional networking package is installed, the
DICOM
Storage Commitment
Server information fields must be completely defined before the ability to per-
form Storage Commitment is enabled.
To specify your Storage Commitment settings, follow these steps:
1.
Press the
SETUP
function key.
2.
Double-click the
DICOM
directory listing and complete the Server Information
detailed in steps 3-9:
3.
Click the
Storage Commitment
subtitle.
4.
Click the [
ADD
] button.
5.
Identify your server by typing a user-defined name in the
Server Name
Field
.
6.
Set the
Application Entity
. This defines the server application that receives
Storage Commitment messages from the Cypress system.
7.
Set the
IP Address
. This is the IP address of the Storage Commitment
where the query will be made.
8.
Set the
Port Number
. This is the remote port on the server that is to receive
the connection request. The default setting is 104.
9.
The
Test Result
field lets you know the test result of the connection to the
Storage Commitment server. This field is used in conjunction with the [
PING
]
and [
VERIFY
] buttons. Before using these buttons, verify with your IT depart-
ment that your Storage Commitment server will respond to a ping and a
DICOM
Verify. Some systems will not respond to a ping due to network secu-
rity issues. Some Storage Commitment servers may not respond to a
DICOM
Verify command — consult the server's
DICOM
Conformance Statement.
After changing or setting up the
DICOM
server settings, test the connection
by clicking the [
PING
] button. The system will test the connection and indi-
cate its results in this field. If the ping is successful, test the
DICOM
connec-
tion by clicking the [
VERIFY
] button.