12. Rotate the cartridge holder with the compartment door partially closed to
locate and remove any additional liquid.
Figure 33 Rotating cartridge compartment
13. Dampen a sponge swab with water or ethanol.
14. Clean the cartridge holder, rotating the cartridge holder as necessary.
15. Locate the vertical grooves inside the cartridge compartment.
16. Locate the front and back slots near the top of the compartment.
17. Locate the leaf spring on one side of the cartridge return spring.
18. With the leaf spring orientated toward the back of the system, complete
the following steps to lower the leaf spring into the system:
a. Hold on to both sides of the cartridge return spring.
b. Pinch the sides together and lower the spring into the system by
sliding the sides of the spring between the vertical grooves in the
compartment.
c. Release the spring.
CAUTION
Do not allow liquid to drip off of the sponge swab into the system.
If liquid drips into the system, you can damage the optics.