SPECIFICATIONS
9
9.3.2
Measured value recording
9.3.3
Enclosure specification
9.3.4
Ambient conditions
Measuring principle
Photometer, interference and gas filter correlation method
Spectral range
VIS version: 300 … 1200 nm
IR version: 1200 … 11000 nm
Measured components
All IR-/NIR-/VIS-active gases and liquids.
Simultaneously up to 6 components, e.g.,
CO, CO
2
, NO, NO
2
, N
2
O, HCl, NH
3
, H
2
O, hydrocarbons, Cl
2
.
External sensors
Read-in and processing of up to 4 external sensors.
Number of measuring ranges
2, automatic measuring range switch-over (adjustable)
Cross-sensitivity compensation
Max. 6 disturbance variables (also external disturbance
variables)
Detection limit
< 2% of respective measuring range
Zero drift
VIS: < 1% of full scale / day
IR: < 2% of full scale / week
Temperature effect
< 2% of respective measuring range / 10 K
Setting time t
90
Approx. 30 ... 120 s, adjustable specific for system and compo-
nent
Limit values
2 limit values per measured component
Sample point switching
Max. 8 sampling points
Enclosure material
Aluminum, coated
Weight
Approx. 30 kg (66 lb) (without cell)
Highest surface temperature
+47 °C (+117 °F)
Degree of protection
IP 65 (cell separate)
Ambient temperature
+5 … +40 °C (+40 … +105 °F)
Storage temperature
–20 … +60 °C (-5 … +140 °F)
Relative humidity
Max. 80% (non-condensing)