EC-1 Needle Guide Bracket Kit for Endocavity Transducers
WARNING:
The EC-1 Needle Guide Bracket kit is packaged non-sterile. Sterilize
this product prior to its first use.
Prior to sterilization, clean the bracket assembly.
To clean the bracket assembly:
1. Rinse the bracket under water and remove all debris. Avoid using a
brush as it may damage the needle guide bracket.
2. Visually inspect the guide to make sure all debris is removed.
3. Wipe off water on the bracket assembly using a sterile cloth or sterile
gauze.
To sterilize the bracket assembly:
Use sterile technique to perform this procedure.
1. Sterilize assembly using high-pressure steam sterilization (130°C [266°F]
for an exposure time of 10 minutes), EOG (ethylene oxide) sterilization,
or hydrogen peroxide plasma sterilization.
2. Perform one of the gas sterilization methods listed below:
Procedure
Sterilization
Method
Gas
Concentration
Temperature Humidity
Pressure Time
Ethylene Oxide Gas
10%
50°C (122°F)
50%
Maximum
operating
pressure
980 hPa
(1.0 kgf/cm
2
)
7 hours
Hydrogen Peroxide
Plasma
STERRAD
For details, refer to the operation manual for the STERRAD®
sterilization system.
3. After gas sterilization, thoroughly degas (aerate) the biopsy adapter to
remove all gas residues.
Содержание ACUSON Antares
Страница 2: ......
Страница 4: ...ii 1 I N S T R U C T I O N S F O R U S E ...
Страница 6: ...iv 1 I N S T R U C T I O N S F O R U S E ...
Страница 10: ...viii 1 I N S T R U C T I O N S F O R U S E ...
Страница 12: ...1 Introduction 1 2 1 I N S T R U C T I O N S F O R U S E ...
Страница 180: ...4 System Setup 4 36 1 I N S T R U C T I O N S F O R U S E ...
Страница 182: ...5 Beginning an Examination 5 2 1 I N S T R U C T I O N S F O R U S E ...
Страница 200: ...5 Beginning an Examination 5 20 1 I N S T R U C T I O N S F O R U S E ...
Страница 230: ...6 Technical Description 6 30 1 I N S T R U C T I O N S F O R U S E ...