Commissioning and parameterization
FLOWSIC200 · Operating Instructions · 8013271/1CJ9/V2-0/2022-01 · © SICK Engineering GmbH
75
4.2.7
Configuring the analog inputs
In directory “Configuration / I/O Configuration / Analog Input / General Configuration”, the
standard analog inputs can be assigned to any measured values to be standardized in the
“Parameter analog input 1” and ”Parameter analog input 2” groups, and the respective
measuring range can be defined.
Fig. 62
Directory “Configuration / I/O Configuration / Analog Output / Analog Input“(example)
NOTICE:
Correction factors CC2, CC1 and CC0 are preset at the factory and may only be
changed by SICK Service.
Field
Parameter
Remark
Aim value in the MCU
measurement block
Measured value 1 to 8
Variable to be assigned to the selected analog input
Range low
Lower measuring range limit
Physical value at live zero
Range high
Upper measuring range limit Physical value at maximum current
Live zero
Zero point value > 0 mA)
Specification of the mA value for measurement range start
Indicate NAMUR error Inactive
No error is reported for underflow or overflow of the set current range
(LZ to 20 mA).
Active
An error is reported for underflow or overflow of the set current range
(LZ to 20 mA).
CC2
Square correction factor
Correction factors for calibrating the input size (factory preset)
Modification only by SICK Service!
CC1
Linear correction factor
CC0
Absolute correction factor
Содержание FLOWSIC200
Страница 1: ...O PERATING INSTRU CTIO NS FLOWSIC200 Flow Velocity Monitor Installation Operation Maintenance...
Страница 30: ...30 FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH Product description...
Страница 100: ...100 FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH Commissioning and parameterization...
Страница 104: ...104 FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH Maintenance...
Страница 121: ...Specification FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH 121...