54
FLOWSIC30 · Operating Instructions · 8020999/173B/V1-4/2020-02 · © SICK Engineering GmbH
Commissioning and operation
Su
bj
ec
t t
o
ch
an
ge
w
ith
ou
t n
ot
ice
4.3
Commissioning with the FLOWgate
TM
operating software
4.3.1
Required tools and accessories
● Service adapter (included in service toolbox, part no. 2089847)
● Service adapter drivers installed on the laptop/PC
● USB cable with connector type A to type B (not included in service toolbox)
● FLOWgate
TM
operating software and manual (provided on the product CD delivered with
the meter)
● Rights for parameter modification (see
4.3.2
Connecting to the device
1
Install the FLOWgate
TM
operating software on a laptop/PC.
2
Connect the FLOWSIC30 to this laptop/PC using the service adapter,
3
Start FLOWgate
TM
, either by selecting the “FLOWgate” entry in the program group
“SICK” or by double-clicking on the desktop icon.
4
Add the FLOWSIC30 to the device manager of the FLOWgate
TM
operating software and
connect to the device.
For detailed information, please refer to the FLOWgate software manual.