36
8013741/16LH/V1-7/2020-04| SICK
O P E R A T I N G I N S T R U C T I O N S | MCS300P Ex
Subject to change without notice
4
OPERATION
4.2
Status and classification
4.2.1
Status (operating state)
The respective operating state (e.g., measuring, heating, etc.) is displayed in the top line of
the operator panel.
4.2.2
Classification, LEDs
The classification (error status) is indicated by LEDs on the operator panel and recorded in
the logbook (SOPAS ET).
Classification
LED
Significance
Measuring
screen
Analog
outputs
[1]
Status signal
[2],[3]
Maintenance
The MCS300P Ex is switched to “
Maintenance
”
via the menu or program.
The status bar shows: “
Status: Maintenance
”
Actual
Frozen
[4]
According to setting
Uncertain
The
uncertain
measured value (e.g., outside
calibration range)
blinks
.
To view the cause: Press the <Diag> button.
Actual
Actual
According to setting
Maintenance request
Yellow
Irregularities (e.g., deviation from check cycle too
high) that require a review of the cause.
To view the cause: Press the <Diag> button.
Actual
Actual
According to setting
Failure
Red
Device failure (e.g., lamp failed)
To view the cause: Press the
<Diag>
button.
Frozen
Frozen
According to setting
[1] Typical default (
→
System Documentation).
[2] Option (
→
System Documentation).
[3] See SOPAS ET in the “Digital outputs” menu.
[4] The last valid measured value is retained.