DCA Vantage Operator’s Guide
33
Op
era
ting t
h
e Syst
em
Performing a DCA Microalbumin/Creatinine Test
You can save unused capillary holders and use them with any lot of DCA
Systems Microalbumin/Creatinine reagent cartridges.
1. Remove a capillary holder and a plunger from the plastic bag.
2. Inspect the capillary holder for the presence of the parts (figure 17).
Figure 17 Capillary holder
If the capillary holder is missing any of the above parts, discard the
capillary holder; also discard it if the starch plug is at the bottom of the
capillary tube.
BIOHAZARD
Wear personal protective equipment, including safety glasses and gloves.
Use universal precautions. Refer to
Safety Information
‚ page 131 for
recommended precautions when working with biohazardous materials.
1
Glass Capillary
2
Starch Plug
3
Absorbent Material
4
Plunger