Low flow cut-off
A third panel in the “Communication” view provides the “Low Flow Cut-Off” parameter. For
volume flow rates being below this threshold, flow rate output will be set to “0”.
Once dropping below this threshold, an information message will be provided via the LC
display: I-1015 Low Flow Cut Off. The administrator user level is required for changing this
threshold.
Fig. 37
Low flow cut-off
After having applied changes, proceed as follows:
1
Press “Write to Device” to update the meter with the new information
2
Click on “Close” to return to the “Parameter Modification” view
4.3.6.6
Pressure/Temperature Sensors
The “Pressure/Temperature Sensors” view allows an authorized user to select pressure
and temperature value sources and define common settings such as units to be displayed
on the LCD and the update interval.
The following pressure sources are available:
● Fixed value
● HART
®
pressure sensor (absolute or relative)
● External live value (absolute or relative)
The following temperature sources are available:
● Fixed value
● Internal Sensor
● External
● HART
®
temperature sensor
The option “Fixed value” is used when no temperature or pressure sensor is connected.
After having applied changes, proceed as follows:
1
Press “Write to Device” to update the meter with the new information.
2
Click on “Close” to return to the “Parameter Modification” view.