Commissioning and operation
FLOWSIC30 · Operating Instructions · 8020999/173B/V1-4/2020-02 · © SICK Engineering GmbH
57
Su
bj
ec
t t
o
ch
an
ge
w
ith
ou
t n
ot
ice
4.3.6
Commissioning wizard
▸
Establish an online connection between FLOWSIC30 and the FLOWgate
TM
operating
software.
▸
Log-in as user level “Admin” (standard password: 3333).
▸
Expand the menu navigation on the right border of the application window. Click on
“Commissioning” and switch the meter into Configuration Mode.
The commissioning wizard leads you step by step through the following procedures:
1
Identification
– Serial numbers
– Device name
2
System/User
– Units
– Device display settings
– User accounts
3
Analog Output
– Configuration
4
Communication
– HART
®
– RS485
– Low flow cut-off
5
Pressure / Temperature Sensor
– Common settings
– Pressure sensor settings
– Temperature sensor settings
6
Volume Converter
– Volume conversion
– Compressibility calculation
7
Archive / Logbook
– Diagnostic Archive
8
Wet gas detection
– Detection of wet gas
– Limits
9
Installation
– Meter installation
10
Finalize
– Finalize
– Generate reports
▸
Check the options and values in each step and adjust them according to the
application, if necessary.
▸
Create and print out a parameter report for reference purposes.
▸
Save to database.
NOTICE:
Function wet gas detection is automatically disabled when changing
the meter orientation from horizontal to vertical.