The printer stops printing when it runs out of paper. If no paper is available at
the start of printing, the printer does not attempt to print.
Microalbumin/Creatinine Test
The following sections explain the test sequence for a Microalbumin/Creatinine
test.
Preparing Patient Samples and Controls for the Microalbumin/Creatinine
Test
Materials Required:
•
DCA 2000 or DCA Systems Microalbumin/Creatinine Reagent Kit (For use
with the DCA Vantage system)
•
Patient Sample
•
Creatinine Control Kit or other control
•
Lint-free tissue
For information on preparing controls, refer to
Quality Control
‚ page 83.
For information on calibrating, refer to
Calibrating for Microalbumin/Creatinine
‚
page 81.
Opening the foil package
1. Remove one foil package (containing a reagent cartridge) from storage.
For more information see
Recommended Procedures for Handling
Reagent Cartridges
in the DCA Systems Microalbumin/Creatinine Reagent
Kit package insert.
CAUTION
Do not use scissors to cut open foil package. Scissors can
damage the reagent cartridge, the flexible pull-tab on the cartridge
or the desiccant bag.
Содержание DCA Vantage
Страница 1: ...From Wilburn Medical USA 1 877 WILBURN 945 2876 www wilburnmedicalusa com ...
Страница 3: ...DCA Vantage Operator s Guide ...
Страница 15: ...14 DCA Vantage Operator s Guide ...
Страница 123: ...122 DCA Vantage Operator s Guide ...
Страница 125: ...124 DCA Vantage Operator s Guide File Management ...
Страница 131: ...130 DCA Vantage Operator s Guide System Configuration ...
Страница 161: ...160 DCA Vantage Operator s Guide ...